Skip to content

HOFFMANN 4933-1-001Recalls & reported problems

Stryker GmbH · Wire Bolt - Short LRF For Wire dia 1.5 to 2.0mm · Class 2

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613252507789), received between May 21, 2024 and May 21, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Use of Device Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Use of Device Problem2

Patient problems

  • Failure of Implant2
  • Implant Pain2
  • Nonunion/Delayed-union Bone Fracture2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 21, 2024Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT THERE WERE 3 WIRE BREAKAGES. AFTER CONCLUSION OF THE INVESTIGATION, IT WAS NOTED THAT THERE WAS A CLEAR INDICATION THAT THE WIRE BOLT WASN¿T TIGHTENED ENOUGH.

    Patient problems: Failure of Implant; Implant Pain; Nonunion/Delayed-union Bone Fracture

    MAUDE report 19367325 ↗
  • InjuryMay 21, 2024Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT THERE WERE 3 WIRE BREAKAGES. AFTER CONCLUSION OF THE INVESTIGATION, IT WAS NOTED THAT THERE WAS A CLEAR INDICATION THAT THE WIRE BOLT WASN¿T TIGHTENED ENOUGH.

    Patient problems: Failure of Implant; Implant Pain; Nonunion/Delayed-union Bone Fracture

    MAUDE report 19367322 ↗

Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices

All APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.