AVIATOR 48811232Recalls & reported problems
Stryker Corporation · TWO-LEVEL ANTERIOR PLATE · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07613252152200), received between Mar 17, 2016 and Feb 17, 2022 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Material Deformation4
- Break2
- Bent1
- Detachment Of Device Component1
- Disconnection1
- Fracture1
- Material Integrity Problem1
Patient problems
- No Known Impact Or Consequence To Patient4
- Discomfort1
- Feeding Problem1
- Laceration(s)1
- Laceration(s) of Esophagus1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 17, 2022Fracture; Fracture
IT WAS REPORTED THAT THE SPRING BAR OF AN AVIATOR TWO LEVEL PLATE FRACTURED AND A SCREW MIGRATED POST-OPERATIVELY. THE PATIENT WILL BE MONITORED AND REVISION SURGERY HAS NOT BEEN SCHEDULED AT THIS TIME. THIS REPORT WILL CAPTURE THE PLATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13549032 ↗InjuryApr 10, 2019Break; Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING THE EXPLORATORY SURGERY DUE TO DISCOMFORT IN PATIENT'S THROAT, THE LOCKING MECHANISM OF THE PLATE WAS FOUND TO BE BROKEN AND CAUSED LACERATION TO THE ESOPHAGUS MUSCLE WHICH IN TURN HAD TO BE SEWN CLOSED. THE BROKEN PLATE WAS REPLACED. THIS PROCESS RESULTED IN SIGNIFICANT SURGICAL DELAY. HOWEVER, THE EXACT DURATION OF THE SURGICAL DELAY WAS NOT REPORTED.
Patient problems: Discomfort; Feeding Problem; Laceration(s); Laceration(s) of Esophagus
MAUDE report 8499068 ↗MalfunctionOct 9, 2018Break; Material Deformation; Material Deformation
SURGEON WAS PERFORMING A C5-C7 ACDF WITH THE PLATE SYSTEM. SURGEON WAS TIGHTENING A 4.35 SELF TAPPING SCREW AT C7 WHEN A SPRING BAR AT C5 POPPED UP. SURGEON REMOVED THAT PLATE AND APPLIED ANOTHER TO FINISH THE CASE. NO ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7950394 ↗MalfunctionJul 19, 2016Disconnection
PER REPORTED EVENT; SURGEON ATTEMPTED TO IMPLANT A 32MM 2 LEVEL PLATE WITH 4.35MM RESCUE SCREWS AND THE METAL LOCKING HINGE CAME DISCONNECTED FROM THE PLATE. SURGEON TOOK PLATE OUT AND INSERTED A NEW ONE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5804237 ↗MalfunctionMay 23, 2016Bent; Material Deformation
IT WAS REPORTED THAT; PLATE WAS BENT AND IMPLANTED INTO THE PATIENT. AFTER SURGEON TOOK AN XRAY AND UPON CLOSING, SURGEON NOTICED THE SILVER LOCKING MECHANISM WAS DEFORMED. TOOK THE PLATE OUT AND REPLACED WITH THE SAME SIZE PLATE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5673191 ↗MalfunctionMar 17, 2016Material Deformation; Material Integrity Problem; Detachment Of Device Component; Material Deformation; Material Integrity Problem
IT WAS REPORTED THAT THE DOCTOR WENT TO BACK A SCREW FROM THE PLATE AND THE EYELID CAME OFF.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5509286 ↗
Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- Elegance® Anterior Cervical Plate 53430353Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430108Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430114Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430240Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430236Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page