Skip to content

iVAS 0306330000Recalls & reported problems

Stryker Corporation · Access Cannula · Class 1

All models of iVAS (10)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Stryker iVAS access cannula is used in vertebroplasty. Reports mostly describe the cannula fracturing or breaking during procedures, sometimes leaving a fragment in the patient. There are 41 adverse event reports naming this device identifier: no deaths, 12 injuries, and 29 malfunctions. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
42
Deaths reported
0
Injuries reported
12
Recalls naming it
0

Summary

The FDA's MAUDE database contains 42 adverse event reports that list this exact device (DI 07613252505990), received between Jun 15, 2018 and Aug 24, 2026 — classified as 12 injuries, 30 malfunctions.

The most frequently reported specific device problem is “Fracture” (31 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture31
  • Break6
  • Detachment of Device or Device Component3
  • Activation Failure1
  • Difficult to Remove1
  • Use of Device Problem1

Patient problems

  • Foreign Body In Patient24
  • No Clinical Signs, Symptoms or Conditions24
  • Insufficient Information3
  • No Consequences Or Impact To Patient3
  • Device Embedded In Tissue or Plaque2
  • Extravasation1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 42 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 24, 2026Fracture

    It was reported that the cannula broke inside the the patient, the case was stopped and the canula was removed the following day.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 50411877 ↗
  • MalfunctionJul 22, 2026Fracture

    This record is a consolidation of records summarized as a part of the FDA Voluntary Malfunction Summary Reporting program.\\Events occurred between (b)(6) 2026.\\This report summarizes 4 events for the failure: Fracture.\ - 4 events had no health consequences or impact.

    Patient problems: Foreign Body In Patient; Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 25908190 ↗
  • MalfunctionMay 12, 2026Fracture

    IT WAS REPORTED THAT A PIECE OF THE CANNULA WAS LEFT IN THE PEDICLE OF THE PATIENT, WHICH WAS CONFIRMED VIA X-RAY. THE PROCEDURE WAS INTENDED TO COMPLETE BOTH SIDES; HOWEVER, THE SECOND SIDE WASN'T COMPLETED. NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION TO PREVENT PERMANENT IMPAIRMENT WERE REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 25148059 ↗
  • MalfunctionApr 22, 2026Fracture

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1 ¿ MARCH 31 2026. THIS REPORT SUMMARIZES 3 EVENTS FOR THE FAILURE: FRACTURE. - 3 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: Foreign Body In Patient

    MAUDE report 24953814 ↗
  • InjuryFeb 27, 2026Fracture

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24471814 ↗
  • MalfunctionJan 23, 2026Fracture

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1 ¿ DECEMBER 31, 2025. THIS REPORT SUMMARIZES 3 EVENTS FOR THE FAILURE: FRACTURE. 2 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT. 1 EVENT HAD CODE NOT FOUND.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 24163103 ↗
  • MalfunctionOct 23, 2025Fracture; Fracture

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30 2025. THIS REPORT SUMMARIZES 2 EVENTS FOR THE FAILURE: FRACTURE. - 2 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: Foreign Body In Patient

    MAUDE report 23368320 ↗
  • MalfunctionJul 29, 2025Fracture; Fracture

    THIS REPORT SUMMARIZES 6 EVENTS FOR THE FAILURE: FRACTURE. 6 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: Foreign Body In Patient

    MAUDE report 22647061 ↗
  • MalfunctionApr 21, 2025Fracture; Fracture

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. 2 EVENTS HAD PATIENT INVOLVEMENT: NO PATIENT IMPACT.

    Patient problems: Foreign Body In Patient

    MAUDE report 21878601 ↗
  • MalfunctionJan 21, 2025Fracture; Fracture

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. TWO EVENTS HAD PATIENT INVOLVEMENT: NO PATIENT IMPACT.

    Patient problems: Foreign Body In Patient

    MAUDE report 21201191 ↗
  • MalfunctionOct 9, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE DURING THE PROCEDURE, THE ACCESS CANNULA BROKE WHEN IT WAS BEING REMOVED. PER A CT SCAN, IT BROKE OFF FLUSH WITH THE VERTEBRAL PEDICLE. IT SHOULD BE NOTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO PATIENT IMPACT, MEDICAL INTERVENTION, OR ADVERSE CONSEQUENCES TO THE PATIENT. THERE WAS A 20 MINUTE DELAY REPORTED AS WELL.

    Patient problems: Foreign Body In Patient

    MAUDE report 20408301 ↗
  • InjurySep 4, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE 11G ACCESS CANNULA BENT AND CONSEQUENTLY BROKE WHILE BEING REMOVED FROM VERTEBRAL BODY. IT SHOULD BE NOTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY, WITH NO PATIENT IMPACT, OR MEDICAL INTERVENTION. THERE WAS A SURGICAL DELAY OF 2 HOURS TO REMOVE THE BROKEN DEVICE FROM THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20136898 ↗
  • InjuryAug 21, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE NEEDLE BROKE DURING USE AND THE TIP OF THE DEVICE REMAINED IN THE PATIENT'S BONE AND COULD NOT BE REMOVED. IT SHOULD BE NOTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS A 2 HOUR DELAY TO TRY AND REMOVE THE BROKEN PIECE AS WELL.

