HOFFMANN 4933-1-010Recalls & reported problems

Stryker GmbH · Connecting Nut - Short LRF M8 x 6mm · Class 2 · Implant

All models of HOFFMANN (45)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 07613252610304), received between Apr 30, 2018 and Aug 31, 2023 — classified as 5 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Break4
  • Fracture2
  • Loosening of Implant Not Related to Bone-Ingrowth2
  • Off-Label Use2
  • Detachment of Device or Device Component1
  • Device Difficult to Setup or Prepare1
  • Loss of Osseointegration1

Patient problems

  • Failure of Implant4
  • Insufficient Information2
  • No Code Available2
  • Injury1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 31, 2023Break; Fracture; Device Difficult to Setup or Prepare; Off-Label Use; Break; Fracture; Device Difficult to Setup or Prepare; Off-Label Use

    NO EVENT TO REPORT AS OF YET, AWAITING FURTHER INFORMATION. UPDATE:- "WIRE BREAKAGE".

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 17666981 ↗
  • MalfunctionAug 31, 2023Break; Fracture; Off-Label Use; Break; Fracture; Off-Label Use

    NO EVENT TO REPORT AS OF YET, AWAITING FURTHER INFORMATION. UPDATE:- "WIRE BREAKAGE"

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 17666978 ↗
  • InjuryFeb 2, 2022Break; Break

    THE FOLLOWING EVENT WAS REPORTED BY THE STRYKER REP: "THE HOFFMANN LRF FRAME WAS PUT ON THE PATIENT ON (B)(6) 2021 FOR AN ANKLE FUSION PROCEDURE TO PROTECT THE INTERNAL HARDWARE. PATIENT NOTIFIED THE SURGEON ON (B)(6) 2022 COMPLAINING OF A LOUD 'POP' OR A LOOSE 'CLICKING SOUND. FRAME WAS REMOVED ON THE FOLLOWING DAY (B)(6) 2022. EVERY WIRE BROKE AND NUTS LOOSENED UP. THE SURGEON MENTIONED THE PATIENT WAS NOT WALKING ON IT THROUGHOUT THE FEW WEEKS IT WAS ON."

    Patient problems: Failure of Implant

    MAUDE report 13424189 ↗
  • InjuryFeb 1, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    THE FOLLOWING EVENT WAS REPORTED BY THE STRYKER REP: "THE HOFFMANN LRF FRAME WAS PUT ON THE PATIENT ON (B)(6) 2021 FOR AN ANKLE FUSION PROCEDURE TO PROTECT THE INTERNAL HARDWARE. PATIENT NOTIFIED THE SURGEON ON (B)(6) 2022 COMPLAINING OF A LOUD 'POP' OR A LOOSE 'CLICKING SOUND. FRAME WAS REMOVED ON THE FOLLOWING DAY (B)(6) 2022. EVERY WIRE BROKE AND NUTS LOOSENED UP. THE SURGEON MENTIONED THE PATIENT WAS NOT WALKING ON IT THROUGHOUT THE FEW WEEKS IT WAS ON."

    Patient problems: Failure of Implant

    MAUDE report 13419620 ↗
  • InjuryOct 4, 2019Loosening of Implant Not Related to Bone-Ingrowth

    AS REPORTED: "A COUPLE OF PROXIMAL PINS IN THE TIBIA IN AN LRF CASE WERE LOOSE." ALSO REPORTED: "THE REVISION WAS DONE DUE TO LOOSENING OF THE THIN WIRES. THE WIRES THAT APPEARED LOOSE WERE REPLACED BY HALF PINS, WHICH BECAME STABLE."

    Patient problems: No Code Available

    MAUDE report 9157093 ↗
  • InjuryOct 4, 2019Loss of Osseointegration; Loosening of Implant Not Related to Bone-Ingrowth; Loss of Osseointegration

    AS REPORTED: "A COUPLE OF PROXIMAL PINS IN THE TIBIA IN AN LRF CASE WERE LOOSE." ALSO REPORTED: "THE REVISION WAS DONE DUE TO LOOSENING OF THE THIN WIRES. THE WIRES THAT APPEARED LOOSE WERE REPLACED BY HALF PINS, WHICH BECAME STABLE."

    Patient problems: Injury; No Code Available

    MAUDE report 9156719 ↗
  • InjuryApr 30, 2018Break

    AS REPORTED: "THE PATIENT WAS WALKING WITH THE HOFFMANN LRF / HEXAPOD FRAME ON AND THE CLAMP BROKE". REP REPORTED THAT THE CLAMP WAS REPLACED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7472342 ↗

FDA clearance and classification

The FDA regulates product code KTT (appliance, fixation, nail/blade/plate combination, multiple component) under 21 CFR 888.3030 as a Class II device. 21 CFR 888.3030 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the HOFFMANN 4933-1-010

Has the HOFFMANN 4933-1-010 been recalled?

No FDA recall record in our data names this exact device identifier (DI 07613252610304), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the HOFFMANN 4933-1-010?

The FDA's MAUDE database lists 7 adverse event reports for this device, received between Apr 30, 2018 and Aug 31, 2023. In those reports, the most common specific device problems are “Break” (4), “Fracture” (2) and “Loosening of Implant Not Related to Bone-Ingrowth” (2); the most common patient problems are “Failure of Implant” (4) and “Injury” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the HOFFMANN 4933-1-010?

None of the reports are classified as deaths, 5 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the HOFFMANN 4933-1-010 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the HOFFMANN 4933-1-010, and is it still sold?

The FDA's device database lists Stryker GmbH as the company responsible for it (brand HOFFMANN, model 4933-1-010). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices

All APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page