Skip to content

TRANSFORM M003SRC03100Recalls & reported problems

Stryker Corporation · C Compliant Occlusion Balloon Catheter · Class 2

All models of TRANSFORM (13)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07613327004410), received between Aug 8, 2018 and Feb 8, 2021 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).

8 other recalls cover devices cleared under the same FDA submission (K122576); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Fluid/Blood Leak2
  • Component Missing1
  • Deflation Problem1
  • Material Rupture1

Patient problems

  • No Consequences Or Impact To Patient2
  • No Patient Involvement2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 8, 2021Component Missing; Component Missing

    IT WAS REPORTED THAT DURING THE PROCEDURE, THE BALLOON (SUBJECT DEVICE) WAS INFLATED HOWEVER, THE RADIOPAQUE CONTRAST INSIDE THE INFLATED BALLOON WAS NOT VISIBLE ON SCREEN. THE INFLATION WAS CONFIRMED BY THE PHYSICIAN BY MOVEMENT OF THE INTRAOPERATIVE ADJACENT COILING ACTIVITY AND RADIOPAQUE CONTRAST WAS STILL NOT VISIBLE. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11291306 ↗
  • MalfunctionOct 19, 2018Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT DURING PREPARATION ON THE BACK TABLE, THE SUBJECT DEVICE BALLOON PAIRED WITH A FLOPPY WIRE AND DURING PREP TEST INFLATION WITH A WIRE, THERE WAS LEAKAGE AT THE DISTAL END. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT DUE TO THE REPORTED EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 7987203 ↗
  • MalfunctionOct 19, 2018Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT DURING PREPARATION ON THE BACK TABLE, THE SUBJECT DEVICE BALLOON PAIRED WITH A FLOPPY WIRE AND DURING PREP TEST INFLATION WITH A WIRE, THERE WAS LEAKAGE AT THE DISTAL END. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT DUE TO THE REPORTED EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 7987200 ↗
  • MalfunctionAug 31, 2018Deflation Problem; Deflation Problem

    IT WAS REPORTED THAT THE BALLOON CATHETER FAILED TO DEFLATE COMPLETELY DURING THE PROCEDURE. THERE WERE NO REPORTED CLINICAL CONSEQUENCES TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7838523 ↗
  • MalfunctionAug 8, 2018Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING PROCEDURE, THE FIRST BALLOON CATHETER (SUBJECT DEVICE) WAS RUPTURED. THE FIRST BALLOON WAS REMOVED FROM THE PATIENT'S BODY AND REPLACE WITH A 2ND BALLOON CATHETER. DURING USE OF THE 2ND BALLOON CATHETER, EXTRAVASATION WAS OBSERVED. NO ADDITIONAL INFORMATION IS AVAILABLE CURRENTLY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7765553 ↗

FDA clearance and classification

Cleared through 510(k) K122576 on Jan 11, 2013 (decision: Substantially Equivalent). Submitted by Stryker Neurovascular as “Transform Occlusion Balloon Catheter (Compliant And Super Compliant”. The FDA review took 141 days. FDA record ↗

The FDA regulates product code MJN (catheter, intravascular occluding, temporary) under 21 CFR 870.4450 as a Class II device. 21 CFR 870.4450 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K122576. One submission often covers many models, so check the product and lot details in each record.

  • Z-0506-2014 · Nov 4, 2013 · Terminated
    Stryker brand COMPLIANT 4MM X 15MM, Catalog Number: M003SRC04150; Product is manufactured by Stryker Neurovascular, Cork, Ireland and distributed by Excelsior SL-10/Excelsior 1018
  • Z-0509-2014 · Nov 4, 2013 · Terminated
    Stryker brand COMPLIANT 5MM X 15MM, Catalog Number: M003SRC05150; Product is manufactured by Stryker Neurovascular, Cork, Ireland and distributed by Stryker Neurovascular, Fremont
  • Z-0508-2014 · Nov 4, 2013 · Terminated
    Stryker brand COMPLIANT 4MM X 30MM, Catalog Number: M003SRC04300; Product is manufactured by Stryker Neurovascular, Cork, Ireland and distributed by Stryker Neurovascular, Fremont,
  • Z-0505-2014 · Nov 4, 2013 · Terminated
    Stryker brand COMPLIANT 4MM X 10MM, Catalog Number: M003SRC04100; Product is manufactured by Stryker Neurovascular, Cork, Ireland and distributed by Stryker Neurovascular, Fremont,
  • Z-0510-2014 · Nov 4, 2013 · Terminated
    Stryker brand COMPLIANT 5MM X 30MM, Catalog Number: M003SRC05300; Product is manufactured by Stryker Neurovascular, Cork, Ireland and distributed by Stryker Neurovascular, Fremont
  • Z-0507-2014 · Nov 4, 2013 · Terminated
    Stryker brand COMPLIANT 4MM X 20MM, Catalog Number: M003SRC04200; Product is manufactured by Stryker Neurovascular, Cork, Ireland and distributed by Stryker Neurovascular, Fremont,
  • Z-0512-2014 · Nov 4, 2013 · Terminated
    Stryker brand SUPER COMPLIANT 7MM X 15MM, Catalog Number: M003SSC07150; Product is manufactured by Stryker Neurovascular, Cork, Ireland and distributed by Stryker Neurovascular, F
  • Z-0511-2014 · Nov 4, 2013 · Terminated
    Stryker brand SUPER COMPLIANT 4MM X 7MM, Catalog Number: M003SSC04070; Product is manufactured by Stryker Neurovascular, Cork, Ireland and distributed by Stryker Neurovascular, Fr

Common questions about the TRANSFORM M003SRC03100

Has the TRANSFORM M003SRC03100 been recalled?

No FDA recall record in our data names this exact device identifier (DI 07613327004410), based on the FDA recall file dated Nov 27, 2026. However, 8 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the TRANSFORM M003SRC03100?

The FDA's MAUDE database lists 5 adverse event reports for this device, received between Aug 8, 2018 and Feb 8, 2021. In those reports, the most common specific device problems are “Fluid/Blood Leak” (2), “Component Missing” (1) and “Deflation Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the TRANSFORM M003SRC03100?

None of the reports are classified as deaths, 0 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the TRANSFORM M003SRC03100 FDA approved?

It was cleared through 510(k) K122576 on Jan 11, 2013. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the TRANSFORM M003SRC03100, and is it still sold?

The FDA's device database lists Stryker Corporation as the company responsible for it (brand TRANSFORM, model M003SRC03100). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter, intravascular occluding, temporary devices

All Catheter, intravascular occluding, temporary devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page