V40 Head 6260-5-228Recalls & reported problems
Howmedica Osteonics Corp. · V40 Femoral Head · Class 2 · Implant
All models of V40 Head (30)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07613327012897), received between Dec 6, 2018 and Sep 23, 2024 — classified as 6 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information2
- Adverse Event Without Identified Device or Use Problem1
- Corroded1
- Degraded1
- Device Dislodged or Dislocated1
- Loss of Osseointegration1
- Material Erosion1
Patient problems
- Inflammation2
- Insufficient Information2
- Metal Related Pathology2
- Reaction2
- Ambulation or Postural Difficulties1
- Foreign Body Reaction1
- Inadequate Osseointegration1
- Injury1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjurySep 23, 2024Degraded; Degraded
AS REPORTED BY THE REP: "IT WAS A COMBINATION OF AN ACCOLADE 1 STEM AND COCR +HEAD. IT LOOKS LIKE THE PATIENT IS POTENTIALLY LOOKING TO CHECK/PURSUE ANY INCOMPATIBILITY AS THEY EXPERIENCED METALLOSIS. I HAVE FORWARDED THE HIP COMPOSITION GUIDE. RIGHT HIP; EVENT WAS NOTICED VIA MRI DONE AT CLINIC WHICH SHOWED PSEUDOTUMOR AND ELEVATED METAL ION LEVELS. ELEVATED METAL ION LEVELS (BOTH COBALT AND CHROMIUM ELEVATED) POST PRIMARY SURGERY. DURING REVISION SURGERY, TRUNNIONOSIS WAS NOTED.".
Patient problems: Foreign Body Reaction; Metal Related Pathology
MAUDE report 20284818 ↗InjurySep 4, 2024Loss of Osseointegration; Device Dislodged or Dislocated; Loss of Osseointegration; Device Dislodged or Dislocated
THE HIP JOINT DISLOCATED, AND A REVISION WAS NECESSARY. THE PATIENT HAD A UHR HEAD AND AN ACCOLADE II STEM. INTRAOPERATIVELY, IT TURNED OUT THAT THE HEAD WAS NOT HELD ON THE STEM, AND THE OPERATOR REMOVED THE HEAD FROM THE STEM CONE WITHOUT ANY RESISTANCE. AFTER REMOVING THE HEAD AND STEM, STIMULAN WAS USED BECAUSE THE PATIENT HAD AN INFECTION.
Patient problems: Inadequate Osseointegration; Joint Dislocation; Periprosthetic fracture; Unspecified Infection
MAUDE report 20140626 ↗InjuryAug 2, 2022Corroded; Material Erosion; Corroded; Material Erosion
IT HAS BEEN REPORTED "A REVISION SURGERY WAS REQUIRED DUE TO "RAISED METAL ION LEVELS". SURGEON REPORTED SIGNS OF METALLOSIS AT THE FEMORAL HEAD/TAPER IMPLANT JUNCTION. THE SURGEON DECIDED THE BEST THING FOR THE PATIENT WAS TO REMOVE THE MILLED BODY AND REPLACE IT WITH THE SAME SIZED ONE AND PUT A CERAMIC HEAD ON".
Patient problems: Insufficient Information; Metal Related Pathology
MAUDE report 15154200 ↗InjuryJun 21, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED BY PATIENT'S ATTORNEY AS A RESULT OF A LEGAL CLAIM THAT ALLEGEDLY ON (B)(6) 2019 THE PATIENT UNDERWENT A LEFT TOTAL HIP ARTHROPLASTY AND WAS IMPLANTED WITH AN ACCOLADE II 127° NECK ANGLE HIP STEM AND A V40 FEMORAL HEAD WHICH RESULTED IN REVISION AND PERSONAL INJURIES. UPDATE: PER LEGAL, "THE ALLEGATIONS ARE FOR THE ACCOLADE II STEM AND THE V40 HEAD. CURRENTLY WE DO NOT KNOW WHY THE PATIENT WAS REVISED OR WHAT DEVICES WERE EXPLANTED."
Patient problems: Insufficient Information
MAUDE report 12030829 ↗InjuryJan 24, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
PHARMACIST FROM THE HOSPITAL REPORTED THE FOLLOWING EVENT: " INVOLVEMENT OF THE MEDICAL DEVICE IN THE OCCURRENCE OF METALLOSIS: CHROME-COBALT POISONING WITH CARDIOPATHY ON PROSTHESIS PLACED IN 2014. "
Patient problems: Inflammation; Reaction
MAUDE report 8273819 ↗InjuryDec 6, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT PATIENT'S RIGHT HIP WAS REVISED DUE TO METALLOSIS AFTER PATIENT COMPLAINT OF PAIN AND DIFFICULTY AMBULATING. A 28 +4 HEAD AND 28 10° E LINER WERE REVISED TO A 32 0° LINER AND CERAMIC HEAD WITH SLEEVE. A RESTORATION PS STEM AND TRIDENT PSL CLUSTER SHELL WERE NOT REVISED.
Patient problems: Ambulation or Postural Difficulties; Inflammation; Injury; Pain; Reaction
MAUDE report 8137848 ↗
Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices
- PINNACLE 1217-25-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED735 reports
- NA 1365-52-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED683 reports
- APEX 1246-03-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 deaths613 reports
- PINNACLE 1217-20-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 deaths590 reports
- PINNACLE 1217-30-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 death484 reports
- NA 1365-53-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED404 reports
- PINNACLE ALTRX 1221-36-054Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED388 reports
- PINNACLE ALTRX 1221-32-050Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED386 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page