UHR UH1-41-26Recalls & reported problems
Howmedica Osteonics Corp. · Bipolar Component · Class 2 · Implant
All models of UHR (40)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613327016802), received between Feb 24, 2021 and Feb 24, 2021 — classified as 1 injury.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Jul 30, 2026Class IIOpen, Classified
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Z-3199-2026 · Root cause: Process change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Dislodged or Dislocated1
Patient problems
- Joint Dislocation1
- Unspecified Infection1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryFeb 24, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENT'S RIGHT HIP WAS REVISED. AS REPORTED BY REP: "RIGHT HIP OPEN REDUCTION, I&D, POSSIBLE REVISION TO DUAL MOBILITY OR CONSTRAINED HIP REPLACEMENT, POSSIBLE RESECTIONAL ARTHROPLASTY, AND CEMENT SPACER. PRE-OP DIAGNOSIS: RIGHT HIP DISLOCATION AND POSSIBLE INFECTION, STATUS POST BIPOLAR HEMIARTHROPLASTY. REMOVED ALL COMPONENTS." AN ACCOLADE II STEM, V40 METAL HEAD, AND BIPOLAR COMPONENT WERE REVISED. REP PROVIDED OPERATIVE REPORTS FOR PRIMARY AND (B)(6) 2020 REVISION, X-RAYS PRIOR TO (B)(6) 2021 REVISION, AND USAGE SHEET FROM (B)(6) 2020 REVISION AND CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Joint Dislocation; Unspecified Infection
MAUDE report 11372221 ↗
FDA clearance and classification
The FDA regulates product code KWL (prosthesis, hip, hemi-, femoral, metal) under 21 CFR 888.3360 as a Class II device. 21 CFR 888.3360 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the UHR UH1-41-26
Has the UHR UH1-41-26 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-3199-2026 (Class II), began on Jul 30, 2026: Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the UHR UH1-41-26?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Feb 24, 2021. That report lists device problem “Device Dislodged or Dislocated” (1); patient problems “Joint Dislocation” (1) and “Unspecified Infection” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the UHR UH1-41-26?
None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the UHR UH1-41-26 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the UHR UH1-41-26, and is it still sold?
The FDA's device database lists Howmedica Osteonics Corp. as the company responsible for it (brand UHR, model UH1-41-26). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Prosthesis, hip, hemi-, femoral, metal devices
- UHR UH1-48-28Howmedica Osteonics Corp. · Prosthesis, hip, hemi-, femoral, metal2 recalls3 reports
- "UNITED" USTAR II Knee System 2515-3111United Orthopedic Corp. · Prosthesis, Hip, Hemi-, Femoral, Metal2 recalls0 reports
- "UNITED" USTAR II Knee System 2515-3213United Orthopedic Corp. · Prosthesis, Hip, Hemi-, Femoral, Metal2 recalls0 reports
- STAGEONE 431207Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal2 recalls0 reports
- "UNITED" USTAR II Knee System 2515-3115United Orthopedic Corp. · Prosthesis, Hip, Hemi-, Femoral, Metal2 recalls0 reports
- ACTIS 101011050Depuy (Ireland) · Prosthesis, hip, hemi-, femoral, metal73 reports
- UHR UH1-45-26Howmedica Osteonics Corp. · Prosthesis, hip, hemi-, femoral, metal1 recall11 reports
- UHR UH1-46-26Howmedica Osteonics Corp. · Prosthesis, hip, hemi-, femoral, metal1 recall11 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page