SONICFUSION 1910-2000Recalls & reported problems
Stryker GmbH · Ultrasonic Generator · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 07613252495079), received between Mar 11, 2019 and Oct 27, 2023 — classified as 14 malfunctions.
The most frequently reported specific device problem is “Failure to Power Up” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Failure to Power Up6
- Device Displays Incorrect Message2
- Use of Device Problem2
- Display or Visual Feedback Problem1
- Image Display Error/Artifact1
- No Display/Image1
- Smoking1
- Visual Prompts will not Clear1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- No Known Impact Or Consequence To Patient4
- Failure of Implant1
- Foreign Body In Patient1
- No Patient Involvement1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 27, 2023Failure to Power Up; Failure to Power Up
AS REPORTED: "PLUGGED IN POWER CORD FOR SONIC ANCHOR GENERATOR AND IT NEVER POWERED UP. TRIED SECOND POWER GENERATOR CORD AND STILL DID NOT TURN ON".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18026185 ↗MalfunctionSep 28, 2023Failure to Power Up; Failure to Power Up
AS REPORTED: "IT WOULD NOT PASS THE STARTING TEST."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17839799 ↗MalfunctionSep 7, 2023Failure to Power Up; Failure to Power Up
AS REPORTED: "MALFUNCTION WITH SONIC ANCHOR CONSULT. LOCKDOWN ERROR L01000".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17697197 ↗MalfunctionSep 4, 2023Use of Device Problem; Use of Device Problem
AS REPORTED: "SURGEON STATES THAT BOTH 26MM SONIC PINS DID NOT FULLY LIQUIFY WHEN THE ULTRASONIC ENERGY WAS APPLIED. SURGEON STATED THE PINS LIQUEFIED DISTALLY AND PROXIMALLY BUT NOT AT THE CENTER ASPECT. SURGEON STATED THIS HAPPENED TWICE IN A ROW WITH THE 26MM PINS. SURGEON REMOVED THE FIRST 26MM PIN BUT ALLOWED THE 2ND 26MM PIN TO REMAIN IN THE PATIENT AFTER SHE REMOVED SOME OF THE HEAD. THE SURGEON USED A 22MM PIN RIGHT AFTER AND STATED SHE HAD NO ISSUE WITH THE 22MM PIN FULLY LIQUEFYING."
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 17676613 ↗MalfunctionJan 27, 2023Failure to Power Up; Failure to Power Up
AS REPORTED: "MY TEAM DID NOT HAVE A REP AVAILABLE TO COVER THIS CASE. AS I WAS IN A SEPARATE CASE, I STEPPED OUT FOR A PHONE CALL FROM MY COLLEAGUE WHO STATED THAT THE FACILITY NEEDED TO TROUBLESHOOT AND SHE WASN'T AWARE. I GAVE THEM A CALL AND WALKED THEM THROUGH THE STRYKER SONIC ANCHOR TROUBLE SHOOTING PER THE TECHNIQUE GUIDE. AS NOBODY WAS PHYSICALLY THERE, I DO NOT KNOW IF MY TROUBLESHOOT ACTIONS WERE TAKING. THE FACILITY STATES THAT THEY DID ALL OF THE ACTIONS (UNPLUGGING MACHINE, TAKING OFF THE TIP OF THE HANDPIECE AND RE ATTACHING, RESTARTING, RUNNING A TEST PRIOR). THEY CLAIM THAT THEY GENERATOR WAS FAULTY AND IT WAS NOT ALLOWING THEM TO RUN THE "TEST RUN". I WAS THEN HUNG UP ON BY THE FACILITY. I GET A TEXT MESSAGE FROM THE FACILITY MATERIALS MANAGER STATING THAT THE SONIC ANCHOR DID NOT WORK, AND HE SENT ME A PHOTO OF THE BOX. APPARENTLY THEY OPENED AND ANCHOR AND TRIED TO USE IT. I DO NOT KNOW EXACTLY WHAT HAPPENED HERE AS THEY SHOULD NOT HAVE OPENED THE ANCHOR PRIOR TO DOING THE TEST RUN. WHILE I WAS ON SPEAKER PHONE TRYING TO TROUBLE SHOOT, THE DOCTOR SAID "THIS PATIENT HAS BEEN UNDER FOR 3 HOURS, THIS ANCHOR DOESN'T WORK. THERE IS NOBODY HERE" AND PROCEEDING TO HANG
