C-Taper Head S-1400-HH34Recalls & reported problems
Howmedica Osteonics Corp. · Femoral Head · Class 2 · Implant
All models of C-Taper Head (18)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07613327016734), received between Sep 21, 2018 and Oct 5, 2018 — classified as 3 injuries.
The most frequently reported specific device problem is “Degraded” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Degraded2
- Material Deformation1
Patient problems
- Injury3
- Reaction3
- Inflammation2
- Edema1
- Pain1
- Unspecified Infection1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryOct 5, 2018Degraded; Degraded
IT WAS REPORTED THAT PATIENT'S RIGHT HIP WAS REVISED DUE TO ELEVATED COBALT AND CHROMIUM LEVELS AFTER PATIENT COMPLAINT OF SEVERE PAIN SINCE 2013. INTRA-OPERATIVELY, TRUNNIONOSIS WAS NOTED. THE HEAD AND LINER WERE EXCHANGED FOR A 36E LINER AND SIZE 36 CERAMIC HEAD WITH A +5 ADAPTER SLEEVE - STEM WAS CLEANED OFF AND RETAINED.
Patient problems: Injury; Pain; Reaction
MAUDE report 7937986 ↗InjurySep 25, 2018Degraded; Degraded
"ON (B)(6), 2008 AN EXPLORATION INCISION AND DRAINAGE OF THE RIGHT HIP AND THIGH WAS PERFORMED FOR A DIAGNOSIS OF METALLOSIS RIGHT HIP AND INTRAPELVIC FLUID."
Patient problems: Edema; Inflammation; Injury; Reaction
MAUDE report 7906345 ↗InjurySep 21, 2018Material Deformation; Material Deformation
REVIEW OF THE MEDICAL REVIEW INDICATED, "ON (B)(6) 2008 AN EXPLORATION OF THE RIGHT TOTAL HIP, INCISION AND DRAINAGE, AND VAC DRESSING APPLICATION WAS PERFORMED FOR A POST-OPERATIVE DIAGNOSIS OF EXTENSIVE METALLOSIS RIGHT HIP. THE OPERATIVE REPORT DESCRIBES A LATERAL APPROACH AND FINDINGS INCLUDING METALLOSIS OF THE RIGHT HIP AND THIGH WITH DAMAGE AND WEAR TO THE POLY LINER AND HEAD. EXTENSIVE METALLOSIS AROUND THE PROXIMAL FEMORAL COMPONENT AND BEHIND THE ACETABULAR COMPONENT WAS NOTED AND THE SURGEON DID NOT CLOSE THE FASCIA AND KEPT A SPONGE IN THE PROXIMAL FEMUR AND ACETABULUM AFTER EXPLANTING THE TOTAL HIP COMPONENTS. ON (B)(6) 2008 DEBRIDEMENT WAS PERFORMED AND A VAC SYSTEM WAS APPLIED. CULTURE WAS NOTED TO BE POSITIVE FOR METHICILLIN SENSITIVE STAPH AUREUS. WHEN SEEN ON (B)(6) 2009 SHE WAS NOTED TO BE 6-CENTIMETERS SHORT ON THE RIGHT LOWER EXTREMITY." "ON (B)(6) 2009 A RIGHT HIP WOUND REVISION WITH INCISION AND DRAINAGE AND INSERTION OF AN ANTIBIOTIC SPACER WAS PERFORMED FOR AN INFECTED GIRDLESTONE WITH CHRONIC OSTEOMYELITIS AND PARTICLE DISEASE OF THE RIGHT HIP." "ON (B)(6) 2009 REMOVAL OF THE ANTIBIOTIC SPACER AND CONVERSION TO A TOTAL HIP ARTHROPLASTY WAS PERFORMED FOR A
Patient problems: Inflammation; Injury; Reaction; Unspecified Infection
MAUDE report 7897736 ↗
Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices
- NA 1365-51-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented753 reports
- NA 1365-50-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented412 reports
- ARTICUL/EZE 1365-21-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall280 reports
- ARTICUL/EZE 1365-11-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 death262 reports
- LFIT 6260-9-236Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, cemented256 reports
- PINNACLE GRIPTION 1217-32-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented2 deaths239 reports
- ELEOS™ 25002111EOnkos Surgical, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented3 deaths230 reports
- Novation 130-36-52Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented2 recalls124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.