POLARSTEM 75102073Recalls & reported problems
Smith & Nephew, Inc. · POLARSTEM VALGUS NON-CEMENTED WITH Ti/HA 1 · Class 2 · Implant
All models of POLARSTEM (63)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07611996142334), received between Nov 17, 2021 and Nov 17, 2021 — classified as 1 injury.
The most frequently reported specific device problem is “Migration” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K143739, K211176); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Migration1
Patient problems
- Bone Fracture(s)1
- Inadequate Osseointegration1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryNov 17, 2021Migration; Migration
IT WAS REPORTED THAT, AFTER A RIGHT THR SURGERY PERFORMED ON (B)(6) 2021, A (B)(6) FEMALE PRESENTED WITH INCREASING RIGHT HIP REGION PAIN ON MOBILIZATION. NO REPORTS OF TRIP, FALLS OR INCIDENTS. CT AND X-RAY SHOWED A FRACTURE OF THE PROXIMAL FEMUR WITH SIGNIFICANT STEM SUBSIDENCE. PATIENT MEDICAL HISTORY INCLUDES OVERWEIGHT (BMI NOT KNOWN). ORIGINAL POLARSTEM SUBSIDED, SO THE STEM AND THE HEAD WERE REMOVED. ORIF WAS COMPLETED AND A REDAPT STEM AND A NEW HEAD WERE IMPLANTED ON (B)(6) 2021. THE ORIGINAL SHELL AND LINER WERE RETAINED. PATIENT CURRENT HEALTH STATUS IS UNKNOWN DUE TO HOSPITAL CONFIDENTIALITY POLICY. NO FURTHER INFORMATION AVAILABLE.
Patient problems: Bone Fracture(s); Inadequate Osseointegration
MAUDE report 12828799 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K143739, K211176. One submission often covers many models, so check the product and lot details in each record.
- Z-0326-2025 · Oct 15, 2024 · Open, ClassifiedTandem Intl Bipolar Cocr Shell/UHMWPE Liner, TI-6AL-4V Retaining Ring, 28MM I.D. 46 MM O.D., REF 71324046, sterile, femoral head bipolar component of the Tandem Bipolar Hip System.
- Z-2827-2020 · Jul 22, 2020 · Open, ClassifiedPOLARSTEM COLLAR Reamer Guide, Catalog Number: 75102205, UPC Codes: 00885556535493 - Product Usage: Once the appropriate trial rasp size is in place, the adapter is removed and the
Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices
- BIOLOX DELTA 1365-32-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented551 reports
- PINNACLE ALTRX 1221-36-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented481 reports
- BIOLOX DELTA 1365-32-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death314 reports
- BIOLOX DELTA 1365-36-720Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death189 reports
- DJO SURGICAL 400-03-362Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented187 reports
- Anthology 71365705Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented176 reports
- BIOLOX DELTA 1365-40-710Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented170 reports
- PINNACLE GRIPTION 1217-32-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented169 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page