    Patient problems: Foreign Body In Patient

    MAUDE report 20019508 ↗
  • MalfunctionJul 24, 2024Fracture; Fracture

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19818342 ↗
  • InjuryJun 13, 2024Fracture; Fracture

    THE USER FACILITY REPORTED THAT THE INCIDENT OCCURRED DURING THE PROCEDURE WHEN THE TROCAR WAS INTRODUCED INTO THE PATIENT'S PEDICLE. WHEN HE TRIED TO REMOVE THE TROCAR, IT BROKE AT BONE LEVEL BUT WAS ABLE TO BE EXTRACTED BY THE DOCTOR. IT SHOULD BE NOTED THAT THE PROCEDURE WAS SUCCESSFULLY COMPLETED, WITHOUT ANY ADVERSE CONSEQUENCES. THERE WAS A 60 MINUTE DELAY IN THE PROCEDURE DUE TO THE BREAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19528702 ↗
  • InjuryMay 1, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE NEEDLE BROKE OFF AND WAS STUCK IN THE PATIENT. THE NEEDLE WAS NOT ABLE TO BE REMOVED EITHER. THE PATIENT WAS SUBJECTED TO A LONGER SURGERY AS WELL AS A LARGER INCISION THAN WHAT WAS PLANNED. WE DID INQUIRE ABOUT THE SURGICAL DELAY BUT THE CUSTOMER DID NOT CLARIFY HOW LONG THE DELAY WAS.

    Patient problems: Foreign Body In Patient

    MAUDE report 19220893 ↗
  • InjuryApr 9, 2024Fracture; Fracture

    THE DOCTOR REPORTED THAT DURING THE PROCEDURE, THE OUTER CANNULA RIPPED OFF IN THE BONE. IT TOOK ABOUT 2 HOURS TO PROGRESSIVELY DRILL OVER IT AND REMOVE THE OUTER TABLE OF THE ILIAC BONE TO REMOVE IT. IT SHOULD BE NOTED THE PROCEDURE WAS DEEMED SUCCESSFULLY COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19070372 ↗
  • MalfunctionOct 24, 2023Fracture; Fracture

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. 2 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 17994383 ↗
  • MalfunctionApr 24, 2023Fracture; Fracture

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. - 2 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16801678 ↗
  • MalfunctionJan 29, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD A COMPONENT DETACH. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16261951 ↗
  • MalfunctionJan 29, 2023Fracture; Fracture

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: Foreign Body In Patient

    MAUDE report 16261868 ↗
  • MalfunctionOct 25, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15666905 ↗
  • MalfunctionJul 26, 2022Fracture; Fracture

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. - 2 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 15101964 ↗
  • MalfunctionMar 14, 2022Fracture; Fracture

    PER THE CUSTOMER, THE OUTER PART OF THE ACCESS CANNULA POPPED OFF THE HUB OF THE NEEDLE. THE DOCTOR WAS REPORTED TO BE EXPOSED TO THE PATIENT'S BLOOD DURING THE EVENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED FROM THE USER FACILITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13757068 ↗
  • InjuryFeb 17, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING THE PROCEDURE, THE NEEDLE/CANNULA BROKE OFF IN THE PATIENT AND THE CASE WAS NOT COMPLETED. MEDICAL INTERVENTION WAS NECESSARY BY ENLARGING THE PEDICLE HOLE TO REMOVE THE DEVICE MATERIAL LEFT BEHIND. A SURGICAL PROLONGATION WAS REPORTED. UPON FOLLOW UP, IT WAS REPORTED THAT THE DEVICE WAS SUCCESSFULLY REMOVED FROM THE PATIENT. IT WAS FURTHER REPORTED THAT THE PATIENT WAS DOING WELL AND NO ADDITIONAL PROCEDURE IS PLANNED AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13549554 ↗

FDA clearance and classification

The FDA regulates product code OAR (injector, vertebroplasty (does not contain cement)) under 21 CFR 888.4200 as a Class I device. 21 CFR 888.4200 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

FDA definition of this device type: Intended for injection of vertebroplasty bone cement for the fixation of pathological fractures of the vertebral body using vertebroplasty or kyphoplasty procedures. Painful vertebral compression fractures may result from osteoporosis, benign lesions (hemangioma), and malignant lesions (metastatic cancers, myeloma).

Common questions about the iVAS 0306330000

Has the iVAS 0306330000 been recalled?

No FDA recall record in our data names this exact device identifier (DI 07613252505990), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the iVAS 0306330000?

The FDA's MAUDE database lists 42 adverse event reports for this device, received between Jun 15, 2018 and Aug 24, 2026. In those reports, the most common specific device problems are “Fracture” (31), “Break” (6) and “Detachment of Device or Device Component” (3); the most common patient problems are “Foreign Body In Patient” (24), “Device Embedded In Tissue or Plaque” (2) and “Extravasation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the iVAS 0306330000?

None of the reports are classified as deaths, 12 as injuries and 30 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the iVAS 0306330000 FDA approved?

Devices of this type (product code OAR) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Stryker Corporation.

Who makes the iVAS 0306330000, and is it still sold?

The FDA's device database lists Stryker Corporation as the company responsible for it (brand iVAS, model 0306330000). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Injector, vertebroplasty (does not contain cement) devices

All Injector, vertebroplasty (does not contain cement) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page