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16254617 ↗MalfunctionDec 3, 2021Use of Device Problem; Use of Device Problem
THE CUSTOMER REPORTED THE FOLLOWING: "ON (B)(6), 2021, THE SONIC ANCHOR IMPLANT WAS NOT 'MELTING' CORRECTLY AND WAS PULLED OUT OF THE PATIENT'S FOOT. THROUGH TRIAL AND ERROR, SEVERAL MORE IMPLANTS ALSO FAILED TO 'MELT' CORRECTLY. DRILLED AND THEN TRY TO IMPLANT BIOABSORBABLE PIECE. SONIC ANCHOR IS ONLY MELTING TOP PORTION OF IMPLANT. TROUBLESHOOTED THE MACHINE TODAY AND IT DID NOT WORK. SURGERY WAS COMPLETED WITH A SONIC ANCHOR FROM ANOTHER COMPANY. REPORTER WILL PROVIDE THE PRODUCT CATALOG INFORMATION FOR THE MACHINE LATER."
Patient problems: Failure of Implant
MAUDE report 12928514 ↗MalfunctionSep 16, 2021Failure to Power Up; Failure to Power Up
IT WAS REPORTED THE SONIC GENERATOR STOPPED TURNING ON, AND COULD NOT BE TURNED ON DURING SURGERY. A COMPETITOR DEVICE WAS USED TO COMPLETE THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12481592 ↗MalfunctionFeb 25, 2021Image Display Error/Artifact; Image Display Error/Artifact
AS REPORTED: GENERATOR WAS PLUGGED IN AND TURNED ON. GENERATOR GAVE A LOCKDOWN ERROR "L01000." ADDITIONALLY REPORTED: SURGEON USED A COMPETITIVE PRODUCT FOR FIXATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11382439 ↗MalfunctionNov 3, 2020Failure to Power Up; Failure to Power Up
IT WAS REPORTED THAT FOR A PROCEDURE TO PLACE SONIC ANCHORS IN THE PATIENT'S RIGHT FOOT, IT WAS DISCOVERED THAT THE 3-PRONG SOCKET IN THE BACK OF THE GENERATOR WAS SUNKEN IN AND THE UNIT WOULD NOT TURN ON. COMPETITOR ANCHORS WERE USED TO COMPLETE THE CASE SUCCESSFULLY WITH A DELAY OF APPROXIMATELY 5 MINUTES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10781443 ↗MalfunctionApr 13, 2020Visual Prompts will not Clear; Visual Prompts will not Clear
AS REPORTED: "THIS WAS OUR SECOND SONIC ANCHOR CASE OF THE DAY. OUR FIRST SONIC ANCHOR CASE WENT PERFECTLY AND THE MACHINE WAS FUNCTIONING FINE. DURING OUR SECOND CASE, THE MACHINE'S TOUCH SCREEN STOPPED WORKING. THE SCREEN WOULD NOT LET ME ADVANCE TO THE TEST SCREEN. BASICALLY WHEN YOU WOULD TOUCH THE SCREEN, NOTHING WOULD HAPPEN. SINCE WE COULD NOT ADVANCE TO THE TEST SCREEN, WE WERE NOT ABLE TO USE THE SONIC ANCHOR IMPLANT." ADDITIONALLY REPORTED: DELAY WAS "10-15 MINUTES TRYING TO TROUBLESHOOT". COMPETITOR DEVICE "MITEK SUTURE ANCHOR" WAS IMPLANTED INSTEAD.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9954496 ↗MalfunctionJan 9, 2020No Display/Image; No Display/Image
IT WAS REPORTED THAT DURING A PRIMARY FOOT PROCEDURE, THE SONIC GENERATOR WAS POWERED UP. THE TOUCH SCREEN WOULD NOT REGISTER ANY TOUCHES, AND A LOUD, HIGH-PITCHED NOISE CAME FROM WITHIN THE GENERATOR. THE SCREEN THEN GENERATED AN ERROR MESSAGE: "F00400 SYSTEM FAULT". THE DEVICE WAS PLUGGED INTO OTHER OUTLETS AND EXHIBITED THE SAME BEHAVIOR. SURGERY WAS COMPLETED SUCCESSFULLY WITH AN OVERALL DELAY OF APPROXIMATELY 25 MINUTES. THE REP CONFIRMED THAT NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9569766 ↗MalfunctionJul 8, 2019Display or Visual Feedback Problem; Display or Visual Feedback Problem; Device Displays Incorrect Message
