HOFFMANN 4933-1-002Recalls & reported problems
Stryker GmbH · Wire Bolt - Medium LRF For Wire dia1.5 to 2.0mm · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 07613252507796), received between Oct 3, 2019 and May 21, 2024 — classified as 5 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Fracture2
- Loosening of Implant Not Related to Bone-Ingrowth2
- Loss of Osseointegration2
- Off-Label Use2
- Use of Device Problem1
Patient problems
- Failure of Implant5
- Injury2
- Insufficient Information2
- No Code Available2
- Implant Pain1
- Nonunion/Delayed-union Bone Fracture1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryMay 21, 2024Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT THERE WERE 3 WIRE BREAKAGES. AFTER CONCLUSION OF THE INVESTIGATION, IT WAS NOTED THAT THERE WAS A CLEAR INDICATION THAT THE WIRE BOLT WASN¿T TIGHTENED ENOUGH.
Patient problems: Failure of Implant; Implant Pain; Nonunion/Delayed-union Bone Fracture
MAUDE report 19367323 ↗MalfunctionAug 4, 2023Break; Fracture; Off-Label Use; Break; Fracture; Off-Label Use
NO EVENT TO REPORT AS OF YET, AWAITING FURTHER INFORMATION. UPDATE 11 JUL 2023:- "WIRE BREAKAGE".
Patient problems: Failure of Implant; Insufficient Information
MAUDE report 17461813 ↗MalfunctionAug 4, 2023Break; Fracture; Off-Label Use; Break; Fracture; Off-Label Use
NO EVENT TO REPORT AS OF YET, AWAITING FURTHER INFORMATION. UPDATE 11 JUL 2023:- "WIRE BREAKAGE".
Patient problems: Failure of Implant; Insufficient Information
MAUDE report 17461568 ↗InjuryFeb 2, 2022Break; Break
THE FOLLOWING EVENT WAS REPORTED BY THE STRYKER REP: "THE HOFFMANN LRF FRAME WAS PUT ON THE PATIENT ON (B)(6) 2021 FOR AN ANKLE FUSION PROCEDURE TO PROTECT THE INTERNAL HARDWARE. PATIENT NOTIFIED THE SURGEON ON (B)(6) 2022 COMPLAINING OF A LOUD 'POP' OR A LOOSE 'CLICKING SOUND. FRAME WAS REMOVED ON THE FOLLOWING DAY (B)(6) 2022. EVERY WIRE BROKE AND NUTS LOOSENED UP. THE SURGEON MENTIONED THE PATIENT WAS NOT WALKING ON IT THROUGHOUT THE FEW WEEKS IT WAS ON."
Patient problems: Failure of Implant
MAUDE report 13424190 ↗InjuryFeb 2, 2022Break; Break
THE FOLLOWING EVENT WAS REPORTED BY THE STRYKER REP: "THE HOFFMANN LRF FRAME WAS PUT ON THE PATIENT ON (B)(6) 2021 FOR AN ANKLE FUSION PROCEDURE TO PROTECT THE INTERNAL HARDWARE. PATIENT NOTIFIED THE SURGEON ON (B)(6) 2022 COMPLAINING OF A LOUD 'POP' OR A LOOSE 'CLICKING SOUND. FRAME WAS REMOVED ON THE FOLLOWING DAY (B)(6) 2022. EVERY WIRE BROKE AND NUTS LOOSENED UP. THE SURGEON MENTIONED THE PATIENT WAS NOT WALKING ON IT THROUGHOUT THE FEW WEEKS IT WAS ON."
Patient problems: Failure of Implant
MAUDE report 13423873 ↗InjuryOct 4, 2019Loss of Osseointegration; Loosening of Implant Not Related to Bone-Ingrowth; Loss of Osseointegration
AS REPORTED: "A COUPLE OF PROXIMAL PINS IN THE TIBIA IN AN LRF CASE WERE LOOSE." ALSO REPORTED: "THE REVISION WAS DONE DUE TO LOOSENING OF THE THIN WIRES. THE WIRES THAT APPEARED LOOSE WERE REPLACED BY HALF PINS, WHICH BECAME STABLE."
Patient problems: Injury; No Code Available
MAUDE report 9156818 ↗InjuryOct 3, 2019Loss of Osseointegration; Loosening of Implant Not Related to Bone-Ingrowth; Loss of Osseointegration
AS REPORTED: "A COUPLE OF PROXIMAL PINS IN THE TIBIA IN AN LRF CASE WERE LOOSE." ALSO REPORTED: "THE REVISION WAS DONE DUE TO LOOSENING OF THE THIN WIRES. THE WIRES THAT APPEARED LOOSE WERE REPLACED BY HALF PINS, WHICH BECAME STABLE."
Patient problems: Injury; No Code Available
MAUDE report 9153695 ↗
Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices
- PRECICE® Plating System LP4.5-DLF31Nuvasive Specialized Orthopedics, Inc. · Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component3 recalls0 reports
- R3 71364451Smith & Nephew, Inc. · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT145 reports
- DFS HA BONE SCREWS HAA60-15040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15050Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAB60-12040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-13040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-18030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page