C-Taper Head 06-2898Recalls & reported problems
Howmedica Osteonics Corp. · Low Friction Ion Treatment · Class 2 · Implant
All models of C-Taper Head (18)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613327016284), received between Apr 16, 2019 and Oct 21, 2019 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Device Difficult to Setup or Prepare” (2 reports).
3 other recalls cover devices cleared under the same FDA submission (K910988, K993601); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device Difficult to Setup or Prepare2
Patient problems
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 21, 2019Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare
PRIMARY PROCEDURE, RIGHT BIPOLAR HIP. IT WAS REPORTED THAT THE 28 -3 LFIT HEAD WOULD NOT LOCK ONTO THE OMNIFIT HFX STEM. THE SURGEON TRIED 3 TIMES. AS THE REP WENT TO PULL ANOTHER FEMORAL HEAD, THE SURGEON GAVE THE HEAD A THOROUGH CLEANING AND MADE 3 MORE ATTEMPTS. A 28 -2.5 C-TAPER BIOLOX HEAD WAS OPENED AND LOCKED ON TO THE SAME STEM AS EXPECTED. THE BIPOLAR COMPONENT WAS THEN IMPLANTED. SURGERY WAS COMPLETED SUCCESSFULLY WITH A DELAY OF APPROXIMATELY 1 MINUTE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9216233 ↗MalfunctionApr 16, 2019Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare
PRIMARY PROCEDURE, LEFT BIPOLAR HIP. IT WAS REPORTED THAT AFTER IMPACTING THE C-TAPER LFIT HEAD, THE HEAD WAS ABLE TO BE REMOVED FROM THE TRUNNION USING ONE FINGER. SURGEON REMOVED THE HEAD, CLEANED AND DRIED IT, RE-IMPACTED IT WITH THE SAME RESULT. A SECOND HEAD WAS OPENED AND PERFORMED AS EXPECTED. SURGERY WAS COMPLETED SUCCESSFULLY WITH A DELAY OF APPROXIMATELY 2 MINUTES. THE REP MIGHT BE ABLE TO OBTAIN THE OPERATIVE REPORT AND REPORTED THAT NO FURTHER INFORMATION WILL BE RELEASED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8518008 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K910988, K993601. One submission often covers many models, so check the product and lot details in each record.
- Z-0390-2010 · Sep 4, 2009 · TerminatedStryker Howmedica Osteonics, Skirted 5 40' Taper V40 Femoral Head (26 mm +8mm offset), hip prosthesis component.
- Z-0391-2010 · Sep 4, 2009 · TerminatedStryker Howmedica Osteonics, Skirtless 5 40' LFIT V40 Femoral Head (32 mm +0 mm offset); hip prosthesis component.
- Z-1822-2009 · Jun 18, 2009 · TerminatedV 40 Femoral Head, Sterile, 22 mm; Catalog number 6260-4-122, Catalog number 6260-5-226; Howmedica Osteonics Corp., Stryker Ireland, Carrigtwohill County Cork, Ireland; Intended f
Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices
- NA 1365-51-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented753 reports
- NA 1365-50-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented412 reports
- ARTICUL/EZE 1365-21-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall280 reports
- ARTICUL/EZE 1365-11-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 death262 reports
- LFIT 6260-9-236Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, cemented256 reports
- PINNACLE GRIPTION 1217-32-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented2 deaths239 reports
- ELEOS™ 25002111EOnkos Surgical, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented3 deaths230 reports
- Novation 130-36-52Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented2 recalls124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page