RemB 6400031000Recalls & reported problems

Stryker Corporation · Electric Oscillating Saw · Class 2

All models of RemB (7)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Stryker RemB electric oscillating saw is a handheld saw used in ear, nose and throat surgery. Its 378 reports mostly describe the saw overheating, staying on after the trigger is released, or leaking fluid. The reports include no injuries and no deaths. Reports fell from 11 in the previous 12 months to 0 in the last 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
378
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 378 adverse event reports that list this exact device (DI 07613154938308), received between Jul 24, 2018 and Jul 21, 2025 — classified as 378 malfunctions.

The most frequently reported specific device problem is “Device Remains Activated” (116 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 2 reportsJan 2024: 3 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 2 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 2 reportsSep 2024: 6 reportsOct 2024: 3 reportsNov 2024: 3 reportsDec 2024: 2 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 3 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Remains Activated116
  • Overheating of Device115
  • Fluid/Blood Leak70
  • Detachment of Device or Device Component49
  • Fracture15
  • Unintended Power Up8
  • Material Fragmentation4
  • Leak/Splash3

Patient problems

  • No Clinical Signs, Symptoms or Conditions256
  • No Patient Involvement101
  • No Consequences Or Impact To Patient22
  • No Known Impact Or Consequence To Patient10
  • Burn(s)1
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 378 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 21, 2025Overheating of Device; Overheating of Device

    THE DEVICE OVERHEATED DURING A SERVICE EVALUATION AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22557608 ↗
  • MalfunctionJul 11, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE DEVICE OVERHEATED DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22481806 ↗
  • MalfunctionJul 11, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE DEVICE OVERHEATED DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22480878 ↗
  • MalfunctionMay 2, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE DEVICE OVERHEATED DURING TESTING. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21950002 ↗
  • MalfunctionApr 22, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE DEVICE OVERHEATED DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21881398 ↗
  • MalfunctionJan 8, 2025Device Remains Activated; Device Remains Activated

    AS PER SERVICE TECHNICIAN, DEVICE KEPT RUNNING AFTER TRIGGER WAS RELEASED. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21105099 ↗
  • MalfunctionDec 9, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE DEVICE WAS LEAKING OIL DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20885104 ↗
  • MalfunctionDec 5, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE DEVICE WAS LEAKING DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20855695 ↗
  • MalfunctionNov 25, 2024Device Remains Activated; Device Remains Activated

    THE DEVICE KEPT RUNNING DURING A SERVICE EVALUATION AT THE MANUFACTURER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20771105 ↗
  • MalfunctionNov 25, 2024Device Remains Activated; Device Remains Activated

    THE DEVICE KEPT RUNNING DURING A SERVICE EVALUATION AT THE MANUFACTURER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20771081 ↗
  • MalfunctionNov 6, 2024Overheating of Device; Overheating of Device

    THE DEVICE OVERHEATED DURING A SERVICE EVALUATION AT THE MANUFACTURER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20621721 ↗
  • MalfunctionOct 31, 2024Device Remains Activated; Device Remains Activated

    THE DEVICE HAD RUN ON DURING A SERVICE EVALUATION AT THE MANUFACTURER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20579805 ↗
  • MalfunctionOct 31, 2024Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE DEVICE GENERATED HEAT DURING A CASE AT THE USER FACILITY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20579801 ↗
  • MalfunctionOct 8, 2024Overheating of Device; Overheating of Device

    THE DEVICE OVERHEATED DURING A SERVICE EVALUATION AT THE MANUFACTURER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20396460 ↗
  • MalfunctionSep 16, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE DEVICE HAD A BROKEN BLADE STUCK IN THE DEVICE DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20231681 ↗
  • MalfunctionSep 16, 2024Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT THE DEVICE WAS SLOW BRAKING DURING A DEMO VISIT AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20230155 ↗
  • MalfunctionSep 11, 2024Material Disintegration; Material Fragmentation; Material Disintegration; Material Fragmentation

    IT WAS REPORTED THAT THE DEVICE WAS SPITTING OUT METAL FRAGMENTS DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20195251 ↗
  • MalfunctionSep 11, 2024Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE DEVICE WAS HEATING UP DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20194172 ↗
  • MalfunctionSep 11, 2024Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE DEVICE WAS HEATING UP DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20194139 ↗
  • MalfunctionSep 11, 2024Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE DEVICE OVERHEATED DURING A PROCEDURE AT THE USER FACILITY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20193923 ↗
  • MalfunctionAug 23, 2024Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE DEVICE WAS HOT DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20054084 ↗
  • MalfunctionAug 22, 2024Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT THE DEVICE HAD RUN ON FOUND DURING A DEMO VISIT AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20037829 ↗
  • MalfunctionJul 22, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE DEVICE APPEARED TO BE LEAKING DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19799289 ↗
  • MalfunctionMay 28, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THE DEVICE WAS LEAKING OIL DURING A CASE AT THE USER FACILITY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19403945 ↗
  • MalfunctionApr 10, 2024Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT THE DEVICE WAS SLOW BRAKING DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19079970 ↗

Other DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE devices

All DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page