AVIATOR 48834018Recalls & reported problems
Stryker Corporation · FIXED, SELF-TAPPING SCREW · Class 2 · Implant
All models of AVIATOR (36)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613252254621), received between Sep 14, 2018 and Sep 14, 2018 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Mechanical Jam” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Mechanical Jam2
- Mechanical Problem2
- Break1
- Detachment of Device or Device Component1
Patient problems
- No Known Impact Or Consequence To Patient2
- No Code Available1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 14, 2018Mechanical Problem; Detachment of Device or Device Component; Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT A 22 MM PLATE WAS TO BE ATTACHED TO THE CERVICAL SPINE, USING 4MM X 18MM SELF-TAPPING SCREWS, PRE-DRILLED WITH 16MM PRE-DRILLS. 3 SCREWS COULD BE VERY TIGHTLY INSERTED WHILE 1 SCREW WAS OVER-TIGHTENED AND COULD NOT BE FIXED. THE PLATE'S LOCKING MECHANISM COULD NOT SLIDE OVER THE SCREW HEADS. WHEN UNSCREWING, PARTS OF THE SCREW THREAD WERE SHEARED OFF IN SITU. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A NEW 20 MM PLATE,16 MM SCREWS AND NEW PLACEMENT FOR THE PLATE ON THE CERVICAL VERTEBRA. THERE WAS A REPORTED 60 MINUTES DELAY AND NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7876476 ↗MalfunctionSep 14, 2018Break; Mechanical Problem; Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT A 22 MM PLATE WAS TO BE ATTACHED TO THE CERVICAL SPINE, USING 4MM X 18MM SELF-TAPPING SCREWS, PRE-DRILLED WITH 16MM PRE-DRILLS. THREE SCREWS COULD BE VERY TIGHTLY INSERTED WHILE 1 SCREW WAS OVER-TIGHTENED AND COULD NOT BE FIXED. THE PLATE'S LOCKING MECHANISM COULD NOT SLIDE OVER THE SCREW HEADS. WHEN UNSCREWING, PARTS OF THE SCREW THREAD WERE SHEARED OFF IN SITU. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A NEW 20 MM PLATE,16 MM SCREWS AND NEW PLACEMENT FOR THE PLATE ON THE CERVICAL VERTEBRA. THERE WAS A REPORTED 60 MINUTES DELAY AND NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Code Available; No Known Impact Or Consequence To Patient
MAUDE report 7876470 ↗
Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- Mariner RDX MX1-000040Seaspine Orthopedics Corporation · Appliance, fixation, spinal intervertebral body1 recall0 reports
- Mariner RDX MX1-000030Seaspine Orthopedics Corporation · Appliance, fixation, spinal intervertebral body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430112Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430224Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- XLP 7565050Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.