Skip to content

HOFFMANN 4933-1-010Recalls & reported problems

Stryker GmbH · Connecting Nut - Short LRF M8 x 6mm · Class 2 · Implant

Adverse event reports
7
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 07613252610304), received between Apr 30, 2018 and Aug 31, 2023 — classified as 5 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Break4
  • Fracture2
  • Loosening of Implant Not Related to Bone-Ingrowth2
  • Off-Label Use2
  • Detachment of Device or Device Component1
  • Device Difficult to Setup or Prepare1
  • Loss of Osseointegration1

Patient problems

  • Failure of Implant4
  • Insufficient Information2
  • No Code Available2
  • Injury1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 31, 2023Break; Fracture; Device Difficult to Setup or Prepare; Off-Label Use; Break; Fracture; Device Difficult to Setup or Prepare; Off-Label Use

    NO EVENT TO REPORT AS OF YET, AWAITING FURTHER INFORMATION. UPDATE:- "WIRE BREAKAGE".

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 17666981 ↗
  • MalfunctionAug 31, 2023Break; Fracture; Off-Label Use; Break; Fracture; Off-Label Use

    NO EVENT TO REPORT AS OF YET, AWAITING FURTHER INFORMATION. UPDATE:- "WIRE BREAKAGE"

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 17666978 ↗
  • InjuryFeb 2, 2022Break; Break

    THE FOLLOWING EVENT WAS REPORTED BY THE STRYKER REP: "THE HOFFMANN LRF FRAME WAS PUT ON THE PATIENT ON (B)(6) 2021 FOR AN ANKLE FUSION PROCEDURE TO PROTECT THE INTERNAL HARDWARE. PATIENT NOTIFIED THE SURGEON ON (B)(6) 2022 COMPLAINING OF A LOUD 'POP' OR A LOOSE 'CLICKING SOUND. FRAME WAS REMOVED ON THE FOLLOWING DAY (B)(6) 2022. EVERY WIRE BROKE AND NUTS LOOSENED UP. THE SURGEON MENTIONED THE PATIENT WAS NOT WALKING ON IT THROUGHOUT THE FEW WEEKS IT WAS ON."

    Patient problems: Failure of Implant

    MAUDE report 13424189 ↗
  • InjuryFeb 1, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    THE FOLLOWING EVENT WAS REPORTED BY THE STRYKER REP: "THE HOFFMANN LRF FRAME WAS PUT ON THE PATIENT ON (B)(6) 2021 FOR AN ANKLE FUSION PROCEDURE TO PROTECT THE INTERNAL HARDWARE. PATIENT NOTIFIED THE SURGEON ON (B)(6) 2022 COMPLAINING OF A LOUD 'POP' OR A LOOSE 'CLICKING SOUND. FRAME WAS REMOVED ON THE FOLLOWING DAY (B)(6) 2022. EVERY WIRE BROKE AND NUTS LOOSENED UP. THE SURGEON MENTIONED THE PATIENT WAS NOT WALKING ON IT THROUGHOUT THE FEW WEEKS IT WAS ON."

    Patient problems: Failure of Implant

    MAUDE report 13419620 ↗
  • InjuryOct 4, 2019Loosening of Implant Not Related to Bone-Ingrowth

    AS REPORTED: "A COUPLE OF PROXIMAL PINS IN THE TIBIA IN AN LRF CASE WERE LOOSE." ALSO REPORTED: "THE REVISION WAS DONE DUE TO LOOSENING OF THE THIN WIRES. THE WIRES THAT APPEARED LOOSE WERE REPLACED BY HALF PINS, WHICH BECAME STABLE."

    Patient problems: No Code Available

    MAUDE report 9157093 ↗
  • InjuryOct 4, 2019Loss of Osseointegration; Loosening of Implant Not Related to Bone-Ingrowth; Loss of Osseointegration

    AS REPORTED: "A COUPLE OF PROXIMAL PINS IN THE TIBIA IN AN LRF CASE WERE LOOSE." ALSO REPORTED: "THE REVISION WAS DONE DUE TO LOOSENING OF THE THIN WIRES. THE WIRES THAT APPEARED LOOSE WERE REPLACED BY HALF PINS, WHICH BECAME STABLE."

    Patient problems: Injury; No Code Available

    MAUDE report 9156719 ↗
  • InjuryApr 30, 2018Break

    AS REPORTED: "THE PATIENT WAS WALKING WITH THE HOFFMANN LRF / HEXAPOD FRAME ON AND THE CLAMP BROKE". REP REPORTED THAT THE CLAMP WAS REPLACED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7472342 ↗

Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices

All APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.