NA 5400121000S1Recalls & reported problems
Stryker Corporation · Electric Handswitch, Left-Handed · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 07613154495634), received between Jun 11, 2019 and Jan 12, 2023 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component5
- Overheating of Device3
- Flaked2
- Material Fragmentation1
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Patient Involvement5
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 12, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THE USER FACILITY REPORTED THAT THE WIRE CLIP CAME OFF DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16146925 ↗MalfunctionMay 17, 2021Overheating of Device; Detachment of Device or Device Component; Overheating of Device; Detachment of Device or Device Component
THE USER FACILITY REPORTED THAT THE DEVICE OVERHEATED DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11832031 ↗MalfunctionMay 17, 2021Overheating of Device; Detachment of Device or Device Component; Overheating of Device; Detachment of Device or Device Component
THE USER FACILITY REPORTED THAT THE DEVICE OVERHEATED DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11832020 ↗MalfunctionOct 16, 2020Unintended Movement; Unintended Movement
THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE HAD RUN-ON WHILE IT WAS BEING DEMONSTRATED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10688287 ↗MalfunctionFeb 17, 2020Flaked; Flaked
THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE DEVICE HAD PAINT CHIPS MISSING DURING TESTING PRIOR TO A PROCEDURE AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Patient Involvement
MAUDE report 9717254 ↗MalfunctionFeb 17, 2020Flaked; Flaked
THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE DEVICE HAD CHIPPED PAINT DURING TESTING PRIOR TO A PROCEDURE AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Patient Involvement
MAUDE report 9717252 ↗MalfunctionFeb 12, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
THE USER FACILITY REPORTED THAT THE DEVICE DISASSEMBLED DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement
MAUDE report 9702372 ↗MalfunctionJan 30, 2020Material Fragmentation; Detachment of Device or Device Component; Material Fragmentation; Detachment of Device or Device Component
THE USER FACILITY REPORTED THAT A METAL ROD FELL OUT OF THE DEVICE DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement
MAUDE report 9647353 ↗MalfunctionJun 11, 2019Overheating of Device; Overheating of Device
THE MANUFACTURER SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED WHILE IT WAS BEING SERVICED AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 8688339 ↗
Other DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE devices
- RemB 6400034000Stryker Corporation · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1,618 reports
- NA XMAXAnspach Effort, The · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1,225 reports
- NA EG1ASynthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE859 reports
- NA CRANI-ASynthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1 recall772 reports
- NA 5450400000Stryker Corporation · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE775 reports
- NA 05.001.010Synthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE770 reports
- RemB 6400015000Stryker Corporation · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE603 reports
- NA QD11Synthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE580 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page