AVIATOR 48844016Recalls & reported problems
Stryker Corporation · FIXED, SELF-TAPPING SCREW · Class 2 · Implant
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07613252254713), received between Oct 17, 2016 and Apr 19, 2022 — classified as 3 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Fracture” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Fracture5
- Break2
- Appropriate Device Problem Term/Code Not Available1
- Device Dislodged or Dislocated1
- Malposition of Device1
- Material Integrity Problem1
- Migration1
- No Apparent Adverse Event1
Patient problems
- Injury2
- No Known Impact Or Consequence To Patient2
- No Clinical Signs, Symptoms or Conditions1
- No Code Available1
- No Consequences Or Impact To Patient1
- No Information1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 19, 2022Fracture; Fracture
A COMPANY REPRESENTATIVE REPORTED THAT THE HEADS OF TWO AVIATOR FIXED SELF-DRILLING SCREWS FRACTURED DURING SURGERY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A THIRTY-MINUTE SURGICAL DELAY AND NO ADVERSE CONSEQUENCE TO THE PATIENT. THIS REPORT CAPTURES THE FIRST OF TWO SCREWS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14141194 ↗InjuryJun 25, 2019Migration; Fracture; Device Dislodged or Dislocated; Migration
PHYSICIAN REPORTED THAT THE SELF-TAPPING SCREW APPEARED TO HAVE BACKED OUT VIA POST OP IMAGES. DURING EXPLORATORY REVISION SURGERY, IT WAS NOTED THAT THE LOCKING WING OF THE PLATE ON THE SAME LEVEL AS THE BACKED-OUT SCREW, FRACTURED. THE BACKED-OUT SCREW WAS ALSO NOTED TO HAVE FRACTURED. THE PHYSICIAN CONFIRMED ALL THE BROKEN PIECES WERE RETRIEVED. THE PATIENTS WOUND WAS INSPECTED WITH BLUE DIE BY AN ENT SURGEON WHO CONFIRMED THERE WERE NO HOLES IN THE ESOPHAGUS. RETRIEVAL OF THE BROKEN PIECES WAS COMPLETED SUCCESSFULLY WITH NO ADVERSE CONSEQUENCES. THE LOCKING WING WOULD HAVE BEEN FORCED FROM ITS POSITION AND EITHER DEFORMED OR FRACTURED AS THE SCREW BACKED-OUT. THIS RECORD REPRESENTS THE BACKED-OUT SCREW.
Patient problems: Injury
MAUDE report 8732909 ↗InjuryApr 10, 2019Malposition of Device; No Apparent Adverse Event; Appropriate Device Problem Term/Code Not Available; No Apparent Adverse Event
IT WAS REPORTED THAT THE PATIENT UNDERWENT REVISION SURGERY DUE TO SCREW PULL OUT POST-OPERATIVELY.
Patient problems: No Code Available; No Consequences Or Impact To Patient; No Information
MAUDE report 8501421 ↗InjuryOct 28, 2018Fracture; Fracture
IT WAS REPORTED THAT THE PLATE SPRING BAR POPPED OFF INSIDE PATIENT POST OP. PATIENT UNDERWENT REVISION TO REMOVE THE PLATE AND PUT A NEW ONE ON WITH RESCUE SCREWS.
Patient problems: Injury
MAUDE report 8011565 ↗MalfunctionMay 3, 2018Fracture; Break; Fracture
IT IS REPORTED THAT FIXATION PIN FOR THE PLATE BROKE AT THE DISTAL END DURING SCREWING IN. THE DISTAL END STAYED INTO THE PATIENT. THE ROUND HEAD PART OF THE SCREW BROKE APART. NEW SCREW WAS IMPLANTED. SURGICAL DELAY OF 30MIN IS REPORTED. NO ADVERSE CONSEQUENCES IS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7483033 ↗MalfunctionOct 17, 2016Fracture; Material Integrity Problem; Break; Fracture; Material Integrity Problem; Positioning Problem
IT WAS REPORTED THAT DURING SURGERY, IMPLANTING A PLATE, THE SURGEON USED AN AWL TO PREPARE THE SCREW HOLE. WHEN INSERTING THE FIXED SELF DRILLING SCREW THE SURGEON HAD TO REPOSITION THE SCREW. WHEN REDIRECTING THE SCREW THE TRAJECTORY (TABS ON THE OUTSIDE OF THE FEMALE HEX) ONE OF THE TABS BROKE OFF. THE BROKEN PIECE WAS RETRIEVED AND THE SAME SIZED SCREW USED TO COMPLETE THE SURGERY. NO ADVERSE CONSEQUENCES REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6034771 ↗
Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- Elegance® Anterior Cervical Plate 53430353Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430108Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430114Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430240Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430236Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.