Skip to content

Excelsior XT-17 M003C1775ST0Recalls & reported problems

Stryker Corporation · 2 Tip Straight Microcatheter · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 07613252714866), received between Mar 20, 2019 and Jun 4, 2026 — classified as 1 injury, 10 malfunctions.

The most frequently reported specific device problem is “Fracture” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture4
  • Material Puncture/Hole3
  • Peeled/Delaminated3
  • Adverse Event Without Identified Device or Use Problem1
  • Flaked1
  • Material Separation1
  • Material Split, Cut or Torn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • No Consequences Or Impact To Patient2
  • Thrombosis/Thrombus1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 4, 2026Material Puncture/Hole

    THE MICROCATHETER (SUBJECT DEVICE) WAS RETURNED WITH THE ATLAS STENT DEVICE AND THE VISUAL/MICROSCOPIC INSPECTION REVEALED THAT THE SDW (STENT DELIVERY WIRE) OF THE ATLAS STENT HAD PIERCED THROUGH THE LUMEN/SHAFT OF THE SUBJECT MICROCATHETER. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25398247 ↗
  • MalfunctionSep 3, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING AN ANEURYSM-ASSISTED EMBOLIZATION PROCEDURE, AFTER THE SUBJECT MICROCATHETER WAS PROPERLY POSITIONED AND THE ATLAS STENT WAS TRANSPORTED INTO THE SYSTEM, BUT IT WAS FOUND THAT THE STENT COULD NOT BE PUSHED FORWARD. SUBSEQUENTLY, WHEN THE SYSTEM WAS WITHDRAWN, IT WAS DISCOVERED THAT THE SUBJECT MICROCATHETER HAD FRACTURED INSIDE THE INTERMEDIATE CATHETER. THE MICROCATHETER WAS REPLACED AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO CLINICAL CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22956925 ↗
  • MalfunctionMar 5, 2025Material Puncture/Hole; Material Puncture/Hole

    ANALYSIS OF THE RETURNED DEVICE REVEALED THAT SUBJECT CATHETER (SUBJECT DEVICE) SHAFT HAD A HOLE/PERFORATION. THERE WAS NO CLINICAL CONSEQUENCE TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21529702 ↗
  • MalfunctionNov 5, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING PCOMA POSTERIOR COMMUNICATING ARTERY ANEURYSM STENT ASSISTED COIL EMBOLIZATION PROCEDURE, WHEN THE STENT REACHED DISTAL OF THE SUBJECT MICROCATHETER IT GOT STUCK AND COULD NOT ADVANCE ANY MORE. THE OPERATOR TRIED FOR SEVERAL TIMES AND THEN FOUND THE TIP OF THE STENT MARKER WAS EXPANDED IN THE VESSEL. HOWEVER, THE STENT WAS NOT COMING OUT FROM TIP. SO, THE OPERATOR FELT THE DISTAL OF SUBJECT MICROCATHETER MIGHT BE FRACTURED, SO HE WITHDREW THE STENT AND SUBJECT MICROCATHETER TOGETHER THROUGH MIDDLE CATHETER. THE SUBJECT DEVICE WAS REPLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20606408 ↗
  • MalfunctionApr 4, 2023Material Split, Cut or Torn; Material Split, Cut or Torn

    THE SUBJECT DEVICE WAS RETURNED FOR ANALYSIS AND THE INVESTIGATION OF THE DEVICE REVEALED THAT THE SUBJECT CATHETER SHAFT HAS HOLES/PERFORATION. THE PHYSICIAN REPLACED IT WITH A NEW DEVICE AND CONTINUED THE PROCEDURE WITHOUT CLINICAL CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16676272 ↗
  • MalfunctionNov 1, 2022Fracture; Peeled/Delaminated; Material Puncture/Hole; Fracture; Peeled/Delaminated; Material Puncture/Hole

    THE CATHETER (SUBJECT DEVICE) WAS RETURNED FOR ANALYSIS AND IT WAS DISCOVERED THAT THE SHAFT WAS DAMAGED AND HAD A HOLE IN THE PROXIMAL SHAFT. PTFE INNER LINER AND CATHETER (SUBJECT DEVICE) FRAGMENTS WERE ALSO PRESENT WHEN THE CATHETER WAS FLUSHED. NO CLINICAL CONSEQUENCES WERE REPORTED DUE TO THIS EVENT. NO OTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15710129 ↗
  • InjuryJan 28, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED DURING A CLINICAL STUDY THAT POST PROCEDURE OF TREATMENT FOR A STENT THROMBOSIS INVOLVING THE CATHETER (SUBJECT DEVICE) ON (B)(6) 2017, A SMALL CLOT WAS VISIBLE DURING LAST CONTROL OF ANGIOGRAM ON THE INFERIOR BRANCH OF THE BIFURCATION. IT LASTED LESS THAN 24 HOURS AND WAS RESOLVED WITHOUT SEQUELAE. THE PATIENTS MEDICAL CONDITION WAS GOOD. THERE WERE NO OTHER CLINICAL CONSEQUENCES TO THE PATIENT.

