Morse Head 01-2200Recalls & reported problems
Howmedica Osteonics Corp. · Femoral Head · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613327016086), received between Mar 13, 2020 and Jun 18, 2020 — classified as 2 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
3 other recalls cover devices cleared under the same FDA submission (K910988, K993601); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated2
Patient problems
- Injury2
- Joint Dislocation2
- Blood Loss1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJun 18, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated
ON (B)(6) 1998 THE FEMALE PATIENT WAS INITIALLY INSERTED THA CUP, WHICH WAS REVISED ON (B)(6) 2020. IT WAS PLANNED TO REPLACE FROM THE LINER OF 22 MM DIAMETER TO THE LINER OF 32 MM DIAMETER AFTER REMOVING THE NECK. BUT THE NECK COULD NOT BE REMOVED, AND IT REMAINED AS IT WAS, AND CLOSED AT THAT TIME. AFTER THAT, SHE BEGAN TO DISLOCATE MORE FREQUENTLY, AND ON (B)(6) 2020, THE STEM SIDE WAS ALSO PREPARED IN CONSIDERATION OF THE POSSIBILITY THAT THE NECK WON¿T BE COME OUT. THE CONDITION OF THE FEMUR BONE WAS ALSO GOOD, AND THE SURGEON WANTED TO REMOVE THE NECK AS MUCH AS POSSIBLE AND ONLY REPLACE THE NECK AND LINER BECAUSE THE STEM IS FIRMLY FIXED. BUT DESPITE THE FACT THAT THE SURGEON WAS CAREFULLY PREPARING FOR THE REMOVAL OF THE NECK, AFTER ALL, THE FEMUR COULD NOT BE PULLED OUT, AND THE FEMUR WAS VERTICALLY SPLIT AND THE STEM WAS REMOVED, AND A NEW STEM, NECK 32 MM, AND LINER WERE REPLACED. INITIALLY, IT WAS PLANNED TO BE LESS THAN 1-HOUR OP IF THE NECK GOT OUT, BUT FINALLY IT TOOK 5 HOURS. THE SURGEON REQUESTED TO INVESTIGATE THE REASON; WHY HE COULD NOT REMOVE THE PRODUCT. HE HAD NEVER EXPERIENCED A CASE WHERE THE NECK COULD NOT COME OUT SO DIFFICULT AND TOOK SO LONG. ORIGINALLY
Patient problems: Blood Loss; Injury; Joint Dislocation
MAUDE report 10170740 ↗InjuryMar 13, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE PATIENT UNDERWENT BIPOLAR HIP ARTHROPLASTY IN 1996. THE BIPOLAR CUP WAS BROKEN, AND DISLOCATION OCCURRED. THEREFORE, THE BIPOLAR CUP AND THE INNER HEAD WERE REMOVED AND THA WAS PERFORMED ON (B)(6) 2020.
Patient problems: Injury; Joint Dislocation
MAUDE report 9828617 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K910988, K993601. One submission often covers many models, so check the product and lot details in each record.
- Z-0390-2010 · Sep 4, 2009 · TerminatedStryker Howmedica Osteonics, Skirted 5 40' Taper V40 Femoral Head (26 mm +8mm offset), hip prosthesis component.
- Z-0391-2010 · Sep 4, 2009 · TerminatedStryker Howmedica Osteonics, Skirtless 5 40' LFIT V40 Femoral Head (32 mm +0 mm offset); hip prosthesis component.
- Z-1822-2009 · Jun 18, 2009 · TerminatedV 40 Femoral Head, Sterile, 22 mm; Catalog number 6260-4-122, Catalog number 6260-5-226; Howmedica Osteonics Corp., Stryker Ireland, Carrigtwohill County Cork, Ireland; Intended f
Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices
- NA 1365-51-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented753 reports
- NA 1365-50-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented412 reports
- ARTICUL/EZE 1365-21-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall280 reports
- ARTICUL/EZE 1365-11-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 death262 reports
- LFIT 6260-9-236Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, cemented256 reports
- PINNACLE GRIPTION 1217-32-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented2 deaths239 reports
- ELEOS™ 25002111EOnkos Surgical, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented3 deaths230 reports
- Novation 130-36-52Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented2 recalls124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page