V40 Head 6260-9-232Recalls & reported problems
Howmedica Osteonics Corp. · LFIT V40 Femoral Head · Class 2 · Implant
All models of V40 Head (30)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Howmedica LFIT V40 femoral head is part of a hip replacement that fits on the stem in the thigh bone. Reports mostly describe the head coming loose from the stem, with corrosion and metal buildup in the body, and mention pain and another operation to replace parts. There are 23 reports that name its identifier, with no deaths and 22 mentioning injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 23
- Deaths reported
- 0
- Injuries reported
- 22
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 23 adverse event reports that list this exact device (DI 07613327013085), received between Jul 3, 2018 and Jun 23, 2025 — classified as 22 injuries, 1 malfunction.
The most frequently reported specific device problem is “Degraded” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem7
- Degraded6
- Detachment of Device or Device Component6
- Insufficient Device Problem Information6
- Device Dislodged or Dislocated3
- Corroded2
- Material Erosion2
- Contamination1
Patient problems
- Injury7
- Metal Related Pathology7
- Foreign Body Reaction6
- Ambulation or Postural Difficulties5
- Reaction5
- Inflammation4
- Insufficient Information3
- Pain3
Recent adverse event reports
The 23 most recent of 23 reports, in the reporter's own words as published by the FDA.
InjuryJun 23, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
CLAIMANT, THROUGH COUNSEL, ALLEGES THAT HE UNDERWENT A RIGHT THA ON (B)(6) 2010, AND WAS IMPLANTED WITH AN LFIT V40 FEMORAL HEAD AND AN ACCOLADE TMZF STEM. HE SUBSEQUENTLY WAS REVISED ON (B)(6) 2024, DUE TO ALLEGED HEAD DISASSOCIATION.
Patient problems: Ambulation or Postural Difficulties
MAUDE report 22311014 ↗InjuryJul 8, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
HIP WASHOUT, EXCHANGE HEAD 32MM +4 V40 LFIT. ROUTINE HEAD CHANGE DURING HIP WASHOUT FOR INFECTION. PROCEDURE AND REVISION OF COMPONENT NOT DUE TO IMPLANT FAILURE.
Patient problems: Unspecified Infection
MAUDE report 19687954 ↗InjuryJun 9, 2024Degraded; Degraded
IT WAS REPORTED BY PATIENT'S COUNSEL AS A RESULT OF A LEGAL CLAIM THAT ALLEGEDLY THE PATIENT UNDERWENT RIGHT THA ON (B)(6) 2016, AND WAS IMPLANTED WITH AN LFIT V40 COCR FEMORAL HEAD AND ACCOLADE II HIP STEM REQUIRING REVISION SURGERY ON (B)(6) 2021, DUE TO ALLEGED METALLOSIS AND CORROSION.
Patient problems: Metal Related Pathology
MAUDE report 19495772 ↗InjuryApr 23, 2024Degraded; Degraded
IT WAS REPORTED BY CLAIMANT'S COUNSEL THAT THE PATIENT UNDERWENT RIGHT THA ON (B)(6) 2016 AND WAS IMPLANTED WITH AN LFIT V40 FEMORAL HEAD AND AN ACCOLADE II HIP STEM. SUBSEQUENTLY SHE PRESENTED WITH INCREASING PAIN AND WORKUP SHOWED ELEVATED COBALT AND CHROMIUM LEVELS. SHE WAS REVISED ON (B)(6) 2022.
Patient problems: Metal Related Pathology; Pain
MAUDE report 19162861 ↗InjuryMar 19, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY CLAIMANT'S COUNSEL THAT ALLEGEDLY THE PATIENT UNDERWENT RIGHT TOTAL HIP ARTHROPLASTY ON (B)(6) 2010, AND WAS IMPLANTED WITH AN LFIT V40 FEMORAL HEAD AND ACCOLADE TMZF STEM. HE UNDERWENT REQUIRED REVISION SURGERY ON (B)(6) 2022, DUE TO ALLEGED HEAD DISASSOCIATION.
