Skip to content

XIA 3 48233550Recalls & reported problems

Stryker Corporation · UNIPLANAR SCREW · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613327001051), received between Jul 20, 2016 and Jul 20, 2016 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment Of Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Detachment Of Device Component2
  • Mechanical Problem2

Patient problems

  • No Code Available2
  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 20, 2016Detachment Of Device Component; Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT; DURING THE DVR PROCEDURE, TUBES WERE USED WITHOUT CLAM SHELLS. STANDARD BLOCKER INSTER WAS USED INSTEAD OF TORQUE WRENCH. AFTER PROCEDURE WAS COMPLETED IT WAS NOTICED IN XRAY THE 2 TULIP HEADS HAD POPPED OFF FROM THE SCREW POST. SCREWS WERE REPLACED NO FURTHER ISSUES. UPDATE: PER ADDITIONAL CORRESPONDENCE WITH SALES REP, THE CAVITY WAS OPEN AND THERE WAS NO NEED TO REOPEN THE PATIENT AT THE TIME THE REPORTED EVENT WAS NOTICED.

    Patient problems: No Code Available; No Known Impact Or Consequence To Patient

    MAUDE report 5808733 ↗
  • MalfunctionJul 20, 2016Detachment Of Device Component; Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT; DURING THE DVR PROCEDURE, TUBES WERE USED WITHOUT CLAM SHELLS. STANDARD BLOCKER INSTER WAS USED INSTEAD OF TORQUE WRENCH. AFTER PROCEDURE WAS COMPLETED IT WAS NOTICED IN XRAY THE 2 TULIP HEADS HAD POPPED OFF FROM THE SCREW POST. SCREWS WERE REPLACED NO FURTHER ISSUES. UPDATE: PER ADDITIONAL CORRESPONDENCE WITH SALES REP, THE CAVITY WAS OPEN AND THERE WAS NO NEED TO REOPEN THE PATIENT AT THE TIME THE REPORTED EVENT WAS NOTICED.

    Patient problems: No Code Available; No Known Impact Or Consequence To Patient

    MAUDE report 5808596 ↗

Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices

All APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page