TRANSFORM M003SRC05150Recalls & reported problems
Stryker Corporation · C Compliant Occlusion Balloon Catheter · Class 2
All models of TRANSFORM (13)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 07613327004472), received between Nov 7, 2018 and Apr 3, 2024 — classified as 2 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Deflation Problem” (3 reports).
8 other recalls cover devices cleared under the same FDA submission (K122576); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Deflation Problem3
- Break1
- Fracture1
- Leak/Splash1
- Material Puncture/Hole1
- Material Rupture1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Foreign Body In Patient1
- No Consequences Or Impact To Patient1
- Thrombus1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 3, 2024Material Puncture/Hole; Material Puncture/Hole
THE BALLOON (SUBJECT DEVICE) WAS RETURNED FOR ANALYSIS AND IT WAS DISCOVERED THAT THERE WAS A TEAR/HOLE/PERFORATION NOTED ON THE BALLOON (SUBJECT DEVICE). THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19035917 ↗MalfunctionJun 19, 2023Deflation Problem; Deflation Problem
IT WAS REPORTED THAT DURING THE PROCEDURE, THE BALLOON (SUBJECT DEVICE) WAS INFLATED. WHEN THE BALLOON WAS ATTEMPTED TO DEFLATE IT WAS UNABLE TO DEFLATE. THE PHYSICIAN REPLACED IT WITH A NEW DEVICE AND CONTINUED THE PROCEDURE WITHOUT CLINICAL CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17154423 ↗InjuryMay 15, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING THE ADDITIONAL EMBOLIZATION PROCEDURE OF A 18 MONTHS PREVIOUSLY TREATED AREA, THE SHAFT OF THE SUBJECT BALLOON CATHETER WAS KINKED AND IT WAS PERFORATED BY A GUIDEWIRE. THE SHAFT OF THE SUBJECT BALLOON CATHETER WAS FRACTURED DURING RETRACTION. ATTEMPTS WERE MADE TO RETRIEVE THE BROKEN FRAGMENT OF THE SUBJECT BALLOON CATHETER USING THE SNARE DEVICE BUT IT WAS DIFFICULT TO BE REMOVED. THE BROKEN FRAGMENT OF THE SUBJECT BALLOON CATHETER WAS LEFT IN THE COMMON HEPATIC ARTERY. IN DOCTOR'S OPINION THERE WAS NO RISK OF COMPLICATIONS AND WOULD NOT IMPACT THE PATIENT. HE BELIEVES THAT THE SUBJECT BALLOON CATHETER SHAFT MAY HAVE BEEN OVERLOADED DUE TO THE MEANDERING ARTERY WITH HIGH DEGREE OF ARTERIOSCLEROSIS. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THE PATIENT'S CONDITION IS STABLE.
Patient problems: Foreign Body In Patient
MAUDE report 16930997 ↗MalfunctionMar 15, 2023Deflation Problem; Deflation Problem
IT WAS REPORTED THAT DURING COIL EMBOLIZATION OF CELIAC ANEURYSM THE OCCLUSION BALLOON CATHETER (SUBJECT DEVICE) WAS ADVANCED TO THE TARGET SITE AFTER SETUP. WHEN INFLATION AND DEFLATION WERE PERFORMED, IT WAS NOTED THAT BOTH MOTIONS WERE SLOW. THE PHYSICIAN REPLACED IT WITH A NEW DEVICE AND CONTINUED THE PROCEDURE WITHOUT CLINICAL CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16546197 ↗MalfunctionFeb 20, 2023Leak/Splash; Material Rupture; Leak/Splash; Material Rupture
IT WAS REPORTED THAT DURING EMBOLIZATION OF RENAL ARTERY ANEURYSM THE BALLOON (SUBJECT DEVICE) WAS ADVANCED INTO THE PATIENT¿S ANATOMY AND AN ATTEMPT TO INFLATE IT WAS MADE. HOWEVER, THE BALLOON (SUBJECT DEVICE) DID NOT INFLATE. WHEN RETRIEVED FROM THE PATIENT'S ANATOMY THE BALLOON (SUBJECT DEVICE) WAS FOUND TO BE RUPTURED AND CONTRAST AGENT WAS LEAKING OUT. THE PHYSICIAN REPLACED IT WITH A NEW DEVICE AND CONTINUED THE PROCEDURE WITHOUT CLINICAL CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16406384 ↗MalfunctionApr 12, 2019Break; Break
IT WAS REPORTED THAT DURING PROCEDURE, THE SURGEON FELT RESISTANCE WITH THE SUBJECT BALLOON CATHETER AT THE TIME OF INSERTION OF THE MICROCATHETER. THEN THE BALLOON ¿BROKE¿. THE SUBJECT DEVICE WAS EXCHANGED WITH THE SAME NEW PRODUCT. THE SAME ISSUE OCCURRED WITH A SECOND BALLOON CATHETER. THE SECOND BALLOON CATHETER WAS EXCHANGED WITH THE SAME NEW PRODUCT. THE PROCEDURE CONTINUED AND WAS COMPLETED WITHOUT PROBLEM. THERE WERE NO CLINICAL CONSEQUENCES TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8511941 ↗InjuryNov 7, 2018Deflation Problem; Deflation Problem
IT WAS REPORTED THAT DURING THE PROCEDURE, THE BALLOON CATHETER WAS UNABLE TO DEFLATE, DUE TO THROMBUS FORMATION INSIDE THE BALLOON LUMEN. THE PHYSICIAN WAS ABLE TO GET RID OF THE THROMBUS BY MANIPULATING THE GUIDEWIRE AND THE BALLOON WAS ABLE TO DEFLATE. THE BALLOON WAS CONTINUED TO BE USED; HOWEVER, SIMILAR ISSUE OCCURRED. THE BALLOON WAS REMOVED FROM THE PATIENT BODY. NO ADDITIONAL INFORMATION IS AVAILABLE CURRENTLY.