Patient problems: Insufficient Information
MAUDE report 18928186 ↗InjuryNov 15, 2022Degraded; Detachment of Device or Device Component; Device Dislodged or Dislocated; Degraded; Detachment of Device or Device Component; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENT'S RIGHT HIP WAS REVISED DUE TO TRUNNION WEAR LEADING TO HEAD DISSOCIATION. THE HEAD AND STEM WERE REVISED. REP CAN PROVIDE PICTURES AND CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Ambulation or Postural Difficulties; Insufficient Information
MAUDE report 15797295 ↗InjuryOct 21, 2022Material Erosion; Material Erosion
IT WAS REPORTED THROUGH THE FILING OF A LAWSUIT THAT ALLEGEDLY THE PATIENT WAS IMPLANTED WITH AN LFIT ANATOMIC COCR V40 FEMORAL HEAD ON HER LEFT HIP ON OR ABOUT (B)(6) 2012 AND WAS REVISED ON (B)(6), 2022. IT IS FURTHER ALLEGED THAT SHE SUFFERED INJURIES AS A RESULT OF IMPLANTATION AND EXPLANTATION OF THE DEVICE(S) AT ISSUE, DEVICE RECALL AND EXCESSIVE LEVELS OF CHROMIUM AND COBALT IN HER BLOOD.
Patient problems: Foreign Body Reaction
MAUDE report 15654883 ↗InjuryOct 12, 2022Degraded; Detachment of Device or Device Component; Degraded; Detachment of Device or Device Component
IT WAS REPORTED THE PATIENT'S HIP WAS REVISED DUE TO FAILURE OF THE FEMORAL NECK (REP CLARIFIED THE TRUNNION LOOKED GROUND DOWN AND THERE WAS A NOTCH IN THE NECK). THE STEM, HEAD, AND LINER WERE REVISED. REP CONFIRMED THERE ARE NO ALLEGATIONS AGAINST THE REVISED LINER.
Patient problems: Ambulation or Postural Difficulties
MAUDE report 15585386 ↗InjuryJul 6, 2022Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
IT WAS REPORTED THROUGH THE FILING OF A LAWSUIT THAT ALLEGEDLY THE PATIENT WAS IMPLANTED WITH AN LFIT ANATOMIC COCR V40 FEMORAL HEAD ON HER RIGHT HIP ON OR ABOUT (B)(6) 2009 AND WAS REVISED ON (B)(6) 2020. IT IS FURTHER ALLEGED THAT SHE SUFFERED INJURIES AS A RESULT OF IMPLANTATION AND EXPLANTATION OF THE DEVICE(S) AT ISSUE, DEVICE RECALL AND EXCESSIVE LEVELS OF CHROMIUM AND COBALT IN HER BLOOD.
Patient problems: Foreign Body Reaction; Metal Related Pathology
MAUDE report 14926578 ↗InjuryMay 27, 2022Material Erosion; Adverse Event Without Identified Device or Use Problem; Material Erosion; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH THE FILING OF A LAWSUIT THAT ALLEGEDLY THE PATIENT WAS IMPLANTED WITH AN LFIT ANATOMIC COCR V40 FEMORAL HEAD ON HIS LEFT HIP ON OR ABOUT (B)(6) 2006 AND WAS REVISED ON (B)(6) 2021. IT IS FURTHER ALLEGED THAT HE SUFFERED INJURIES AS A RESULT OF IMPLANTATION AND EXPLANTATION OF THE DEVICE(S) AT ISSUE, DEVICE RECALL AND EXCESSIVE LEVELS OF CHROMIUM AND COBALT IN HIS BLOOD.
Patient problems: Foreign Body Reaction; Metal Related Pathology
MAUDE report 14521031 ↗InjuryFeb 9, 2022Corroded; Adverse Event Without Identified Device or Use Problem; Corroded; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH THE FILING OF A LAWSUIT THAT ALLEGEDLY THE PATIENT WAS IMPLANTED WITH AN LFIT ANATOMIC COCR V40 FEMORAL HEAD ON HIS LEFT HIP ON OR ABOUT (B)(6) 2009 AND WAS REVISED ON (B)(6) 2021. IT IS FURTHER ALLEGED THAT HE SUFFERED INJURIES AS A RESULT OF IMPLANTATION AND EXPLANTATION OF THE DEVICE(S) AT ISSUE, DEVICE RECALL AND EXCESSIVE LEVELS OF CHROMIUM AND COBALT IN HIS BLOOD.
Patient problems: Discomfort; Foreign Body Reaction; Metal Related Pathology
MAUDE report 13502313 ↗InjuryJan 31, 2022Degraded; Detachment of Device or Device Component; Degraded; Detachment of Device or Device Component
IT WAS REPORTED THAT THE PATIENT REQUIRED REVISION HIP SURGERY DUE TO A DISSOCIATION BETWEEN THE HEAD AND FEMORAL STEM AND THE SUBSTANTIAL DAMAGE TO THE FEMORAL TRUNION. IT WAS FURTHER REPORTED THAT THE REVISION WAS PERFORMED AFTER THE PATIENT PRESENTED TO EMERGENCY DEPARTMENT WITH A DISLOCATION THAT APPEARED UNUSUAL ON XRAY >10 YEARS POST HIP REPLACEMENT.