AS REPORTED: "AN ERROR MESSAGE OCCURRED. THE MACHINE WILL NOT GO PAST THE ERROR SCREEN". REPORTER INDICATED THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT/ USER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8767255 ↗MalfunctionMay 10, 2019Device Displays Incorrect Message
WHEN RUNNING THE TEST ON THE GENERATOR IT WOULD FLASH 100% AND THE "X" AT THE SAME TIME, THEN REQUIRE A TEST AGAIN. COULD NOT GET PAST THE TEST SCREEN TO USE FOR PROCEDURE. EVENT OCCURRED IN OPERATING ROOM WHILE PATIENT WAS UNDER ANESTHESIA FOR A KIDNER PROCEDURE. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO ADVERSE CONSEQUENCES OR SURGICAL DELAY REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8601757 ↗MalfunctionMar 11, 2019Smoking
MACHINE STARTED SMOKING INSIDE OPERATING ROOM AND CLEANING CREW NOTICED. THIS HAPPENED AFTER HOURS AND DID NOT OCCUR DURING A CASE.
Patient problems: No Patient Involvement
MAUDE report 8409372 ↗
FDA clearance and classification
The FDA regulates product code HTY (pin, fixation, smooth) under 21 CFR 888.3040 as a Class II device. 21 CFR 888.3040 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the SONICFUSION 1910-2000
Has the SONICFUSION 1910-2000 been recalled?
No FDA recall record in our data names this exact device identifier (DI 07613252495079), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the SONICFUSION 1910-2000?
The FDA's MAUDE database lists 14 adverse event reports for this device, received between Mar 11, 2019 and Oct 27, 2023. In those reports, the most common specific device problems are “Failure to Power Up” (6), “Device Displays Incorrect Message” (2) and “Use of Device Problem” (2); the most common patient problems are “Failure of Implant” (1) and “Foreign Body In Patient” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the SONICFUSION 1910-2000?
None of the reports are classified as deaths, 0 as injuries and 14 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the SONICFUSION 1910-2000, and is it still sold?
The FDA's device database lists Stryker GmbH as the company responsible for it (brand SONICFUSION, model 1910-2000). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other PIN, FIXATION, SMOOTH devices
- RIGIDFIX 210133Medos International Sàrl · PIN, FIXATION, SMOOTH91 reports
- SMART TOE ST0A-16PStryker GmbH · PIN, FIXATION, SMOOTH1 recall5 reports
- SMART TOE ST0-19PStryker GmbH · PIN, FIXATION, SMOOTH1 recall2 reports
- SMART TOE ST0-22PStryker GmbH · PIN, FIXATION, SMOOTH1 recall0 reports
- SMART TOE ST0XS-11Stryker GmbH · PIN, FIXATION, SMOOTH1 recall0 reports
- ARROW-LOK 1Arrowhead De LLC · Pin, Fixation, Smooth1 recall0 reports
- ARROW-LOK 1Arrowhead De LLC · Pin, Fixation, Smooth1 recall0 reports
- ARROW-LOK 1Arrowhead De LLC · Pin, Fixation, Smooth1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page