Patient problems: Thrombus
MAUDE report 8049080 ↗
FDA clearance and classification
Cleared through 510(k) K122576 on Jan 11, 2013 (decision: Substantially Equivalent). Submitted by Stryker Neurovascular as “Transform Occlusion Balloon Catheter (Compliant And Super Compliant”. The FDA review took 141 days. FDA record ↗
The FDA regulates product code MJN (catheter, intravascular occluding, temporary) under 21 CFR 870.4450 as a Class II device. 21 CFR 870.4450 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K122576. One submission often covers many models, so check the product and lot details in each record.
- Z-0508-2014 · Nov 4, 2013 · TerminatedStryker brand COMPLIANT 4MM X 30MM, Catalog Number: M003SRC04300; Product is manufactured by Stryker Neurovascular, Cork, Ireland and distributed by Stryker Neurovascular, Fremont,
- Z-0506-2014 · Nov 4, 2013 · TerminatedStryker brand COMPLIANT 4MM X 15MM, Catalog Number: M003SRC04150; Product is manufactured by Stryker Neurovascular, Cork, Ireland and distributed by Excelsior SL-10/Excelsior 1018
- Z-0510-2014 · Nov 4, 2013 · TerminatedStryker brand COMPLIANT 5MM X 30MM, Catalog Number: M003SRC05300; Product is manufactured by Stryker Neurovascular, Cork, Ireland and distributed by Stryker Neurovascular, Fremont
- Z-0507-2014 · Nov 4, 2013 · TerminatedStryker brand COMPLIANT 4MM X 20MM, Catalog Number: M003SRC04200; Product is manufactured by Stryker Neurovascular, Cork, Ireland and distributed by Stryker Neurovascular, Fremont,
- Z-0511-2014 · Nov 4, 2013 · TerminatedStryker brand SUPER COMPLIANT 4MM X 7MM, Catalog Number: M003SSC04070; Product is manufactured by Stryker Neurovascular, Cork, Ireland and distributed by Stryker Neurovascular, Fr
- Z-0509-2014 · Nov 4, 2013 · TerminatedStryker brand COMPLIANT 5MM X 15MM, Catalog Number: M003SRC05150; Product is manufactured by Stryker Neurovascular, Cork, Ireland and distributed by Stryker Neurovascular, Fremont
- Z-0505-2014 · Nov 4, 2013 · TerminatedStryker brand COMPLIANT 4MM X 10MM, Catalog Number: M003SRC04100; Product is manufactured by Stryker Neurovascular, Cork, Ireland and distributed by Stryker Neurovascular, Fremont,
- Z-0512-2014 · Nov 4, 2013 · TerminatedStryker brand SUPER COMPLIANT 7MM X 15MM, Catalog Number: M003SSC07150; Product is manufactured by Stryker Neurovascular, Cork, Ireland and distributed by Stryker Neurovascular, F
Common questions about the TRANSFORM M003SRC05150
Has the TRANSFORM M003SRC05150 been recalled?
No FDA recall record in our data names this exact device identifier (DI 07613327004472), based on the FDA recall file dated Nov 27, 2026. However, 8 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the TRANSFORM M003SRC05150?
The FDA's MAUDE database lists 7 adverse event reports for this device, received between Nov 7, 2018 and Apr 3, 2024. In those reports, the most common specific device problems are “Deflation Problem” (3), “Break” (1) and “Fracture” (1); the most common patient problems are “Foreign Body In Patient” (1) and “Thrombus” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the TRANSFORM M003SRC05150?
None of the reports are classified as deaths, 2 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the TRANSFORM M003SRC05150 FDA approved?
It was cleared through 510(k) K122576 on Jan 11, 2013. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the TRANSFORM M003SRC05150, and is it still sold?
The FDA's device database lists Stryker Corporation as the company responsible for it (brand TRANSFORM, model M003SRC05150). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter, intravascular occluding, temporary devices
- Pruitt F3 Carotid Shunt 2012-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary1 death158 reports
- GORE Molding Occlusion Balloon Catheter MOB37W. L. Gore & Associates, Inc. · Catheter, Intravascular Occluding, Temporary6 deaths152 reports
- Bridge Occlusion Balloon 590-001Spectranetics Corporation · Catheter, Intravascular Occluding, Temporary4 deaths57 reports
- Pruitt F3 Carotid Shunt e2012-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary53 reports
- Pruitt F3 Carotid Shunt 2013-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary46 reports
- Equalizer™ M001171100Boston Scientific Corporation · Catheter, intravascular occluding, temporary18 reports
- Pruitt F3 Carotid Shunt e2013-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary17 reports
- TRANSFORM M003SSC07070Stryker Corporation · Catheter, intravascular occluding, temporary17 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page