Patient problems: Ambulation or Postural Difficulties; Insufficient Information
MAUDE report 13392224 ↗InjuryJan 21, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INFORMATION RECEIVED FROM BROADSPIRE: IT WAS REPORTED THAT THE PATIENT HAS PAIN, METALLOSIS - ELEVATED LEVELS OF CHROMIUM & COBALT , CLICKING NOISE. ADDITIONAL INFORMATION RECEIVED FROM LEGAL: PLAINTIFF ALLEGES THE LEFT REJUVENATE HIP IMPLANTED ON OR ABOUT (B)(6) 2010 FAILED CAUSING PAIN AND ELEVATED METAL ION LEVELS. PLAINTIFF HAS NOT YET SCHEDULED REVISION SURGERY. AS REPORTED BY REP: "REVISION OF LEFT TOTAL HIP. PATIENT HAD REJUVENATE STEM REMOVED...PATIENT COBALT LEVEL WAS 18." SPOKE TO REP. A REJUVENATE MODULAR STEM AND NECK, AND COCR HEAD WERE REVISED. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Foreign Body Reaction
MAUDE report 13313428 ↗InjuryJan 14, 2022Unstable; Unstable
THIS PI IS FOR THE RECURRENT DISLOCATIONS ENDING IN (B)(6) 2021. IT WAS REPORTED THROUGH THE FILING OF A LAWSUIT THAT ALLEGEDLY THE PATIENT WAS IMPLANTED WITH AN LFIT ANATOMIC COCR (B)(4) FEMORAL HEAD ON HER RIGHT HIP ON OR ABOUT (B)(6) 2012 AND WAS REVISED ON (B)(6) 2021. IT IS FURTHER ALLEGED THAT SHE SUFFERED INJURIES AS A RESULT OF IMPLANTATION AND EXPLANTATION OF THE DEVICE(S) AT ISSUE, DEVICE RECALL AND EXCESSIVE LEVELS OF CHROMIUM AND COBALT IN HER BLOOD. MEDICAL RECORDS INDICATED THAT THE PATIENT HAD RECURRENT DISLOCATIONS, WITH THE FINAL DISLOCATION OCCURRING IN (B)(6) 2021.
Patient problems: Joint Dislocation; Joint Laxity
MAUDE report 13257063 ↗InjurySep 30, 2021Detachment of Device or Device Component; Device Dislodged or Dislocated; Detachment of Device or Device Component; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENT'S RIGHT HIP WAS REVISED. THE PATIENT HAD FALLEN AND THE HEAD DISASSOCIATED FROM THE STEM. INTRAOPERATIVELY, THE SURGEON NOTICED THAT THERE WAS A LOT OF WEAR AROUND THE TRUNNION. REP NOTED THAT "THE TIP OF THE TRUNNION WAS GONE." REP HAS THE IMPLANTS FOR RETURN.
Patient problems: Ambulation or Postural Difficulties; Fall
MAUDE report 12554362 ↗MalfunctionSep 2, 2020Contamination; Contamination
CONTAMINATED FEMORAL HEAD, MOISTURE APPEARS TO HAVE PENETRATED BOTH BOXES INSIDE PACKAGING BUT OUTER PACKAGE AND PLASTIC WERE INTACT AND NO SIGN OF MOISTURE.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10482174 ↗InjuryJun 13, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
AS REPORTED: "PATIENT EXPERIENCED PAIN IN BACK/ HIP AROUND (B)(6) 2018. IN SUBSEQUENT VISITS A FLUID/ TISSUE MASS WAS SEEN ON IMAGING. REVISION SURGERY WAS PERFORMED ON (B)(6) 2019. ORIGINAL SURGERY WAS IN 2009. DARK FLUID AND TISSUE WAS SEEN IN JOINT. STEM WAS REMOVED AND REPLACED WITH A LONG CEMENTED STEM, CUP WAS RETAINED".
Patient problems: Injury; Pain
MAUDE report 8697204 ↗InjuryFeb 27, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THROUGH THE ATTORNEY FOR THE PATIENT, AS A RESULT OF A LEGAL CLAIM, THAT ALLEGEDLY THE PATIENT RECEIVED A TRIDENT ACETABULAR HIP SYSTEM. IT WAS FURTHER ALLEGED THAT, THE PATIENT IS EXPERIENCING "LOOSENING AND DISCOMFORT" FROM THE SYSTEM. UPDATE FROM MEDICAL REVIEW: A REVISION OF THE RIGHT TOTAL HIP ARTHROPLASTY ACETABULAR COMPONENT WAS PERFORMED FOR A DIAGNOSIS OF ACETABULAR LOOSENING AND METALLOSIS. THE LINER SHOWED "ANTERIOR IMPINGEMENT". THE STEM WAS RETAINED AS THE TAPER WAS NOTED TO BE "ACCEPTABLE".
