Skip to content

Trident 502-01-54ERecalls & reported problems

Howmedica Osteonics Corp. · Hemispherical Acetabular Shell · Class 2 · Implant

All models of Trident (160)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07613153073093), received between Jan 2, 2019 and May 28, 2019 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Osseointegration Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Osseointegration Problem2
  • Insufficient Device Problem Information1
  • Malposition of Device1
  • Use of Device Problem1

Patient problems

  • Injury3
  • Inadequate Osseointegration2
  • No Known Impact Or Consequence To Patient1
  • Osteolysis1
  • Pain1
  • Unspecified Infection1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 28, 2019Osseointegration Problem; Osseointegration Problem

    THE FOLLOWING REVISION EVENT WAS INCLUDED IN A REPORT RECEIVED AS PART OF THE (B)(6) IIS: PATIENT HAD A RIGHT PRIMARY THA DUE TO OSTEONECROSIS; UNDERWENT REVISION DUE TO INFECTED THA AND ACETABULAR LOOSENING; ALL COMPONENTS REVISED. PATIENT'S BMI IS REPORTED AS 35.5.

    Patient problems: Inadequate Osseointegration; Injury; Unspecified Infection

    MAUDE report 8645265 ↗
  • InjuryMar 12, 2019Osseointegration Problem; Osseointegration Problem

    IT WAS REPORTED THAT THE PATIENT'S LEFT HIP WAS REVISED DUE TO ASEPTIC LOOSENING OF THE SHELL. A 54E SHELL, POLY LINER, AND 36 +0 LFIT V40 HEAD WERE REVISED TO A TRIDENT REVISION SHELL, 36 LINER, AND 36 BIOLOX CERAMIC HEAD. REP CONFIRMED THAT NO ALLEGATIONS AGAINST THE LINER OR THE HEAD WERE REPORTED BY THE SURGEON OR ANY DOCUMENTATION SEEN. REP REPORTED THAT X-RAYS, MEDICAL RECORDS, AND ADDITIONAL INFORMATION ARE NOT AVAILABLE.

    Patient problems: Inadequate Osseointegration; Injury

    MAUDE report 8412225 ↗
  • InjuryFeb 5, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    PATIENT HAD A TOTAL HIP IN 2007 AND PRESENTED WITH RECENT HIP PAIN WITH X-RAYS INDICATING LYSIS AROUND THE SCREW IN THE CUP. UPDATE 11/JANUARY/2019: AS REPORTED IN ADVERSE CONSEQUENCE DETAILS: "PATIENT REQUIRED REVISION OF THE CUP WITH ADAPTOR SLEEVE AND CERAMIC HEAD". SPOKE TO REP. A STRYKER SCREW WAS ALSO REVISED. REP REPORTED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Injury; Osteolysis; Pain

    MAUDE report 8304961 ↗
  • MalfunctionJan 2, 2019Use of Device Problem; Malposition of Device; Use of Device Problem; Malposition of Device

    EXETER STEM BROKE 1.5 YEARS AFTER IMPLANTATION. X-RAY ATTACHED. MEDICAL REVIEW COMPLETED ON (B)(6) 2018 STATES; {...} A REVIEW OF THE PROVIDED MEDICAL RECORDS AND X-RAYS BY A CLINICAL CONSULTANT INDICATED: {...} CUP MALPOSITION IN LOW INCLINATION WITH LACK OF ANTEVERSION AND A MEDIALIZED LOWER CUP RIM IN THE ACETABULAR BONE IN COMBINATION WITH INAPPROPRIATE USE OF AN OFFSET CUP LINER AND USE OF A SKIRTED FEMORAL HEAD IN A MORBIDLY OBESE PATIENT HAVE CONTRIBUTED TO AN OVERLOAD CONDITION IN THE ARTHROPLASTY WITH FATIGUE ACCUMULATION AND ULTIMATELY A FATIGUE FRACTURE OF THE EXETER STEM NECK EXACTLY AT THE LEVEL OF PEAK OVERLOAD. {...}.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8209648 ↗

FDA clearance and classification

The FDA regulates product code KWL (prosthesis, hip, hemi-, femoral, metal) under 21 CFR 888.3360 as a Class II device. 21 CFR 888.3360 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Trident 502-01-54E

Has the Trident 502-01-54E been recalled?

No FDA recall record in our data names this exact device identifier (DI 07613153073093), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Trident 502-01-54E?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Jan 2, 2019 and May 28, 2019. In those reports, the most common specific device problems are “Osseointegration Problem” (2), “Malposition of Device” (1) and “Use of Device Problem” (1); the most common patient problems are “Injury” (3), “Inadequate Osseointegration” (2) and “Osteolysis” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Trident 502-01-54E?

None of the reports are classified as deaths, 3 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Trident 502-01-54E MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the Trident 502-01-54E, and is it still sold?

The FDA's device database lists Howmedica Osteonics Corp. as the company responsible for it (brand Trident, model 502-01-54E). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Prosthesis, hip, hemi-, femoral, metal devices

All Prosthesis, hip, hemi-, femoral, metal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page