Citation 6265-5117Recalls & reported problems
Howmedica Osteonics Corp. · 132 Degree Neck Angle V40 Hip Stem - Right · Class 2 · Implant
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613153074403), received between Aug 14, 2019 and Dec 12, 2019 — classified as 2 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Insufficient Device Problem Information2
Patient problems
- Injury2
- Bone Fracture(s)1
- Inflammation1
- Pain1
- Reaction1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryDec 12, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
PERIPROSTHETIC FRACTURE AROUND HIP IMPLANT.
Patient problems: Bone Fracture(s); Injury
MAUDE report 9459238 ↗InjuryAug 14, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT AFTER A RIGHT HIP REPLACEMENT IN 2011, PATIENT HAS BEEN EXPERIENCING PAIN SINCE THE FIRST YEAR THE IMPLANT WAS IN. SURGEON REPORTED A POTENTIAL PSEUDO-TUMOR AFTER TAKING MRI'S. PATIENT HAS NOT BEEN REVISED AS OF THE TIME OF REPORT. UPDATE ON 07/18/ 2019 BY QS: BASED ON THE PROVIDED MEDICAL RECORDS, THE REVISION SURGERY TOOK PLACE ON (B)(6) 2018. THE FOLLOWING INFORMATION IS ALSO NOTICED IN THE MEDICAL REVIEW: ¿ON (B)(6) 2014 STATES, ¿¿ CONTINUES TO DO WELL ¿ LEG LENGTH INEQUALITY FROM PRIOR HIP SURGERY MANY YEARS AGO; ¼¿ TO ½¿ SHORTER ON RIGHT¿ [¿]"ON (B)(6) 2018 AN OPERATIVE REPORT NOTES A PROCEDURE OF "1. RIGHT THA LIMITED REVISION FEMORAL HEAD AND POLY CHANGE; AND 2. RIGHT HIP EXCISION PSEUDO ACETABULUM, FOREIGN BODY GRANULOMA AND GANGLION CYST" FOR A PRE- AND POST- OPERATIVE DIAGNOSIS OF "(1) FAILED RIGHT TOTAL HIP ARTHROPLASTY SECONDARY TO TRUNNION METALLOSIS, POLYETHYLENE WEAR; AND (2) PSEUDO ACETABULUM WITH FOREIGN BODY GRANULATION AND GANGLION CYST FORMATION"." [...] "MODERATE ELEVATION OF ONE SERUM COBALT LEVEL AND NORMAL CHROMIUM LEVELS IN A PATIENT WITH BILATERAL TOTAL HIP ANGIOPLASTIES, ONE IN SITU EIGHTEEN YEARS AND THE OTHER SEVEN YEARS, IS NOT DI
Patient problems: Inflammation; Injury; Pain; Reaction
MAUDE report 8892496 ↗
Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices
- NA 1365-51-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented753 reports
- NA 1365-50-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented412 reports
- ARTICUL/EZE 1365-21-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall280 reports
- ARTICUL/EZE 1365-11-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 death262 reports
- LFIT 6260-9-236Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, cemented256 reports
- PINNACLE GRIPTION 1217-32-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented2 deaths239 reports
- ELEOS™ 25002111EOnkos Surgical, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented3 deaths230 reports
- Novation 130-36-52Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented2 recalls124 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.