Skip to content

SMART TOE ST0A-19PRecalls & reported problems

Stryker GmbH · Intramedullary Arthrodesis Implant · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07613252263432), received between Mar 11, 2019 and Sep 23, 2024 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Activation Failure” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation Failure1
  • Fracture1
  • Manufacturing, Packaging or Shipping Problem1
  • Premature Activation1

Patient problems

  • Failure of Implant2
  • Implant Pain1
  • Insufficient Information1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjurySep 23, 2024Premature Activation; Premature Activation

    AS REPORTED BY THE REP: "IMPLANTS EXPANDED IMMEDIATELY AFTER TAKING OUT FROM THE PACKAGING SO THE IMPLANT COULD NOT BE INSERTED." THE CUSTOMER CONFIRMED THAT THE IMPLANTS WERE STORED AT 0 DEGREES CELSIUS, OR BELOW, FOR A MINIMUM OF 2 HOURS BEFORE THE PROCEDURE AND WERE ONLY REMOVED FROM COLD STORAGE AT THE TIME OF IMPLANTING, AND NOT PRIOR. THE SURGEON HAD TO SWITCH TO INSERTING A TEMPORARY GUIDEWIRE TO TREAT THE PATIENT¿S HAMMERTOE. THIS WAS DIFFERENT TO WHAT WAS PLANNED BUT WAS THE ONLY WAY TO STABILIZE THE JOINT.

    Patient problems: Insufficient Information

    MAUDE report 20284633 ↗
  • MalfunctionAug 26, 2021Activation Failure; Activation Failure

    IT WAS REPORTED: "WHEN PLACING THE SMART TOE 2 IMPLANT DID NOT EXPAND UPON HEATING UP. NON EXPANDING IMPLANT WAS REMOVED AND EXCHANGED FOR DIFFERENT SMART TOE IMPLANT THAT PERFORMED AS EXPECTED. A SURGICAL DELAY OF 45MIN WAS REPORTED."

    Patient problems: Failure of Implant

    MAUDE report 12374426 ↗
  • InjuryNov 2, 2020Fracture; Fracture

    AS REPORTED: "PRIMARY SURGERY TOOK PLACE IN (B)(6). THE PATIENT REPORTED UNCHANGED PRE-EXISTING LOAD-DEPENDENT PAIN IN THE LEFT FOREFOOT, WITH THE TOES LL - LV EQUALLY PAINFUL. LEFT FOOT: INTEGUMENT INTACT WITH SCARRING WITHOUT IRRITATION. PAIN PROVOCATION WHEN MANIPULATING THE PRP JOINT DIG. LL - LV. DLP JOINT _DEG DIG. IV IN A SUBLUXATED PLANTAR POSITION. BEGINNING MALLET TOE SHAPE ALSO LL AND LLL, BUT CAN BE REPOSITIONED. SQUEEZE-IEST NEGATIVE. NO MULDER'S CLICK. MODERATE PLANTAR HYPERTENSION MTP V-JOINT. PO-NAS INTACT. X-RAY FOOT TO THE SIDE / OBLIQUE LEFT FROM (B)(6) 2020: NEW LYSIS BORDER AROUND THE DISTAL IMPLANT COMPONENT IN THE MIDDLE PHALANX IV AND FRACTURE OF THE PROXIMAL IMPLANT COMPONENT IN THE PROXIMAL PHALANX IV. THE SMART TOE IMPLANTS IN THE 2ND AND 3RD BEAM INTACT AND WITHOUT SIGNS OF EASING WITH ALSO ADVANCED CONSOLIDATION. WELL KNOWN ST.N. HAMME TOE CORRECTION DIG. LL-LV USING SMART TOE LMPLANTATEN WITH NEW FRACTURE AND LOOSENING OF THE IMPLANT IN DIG. LV AND LACK OF CONSOLIDATION THERE. NO PROGRESSIVE DEGENERATION. NOT A MORTON'S NEUROMA. A NEW CLINICAL AND RADIOLOGICAL RUPTURE OF THE SMART TOE IMPLANT OF THE DIG IS EVIDENT. IV. IN ADDITION, THERE WAS A SLIGHT L

    Patient problems: Failure of Implant; Implant Pain

    MAUDE report 10771191 ↗
  • MalfunctionMar 11, 2019Manufacturing, Packaging or Shipping Problem

    CUSTOMER REPORTED THAT INCORRECT IMPLANTS WERE FOUND IN PACKAGING UPON OPENING THE PRODUCT. THIS IMPLANT WAS NOT FITTED AND A NEW IMPLANT WAS USED WITH NO ISSUES.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8409347 ↗

Other PIN, FIXATION, SMOOTH devices

All PIN, FIXATION, SMOOTH devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page