VertaPlex 0406622000Recalls & reported problems

Stryker Corporation · High Viscosity Radiopaque Bone Cement · Class 2 · Implant

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07613252254430), received between Aug 12, 2022 and May 15, 2025 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Migration” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration2
  • Device Ingredient or Reagent Problem1
  • Expiration Date Error1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Pain1
  • Vertebral Fracture1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 15, 2025Migration; Migration

    NO NEW INFORMATION.

    Patient problems: Pain; Vertebral Fracture

    MAUDE report 22026466 ↗
  • MalfunctionSep 4, 2024Device Ingredient or Reagent Problem; Device Ingredient or Reagent Problem

    IT WAS REPORTED THAT DURING A PROCEDURE, THE MONOMER AND POLYMER DID NOT MIX TO THE CORRECT CONSISTENCY, CAUSING IT TO BE TOO RUNNY FOR USE. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20138223 ↗
  • InjuryMar 4, 2024Migration; Migration

    IT WAS REPORTED THAT DURING A SPINEJACK PROCEDURE AT T11, THE RADIOPAQUE MARKER FROM THE CANNULA PLUG REMAINED IN THE VERTEBRAL BODY OF THE PATIENT. THE SURGEON WAS ABLE TO POSITION THE SPINEJACKS. UPON CEMENT INJECTION FROM THE LEFT, A RAPIDLY SIGNIFICANT LEAKAGE OF CEMENT WAS OBSERVED REQUIRING IMMEDIATE SURGICAL INTERVENTION TO REMOVE THE LEAKED CEMENT. THE OPERATION LASTED AN EXTRA 2 HOURS. THE PATIENT'S CONDITION WAS CLEAR ON DISCHARGE (D+2) FOLLOWING EXTRACTION OF THE FOREIGN BODY (LEAKED CEMENT).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18826039 ↗
  • MalfunctionAug 12, 2022Expiration Date Error; Expiration Date Error

    IT WAS REPORTED THAT PRIOR TO A SURGICAL PROCEDURE AND UPON RECEIPT, THE CEMENT PRODUCT WAS OBSERVED TO BE EXPIRED. IT WAS ALSO REPORTED THAT THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, AND THERE WAS NO DELAY OR ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15218505 ↗

Other CEMENT, BONE, VERTEBROPLASTY devices

All CEMENT, BONE, VERTEBROPLASTY devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.