Patient problems: Discomfort; Inadequate Osseointegration; Inflammation; Injury; Reaction
MAUDE report 8373695 ↗InjuryFeb 25, 2019Corroded; Corroded
IT WAS REPORTED THAT THE PATIENT'S LEFT HIP WAS REVISED DUE TO ELEVATED CHROMIUM LEVELS, AND SUSPECTED CORROSION OF THE HEAD/ TRUNNION INTERFACE. INTRA-OPERATIVELY, CORROSION WAS CONFIRMED. A STEM, HEAD AND LINER WERE REVISED TO A RESTORATION MODULAR STEM CONSTRUCT, CERAMIC HEAD, AND MDM/ADM LINER CONSTRUCT. REP PROVIDED EXPLANT PICTURES AND REPORTED THAT X-RAYS, MEDICAL RECORDS, AND ADDITIONAL INFORMATION ARE NOT AVAILABLE.
Patient problems: Inflammation; Injury; Reaction
MAUDE report 8367293 ↗InjuryFeb 22, 2019Degraded; Insufficient Device Problem Information; Degraded; Insufficient Device Problem Information
AS REPORTED: "PATIENT WAS BROUGHT IN FOR A HIP REVISION. PATIENT WAS HAVING PAIN AND HIGH COBALT LEVELS INDICATED METALLOSIS PER [SURGEON]. EXPLANTED HEAD AND LINER BUT RETAINED STEM. METAL DEBRIS AND SOFT TISSUE DAMAGE NOTED, DAMAGE TO TRUNNION OF STEM BUT DIDN'T WARRANT PULLING IMPLANT. [SURGEON] HAD TO DEBRIDE SOFT TISSUE AND REPAIR TO THE BEST OF HIS ABILITY. REPLACED HEAD AND LINER". PATIENT WAS REVISED TO A 36 +5 BIOLOX HEAD AND SLEEVE AND 36E LINER.
Patient problems: Foreign Body Reaction; Injury; Pain; Reaction
MAUDE report 8362372 ↗InjuryOct 16, 2018Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
IT WAS REPORTED THROUGH THE FILING OF A LAWSUIT THAT ALLEGEDLY THE PATIENT WAS IMPLANTED WITH AN LFIT ANATOMIC COCR V40 FEMORAL HEAD ON HIS LEFT HIP ON (B)(6) 2011 AND WAS REVISED ON (B)(6) 2015. IT IS FURTHER ALLEGED THAT THE PATIENT SUFFERED INJURIES AS A RESULT OF IMPLANTATION AND EXPLANTATION OF THE DEVICE AT ISSUE AND EXCESSIVE LEVELS OF CHROMIUM AND COBALT IN HIS BLOOD.
Patient problems: Inflammation; Injury; Metal Related Pathology; Reaction
MAUDE report 7968762 ↗InjurySep 13, 2018Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
IT WAS REPORTED THROUGH THE FILING OF A LAWSUIT THAT ALLEGEDLY THE PATIENT WAS IMPLANTED WITH AN LFIT ANATOMIC COCR V40 FEMORAL HEAD ON HER LEFT HIP ON OR ABOUT (B)(6) 2009 AND WAS REVISED ON (B)(6) 2018. IT IS FURTHER ALLEGED THAT SHE SUFFERED INJURIES AS A RESULT OF IMPLANTATION AND EXPLANTATION OF THE DEVICE AT ISSUE, AND EXCESSIVE LEVELS OF CHROMIUM AND COBALT.
Patient problems: Inflammation; Injury; Metal Related Pathology; Reaction
MAUDE report 7873923 ↗InjuryJul 3, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated
DISASSOCIATION OF TRUNION/FEMORAL HEAD. PATIENT IDENTIFIED AFTER 2007 PRIMARY HIP SURGERY. PATIENT REQUIRED REVISION HIP SURGERY.
Patient problems: Foreign Body In Patient; Injury
MAUDE report 7659722 ↗
Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices
- NA 1365-51-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented753 reports
- NA 1365-50-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented412 reports
- ARTICUL/EZE 1365-21-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall280 reports
- ARTICUL/EZE 1365-11-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 death262 reports
- LFIT 6260-9-236Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, cemented256 reports
- PINNACLE GRIPTION 1217-32-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented2 deaths239 reports
- ELEOS™ 25002111EOnkos Surgical, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented3 deaths230 reports
- Novation 130-36-52Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented2 recalls124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page