Skip to content

AVIATOR 48811357Recalls & reported problems

Stryker Corporation · THREE-LEVEL ANTERIOR PLATE · Class 2 · Implant

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07613252152323), received between Jun 12, 2017 and Feb 2, 2024 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Deformation Due to Compressive Stress1
  • Device Dislodged or Dislocated1
  • Device Slipped1
  • Fracture1
  • Material Deformation1
  • Material Protrusion/Extrusion1

Patient problems

  • No Code Available2
  • Injury1
  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 2, 2024Fracture; Fracture

    A COMPANY REPRESENTATIVE REPORTED THAT A PATIENT UNDERWENT REVISION SURGERY TO ADDRESS THE MIGRATION OF AN AVIATOR VARIABLE SELF-DRILLING SCREW AT LEFT-SIDE C7. DURING SURGERY, IT WAS DISCOVERED THAT THE LOCKING WING OF THE AVIATOR THREE-LEVEL PLATE WAS FRACTURED AT THAT SAME LEVEL. THIS REPORT CAPTURES THE AVIATOR THREE-LEVEL PLATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18631238 ↗
  • InjurySep 24, 2018Device Slipped; Deformation Due to Compressive Stress; Material Protrusion/Extrusion; Deformation Due to Compressive Stress

    PLATE WAS IDENTIFIED WITH 2 LOWEST SCREWS BACKED OUT. PATIENT WAS BROUGHT TO SURGERY TO REMOVE THE TWO PROTRUDING SCREWS. THE 2 SCREWS WERE REMOVED AND THE PLATE REMAINS IMPLANTED.

    Patient problems: Injury; No Code Available

    MAUDE report 7903561 ↗
  • MalfunctionJun 12, 2017Break; Device Dislodged or Dislocated; Material Deformation; Material Deformation

    IT WAS REPORTED THAT; SURGEON WAS DOING A 3-6 ACDF WITH A 57MM PLATE. HE INITIALLY PUT ALL 8 SCREWS IN 2/3 OF THE WAY, NOT USING THE DRILL GUIDE. HE TIGHTENED DOWN THE SCREWS AT THE TOP LEVEL AND BOTTOM LEVEL, AND WHEN HE WENT TO TIGHTEN DOWN THE SCREWS AT THE MIDDLE LEVELS, HE NOTICED ONE OF THE WINGS HAD SNAPPED OUT OF THE PLATE. HE BACKED OUT ALL THE SCREWS, DAMAGING SOME OF THE OTHER LOCKING WINGS. A NEW 57MM PLATES WAS INSTALLED SUCCESSFULLY.

    Patient problems: No Code Available; No Known Impact Or Consequence To Patient

    MAUDE report 6635340 ↗

Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices

All APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.