V40 Head 6260-5-128Recalls & reported problems
Howmedica Osteonics Corp. · V40 Femoral Head · Class 2 · Implant
All models of V40 Head (30)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07613327012866), received between Jan 24, 2017 and Aug 31, 2022 — classified as 4 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Unstable” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Unstable2
- Adverse Event Without Identified Device or Use Problem1
- Degraded1
- Device Damaged Prior to Use1
- Device Dislodged or Dislocated1
- Insufficient Device Problem Information1
- Loose or Intermittent Connection1
Patient problems
- Joint Laxity2
- No Consequences Or Impact To Patient2
- Foreign Body Reaction1
- Inflammation1
- Injury1
- Insufficient Information1
- Joint Dislocation1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryAug 31, 2022Unstable; Device Dislodged or Dislocated; Unstable; Device Dislodged or Dislocated
AS REPORTED: "TOTAL HIP ARTHROPLASTY REVISION - LEFT. DIAGNOSIS: UNSTABLE HEMI-ARTHROPLASTY. NO FURTHER INFORMATION AVAILABLE PER HOSPITAL." A BIPOLAR COMPONENT AD FEMORAL HEAD WERE REVISED TO A TOTAL HIP.
Patient problems: Joint Dislocation; Joint Laxity
MAUDE report 15325747 ↗InjuryFeb 17, 2022Degraded; Degraded
THE FOLLOWING WAS REPORTED: COULD BE ARMD(ADVERSE REACTION TO METALLIC DEBRIS) DUE TO TRUNNIONOSIS. REVISION.
Patient problems: Foreign Body Reaction; Insufficient Information
MAUDE report 13552670 ↗InjuryNov 7, 2021Unstable; Unstable
IT WAS REPORTED THAT THE PATIENT'S HIP WAS REVISED DUE TO ECCENTRIC WEAR OF THE POLY LINER AND INSTABILITY. A HEAD AND LINER EXCHANGE WAS PERFORMED. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Joint Laxity
MAUDE report 12757731 ↗MalfunctionJul 30, 2019Device Damaged Prior to Use; Device Damaged Prior to Use
DURING THE SURGERY THE DOCTOR NOTICED THAT TWO ITEMS WERE DEFECTIVE. ADDITIONAL INFORMATION: DURING THE SURGERY, THE OPERATOR STATED THAT THE HEAD THAT WAS SLAMMED IN THE HEAD OF UH1 WEAKLY MOVES AND WILL CAUSE DISLOCATION IN THE JOINT, HE DECIDED TO CHANGE THE HEAD AND HEAD OF UH1.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8842168 ↗InjuryJan 14, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
A REPRESENTATIVE FROM THE CLINIC REPORTED THE FOLLOWING EVENT : "INSTALLATION OF AN ABG IL HIP PROSTHESIS: ALLERGY TO METALS".
Patient problems: Inflammation; Injury; Reaction
MAUDE report 8245227 ↗MalfunctionJan 24, 2017Loose or Intermittent Connection; Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
WHEN SURGEON WAS PERFORMING THE HIP-JOINT REPLACEMENT SURGERY, HE FOUND THAT AFTER HE CONNECTED THE ACCOLADE HFX HANDLE TO 28MM STD V40 TAPER VIT HEAD, THEY WERE LOOSENING AND COULD NOT BE FASTENED. THE SURGERY PROCESS HAS BEEN AFFECTED, AFTER COORDINATED, THE SURGEON REPLACED WITH A TAPER VIT HEAD WHICH HAS THE SAME CONE ANGEL, AND FINISHED THE SURGERY. THE SURGEON REPLACED WITH A TAPER VIT HEAD WITH A LARGER DIAMETER AND THE SAME CON ANGEL, THE SUBSTITUTE PRODUCT ITEM: 6260-5-228,28MM+4MM, AFTERWARDS THE SURGERY WAS FINISHED.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 6272897 ↗
Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices
- BIOLOX DELTA 1365-32-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented551 reports
- PINNACLE ALTRX 1221-36-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented481 reports
- BIOLOX DELTA 1365-32-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death314 reports
- BIOLOX DELTA 1365-36-720Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death189 reports
- DJO SURGICAL 400-03-362Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented187 reports
- Anthology 71365705Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented176 reports
- BIOLOX DELTA 1365-40-710Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented170 reports
- PINNACLE GRIPTION 1217-32-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented169 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page