AVIATOR 48824016Recalls & reported problems
Stryker Corporation · VARIABLE, SELF-DRILLING SCREW · Class 2 · Implant
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 07613252152613), received between Apr 10, 2019 and Jun 23, 2020 — classified as 6 injuries, 1 malfunction.
The most frequently reported specific device problem is “Migration” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Migration5
- Device Dislodged or Dislocated3
- Device Slipped2
- Unintended Movement2
- Appropriate Device Problem Term/Code Not Available1
Patient problems
- Discomfort4
- Injury4
- Feeding Problem2
- Laceration(s) of Esophagus2
- Failure of Implant1
- No Code Available1
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryJun 23, 2020Migration; Migration
IT WAS REPORTED THAT AN AVIATOR ASSY 3-LEVEL PLATE SECURING MECHANISM FRACTURED AND 3 AVIATOR SCREWS MIGRATED POST-OPERATIVELY THREE WEEKS AFTER IMPLANTATION. THE PATIENT HAS UNDERGONE REVISION SURGERY AND THE SCREWS HAVE BEEN EXPLANTED. THE PLATE REMAINS IMPLANTED. THIS REPORT WILL CAPTURE THE THIRD OF THREE SCREWS.
Patient problems: Injury
MAUDE report 10186781 ↗InjuryJun 23, 2020Migration; Migration
IT WAS REPORTED THAT AN AVIATOR ASSY 3-LEVEL PLATE SECURING MECHANISM FRACTURED AND 3 AVIATOR SCREWS MIGRATED POST-OPERATIVELY THREE WEEKS AFTER IMPLANTATION. THE PATIENT HAS UNDERGONE REVISION SURGERY AND THE SCREWS HAVE BEEN EXPLANTED. THE PLATE REMAINS IMPLANTED. THIS REPORT WILL CAPTURE THE FIRST OF THREE SCREWS.
Patient problems: Injury
MAUDE report 10186773 ↗InjuryJun 12, 2019Device Dislodged or Dislocated
PATIENT UNDERWENT REVISION SURGERY TO REMOVE TWO SCREWS WHICH BACKED OUT POST-OPERATIVELY. ADDITIONAL INFORMATION RECEIVED FROM PATIENT INDICATES THEY EXPERIENCED DISCOMFORT SWALLOWING AND, ¿PRESSURE ON THE BACK OF MY ESOPHAGUS,¿ PRIOR TO REVISION SURGERY. THIS REPORT CAPTURES THE FIRST OF THE TWO SCREWS.
Patient problems: Discomfort; Injury
MAUDE report 8692029 ↗InjuryJun 12, 2019Device Dislodged or Dislocated
PATIENT UNDERWENT REVISION SURGERY TO REMOVE TWO SCREWS WHICH BACKED OUT POST-OPERATIVELY. THIS REPORT CAPTURES THE SECOND OF THE TWO SCREWS."
Patient problems: Discomfort; Injury
MAUDE report 8692028 ↗MalfunctionApr 16, 2019Device Slipped; Migration; Migration
IT WAS REPORTED THAT THE PATIENT WAS SCHEDULED FOR 1 LEVEL ACDF REVISION PROCEDURE DUE TO DEGENERATIVE DISC DISEASE. RIGHT BEFORE THE PROCEDURE, X-RAY REVIEW REVEALED THAT SCREW FROM PLATE WAS BACKING OUT FROM PREVIOUS SURGERY. THEREFORE, THE SURGEON REPLACED THE PLATE AND SCREWS DURING REVISION SURGERY AND THIS RESULTED IN 30 MINUTES SURGICAL DELAY.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 8518411 ↗InjuryApr 10, 2019Device Dislodged or Dislocated; Unintended Movement; Appropriate Device Problem Term/Code Not Available; Migration; Unintended Movement; Migration
IT WAS REPORTED THAT THE PATIENT UNDERWENT REVISION SURGERY DUE TO SCREW PULL OUT POST-OPERATIVELY.
Patient problems: Discomfort; Failure of Implant; Feeding Problem; Laceration(s) of Esophagus
MAUDE report 8501409 ↗InjuryApr 10, 2019Device Slipped; Unintended Movement; Migration; Unintended Movement; Migration
IT WAS REPORTED THAT THE PLATE SECURING MECHANISM WAS FOUND DEFORMED/FRACTURED AT C7 LEVEL. THE REVISION SURGERY WAS EXTENDED AN ADDITIONAL HOUR DUE TO THE NEED TO CLOSE THE MEMBRANE SURROUNDING THE ESOPHAGUS. THE PATIENT WAS REQUIRED TO HAVE A FEEDING TUBE FOR A PERIOD OF TIME; DURATION NOT REPORTED. THE REVISION SURGERY WAS COMPLETED SUCCESSFULLY WITH A DIFFERENT BRAND CERVICAL PLATE. THERE WERE NO COMPLICATIONS DURING THE INITIAL SURGERY. IT WAS CONFIRMED THAT THE SCREWS WERE POSITIONED BELOW THE SPRING BAR AND THE PLATE¿S LOCKING MECHANISM WAS IN THE LOCKED POSITION. SCREW HOLES WERE PREPARED WITH VARIABLE AWL AND WERE IMPLANTED AT A DIFFICULT ANGLE. PATIENT WAS REPORTED TO HAVE GOOD BONE PURCHASE, DID NOT DO ANY STRENUOUS ACTIVITY, AND DID NOT EXPERIENCE A FALL.
Patient problems: Discomfort; Feeding Problem; Laceration(s) of Esophagus; No Code Available
MAUDE report 8499124 ↗
Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- Elegance® Anterior Cervical Plate 53430353Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430108Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430114Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430240Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430236Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.