Skip to content

PCA Head 2284-0-428Recalls & reported problems

Howmedica Osteonics Corp. · Femoral Head Component · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613327012774), received between Nov 11, 2020 and Nov 11, 2020 — classified as 1 injury.

The most frequently reported specific device problem is “Corroded” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Corroded1
  • Device Dislodged or Dislocated1

Patient problems

  • Inflammation1
  • Joint Dislocation1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 11, 2020Corroded; Device Dislodged or Dislocated; Corroded; Device Dislodged or Dislocated

    REVISION SURGERY. LAST REVISION WAS COMPLETED ON (B)(6)2011. UPON SURGERY, IT WAS NOTICED THAT THE FEMORAL HEAD DISASSOCIATED WITH THE OUTER HEAD COMPONENT. SEVERE METALLOSIS WAS NOTED. LEFT SIDE. UPDATE: ORIGINAL REVISION DATE WAS 11/21/11. NO OP-NOTES OR ANY OTHER INFORMATION FROM THE PHYSICIAN. THE LAST REVISION WHEN THE IMPLANTS WERE RETRIEVED ((B)(6)2020), IT WAS CLEAR THE INNER PCA HEAD WAS DISASSOCIATED WITH THE OUTER MDM HEAD (AS EVIDENCED BY THE X RAY). THE PCA HEAD WAS ARTICULATING ON THE MDM METAL LINER INSIDE THE CUP. THE TISSUE WAS GRAY AS WAS FLUID INSIDE THE JOINT WHICH WAS SUCTIONED OUT.

    Patient problems: Inflammation; Joint Dislocation

    MAUDE report 10821470 ↗

Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices

All PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices →