Skip to content

AVIATOR 48834518Recalls & reported problems

Stryker Corporation · FIXED, SELF-TAPPING SCREW · Class 2 · Implant

All models of AVIATOR (36)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613252254683), received between Sep 14, 2018 and Sep 14, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break1
  • Mechanical Jam1
  • Mechanical Problem1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 14, 2018Break; Mechanical Problem; Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT A 22 MM PLATE WAS TO BE ATTACHED TO THE CERVICAL SPINE, USING 4MM X 18MM SELF-TAPPING SCREWS, PRE-DRILLED WITH 16MM PRE-DRILLS. THREE SCREWS COULD BE VERY TIGHTLY INSERTED WHILE 1 SCREW WAS OVER-TIGHTENED AND COULD NOT BE FIXED. THE PLATE'S LOCKING MECHANISM COULD NOT SLIDE OVER THE SCREW HEADS. WHEN UNSCREWING, PARTS OF THE SCREW THREAD WERE SHEARED OFF IN SITU. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A NEW 20 MM PLATE,16 MM SCREWS AND NEW PLACEMENT FOR THE PLATE ON THE CERVICAL VERTEBRA. THERE WAS A REPORTED 60 MINUTES DELAY AND NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7876472 ↗

Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices