NA 5450040002Recalls & reported problems
Stryker Corporation · Angled Attachment · Class 2
Quick summary
This Stryker angled attachment for surgical drills appears in 117 malfunction reports, with no deaths and no injuries. Most reports describe the attachment overheating, often during testing, servicing or a procedure, and others describe paint chipping off its color band. Many reports state that no adverse consequences followed.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 117
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 117 adverse event reports that list this exact device (DI 07613252508090), received between Jul 24, 2018 and Mar 3, 2026 — classified as 117 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (88 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Overheating of Device88
- Flaked26
- Detachment of Device or Device Component4
- Fluid/Blood Leak1
Patient problems
- No Clinical Signs, Symptoms or Conditions91
- No Patient Involvement26
- No Known Impact Or Consequence To Patient6
- No Consequences Or Impact To Patient4
- Insufficient Information2
Recent adverse event reports
The 25 most recent of 117 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 3, 2026Flaked
THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE HAD PAINT CHIPS MISSING FROM THE COLOR BAND WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24494543 ↗MalfunctionNov 11, 2025Overheating of Device; Overheating of Device
THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23522370 ↗MalfunctionSep 9, 2025Flaked; Flaked
THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE HAD PAINT CHIPPING FROM THE COLOR BAND WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23003443 ↗MalfunctionAug 4, 2025Overheating of Device; Overheating of Device
THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE DEVICE OVERHEATED DURING A PROCEDURE AT THE USER FACILITY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22706360 ↗MalfunctionApr 23, 2025Overheating of Device; Overheating of Device
THE USER FACILITY REPORTED THAT THE DEVICE OVERHEATED DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21894673 ↗MalfunctionJan 17, 2025Overheating of Device; Overheating of Device
THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE DEVICE OVERHEATED DURING A PROCEDURE AT THE USER FACILITY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21186127 ↗MalfunctionSep 20, 2024Overheating of Device; Overheating of Device
THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE DEVICE OVERHEATED DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20271079 ↗MalfunctionSep 17, 2024Overheating of Device; Overheating of Device
THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE DEVICE OVERHEATED DURING A PROCEDURE AT THE USER FACILITY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20236416 ↗MalfunctionAug 27, 2024Overheating of Device; Overheating of Device
THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20076077 ↗MalfunctionJul 19, 2024Overheating of Device; Overheating of Device
THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE DEVICE OVERHEATED DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19784781 ↗MalfunctionMay 15, 2024Overheating of Device; Overheating of Device
THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE DEVICE OVERHEATED DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19326209 ↗MalfunctionApr 11, 2024Overheating of Device; Overheating of Device
THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE DEVICE OVERHEATED DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19086340 ↗MalfunctionMar 25, 2024Overheating of Device; Overheating of Device
THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18975551 ↗MalfunctionMar 25, 2024Flaked; Flaked
THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE HAD CHIPS MISSING FROM THE COLOR BAND WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18975480 ↗MalfunctionMar 25, 2024Flaked; Flaked
THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE COLOR BAND WAS CHIPPED WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18975364 ↗MalfunctionMar 23, 2024Overheating of Device; Overheating of Device
THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18965547 ↗MalfunctionJan 22, 2024Overheating of Device; Overheating of Device
THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18554731 ↗MalfunctionDec 15, 2023Flaked; Flaked
THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE HAD A CHIPPED COLOR BAND WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18340003 ↗MalfunctionDec 14, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER REPORTED THAT A SMALL PIECE BROKE OFF THE DEVICE PRIOR TO A PROCEDURE AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18329075 ↗MalfunctionJun 19, 2023Overheating of Device; Overheating of Device
THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE DEVICE OVERHEATED DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17154439 ↗MalfunctionJun 15, 2023Overheating of Device; Overheating of Device
THE MANUFACTURER SALES REPRESENTATIVE REPORTED, THAT THE DEVICE OVERHEATED DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17136688 ↗MalfunctionFeb 22, 2023Overheating of Device; Overheating of Device
THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE DEVICE OVERHEATED AT THE USER FACILITY. NO INFORMATION WAS AVAILABLE REGARDING PATIENT INVOLVEMENT, ADVERSE CONSEQUENCES, SURGICAL DELAY OR MEDICAL INTERVENTION.
Patient problems: Insufficient Information
MAUDE report 16419467 ↗MalfunctionFeb 20, 2023Overheating of Device; Overheating of Device
THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE DEVICE OVERHEATED DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16402051 ↗MalfunctionFeb 9, 2023Flaked; Flaked
THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE HAD A CHIPPED PAINT BAND WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16342440 ↗MalfunctionJan 12, 2023Overheating of Device; Overheating of Device
THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE DEVICE OVERHEATED AT THE USER FACILITY. NO INFORMATION WAS AVAILABLE REGARDING PATIENT INVOLVEMENT, ADVERSE CONSEQUENCES, SURGICAL DELAY OR MEDICAL INTERVENTION.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 16146893 ↗
Other DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE devices
- RemB 6400034000Stryker Corporation · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1,618 reports
- NA XMAXAnspach Effort, The · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1,225 reports
- NA EG1ASynthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE859 reports
- NA CRANI-ASynthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1 recall772 reports
- NA 5450400000Stryker Corporation · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE775 reports
- NA 05.001.010Synthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE770 reports
- RemB 6400015000Stryker Corporation · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE603 reports
- NA QD11Synthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE580 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.