    Patient problems: Thrombosis/Thrombus

    MAUDE report 11246185 ↗
  • MalfunctionJan 23, 2021Fracture; Fracture

    IT WAS REPORTED THAT THE DURING THE PROCEDURE, RESISTANCE WAS FELT WHEN GUIDING THE SUBJECT CATHETER TO THE MIDDLE CEREBRAL ARTERY WITH THE GUIDEWIRE USED. PHYSICIAN THEN REPLACED THE GUIDEWIRE WITH THE SAME PRODUCT AND THE PROCEDURE WAS CONTINUED. A STENT WAS THEN INSERTED INTO THE SUBJECT CATHETER AND PHYSICIAN NOTED THERE WAS RESISTANCE. PHYSICIAN FELT A STRONG RESISTANCE IN FRONT OF THE TARGET SITE WHEN ATTEMPTING TO PLACE STENT. AS PHYSICIAN PUSHED THE STENT FORWARD, THE TIP OF THE STENT WAS PROTRUDING TO THE OUTSIDE OF THE SUBJECT CATHETER IN FRONT OF THE SECOND MARKER. PHYSICIAN PROCEEDED TO REMOVE THE SUBJECT CATHETER AND STENT FROM PATIENT. BOTH DEVICES WERE REPLACED WITH THE SAME PRODUCTS AND THE PROCEDURE CONTINUED AND WAS COMPLETED SUCCESSFULLY WITH NO CLINICAL CONSEQUENCES TO THE PATIENT. ANALYSIS OF THE RETURNED SUBJECT CATHETER REVEALED THAT THE DEVICE HAD FRACTURED DURING USE. THEREFORE, BASED ON THIS INFORMATION THE EVENT IS DEEMED REPORTABLE AND EMDR WILL BE FILED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11220341 ↗
  • MalfunctionSep 30, 2020Flaked; Peeled/Delaminated; Flaked; Peeled/Delaminated

    IT WAS REPORTED THAT THE COATING OF THE DISTAL SHAFT OF THE MICROCATHETER (SUBJECT DEVICE) WAS PEELING AWAY WHEN REMOVED FROM THE DISPENSER HOOP. THE PHYSICIAN REPLACED IT WITH A NEW DEVICE AND CONTINUED THE PROCEDURE WITHOUT CLINICAL CONSEQUENCES TO THE PATIENT. UPDATE: THE PRODUCT INVESTIGATION CONFIRMED THAT THE HYDROPHILIC PEELING ISSUE OF THE SUBJECT CATHETER WAS NOT CONFIRMED BASED ON ANALYSIS. THERE WERE SMALL FLAKES OF COATING IDENTIFIED ON THE CATHETER PROXIMAL SHAFT WHICH MAY HAVE BEEN PERCEIVED AS COATING PEELING BY THE PHYSICIAN. IT IS HIGHLY UNLIKELY THAT THE SMALL FLAKES OF COATING MAY CONTRIBUTED TO AND/OR CAUSE ANY PATIENT INJURY; THEREFORE, THIS RECORD HAS BEEN DEEMED NON-REPORTABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10610724 ↗
  • MalfunctionAug 13, 2019Peeled/Delaminated; Peeled/Delaminated

    ANALYSIS OF THE RETURNED DEVICE FOUND THAT THE CATHETER PTFE (POLYTETRAFLUOROETHYLENE) INNER LINING PEELING. THERE WERE NO CLINICAL CONSEQUENCES TO THE PATIENT REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8889028 ↗
  • MalfunctionMar 20, 2019Material Separation; Material Separation

    IT WAS REPORTED THAT THE MARKER OF THE MICROCATHETER APPEARED FRACTURED UNDER FLUOROSCOPY THE PHYSICIAN REMOVED THE MARKER WITH THE MICROCATHETER. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8438539 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page