Skip to content

UHR UH1-44-26Recalls & reported problems

Howmedica Osteonics Corp. · Bipolar Component · Class 2 · Implant

All models of UHR (40)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
7
Deaths reported
0
Injuries reported
7
Recalls naming it
1

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 07613327016888), received between Nov 9, 2017 and Nov 12, 2025 — classified as 7 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (5 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 30, 2026Class IIOpen, Classified

    Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

    Z-3199-2026 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Dislodged or Dislocated5
  • Adverse Event Without Identified Device or Use Problem1
  • Insufficient Device Problem Information1

Patient problems

  • Injury4
  • Joint Dislocation4
  • Hematoma1
  • Unspecified Infection1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 12, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THE FOLLOWING INFORMATION WAS PROVIDED: IT WAS REPORTED THAT THE PATIENT'S LEFT HIP WAS REVISED DUE TO RECURRENT DISLOCATIONS OF THE BIPOLAR COMPONENT FROM THE PATIENT'S NATIVE ACETABULUM. THE HIP WAS CONVERTED TO A TOTAL HIP (STEM WAS RETAINED). REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Joint Dislocation

    MAUDE report 23536384 ↗
  • InjuryMar 12, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT'S LEFT HEMI HIP WAS REVISED DUE TO SUSPECTED INFECTION. A BIPOLAR COMPONENT AND FEMORAL HEAD WERE REVISED. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Unspecified Infection

    MAUDE report 18884448 ↗
  • InjuryOct 21, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THE PATIENT'S LEFT HIP WAS REVISED DUE TO DISLOCATION OF THE BIPOLAR COMPONENT FROM THE PATIENT'S NATIVE ACETABULUM. THE BIPOLAR COMPONENT AND HEAD WERE REVISED TO A TOTAL HIP. REP CONFIRMED THERE ARE NO ALLEGATIONS AGAINST THE REVISED FEMORAL HEAD. BRANCH PROVIDED THE PRIMARY USAGE SHEET, AND REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Joint Dislocation

    MAUDE report 10709898 ↗
  • InjuryJul 24, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE PATIENT'S RIGHT HIP WAS REVISED DUE TO HEMATOMA WITH NO ACUTE INFLAMMATION. AN IRRIGATION AND DEBRIDEMENT WITH WASHOUT AND EXCHANGE OF THE FEMORAL HEAD AND BIPOLAR COMPONENT WERE PERFORMED. PATIENT WAS REVISED TO DEVICES OF THE SAME CATALOG NUMBER. REP REPORTED THAT NO FURTHER INFORMATION WILL BE AVAILABLE.

    Patient problems: Hematoma; Injury

    MAUDE report 8821678 ↗
  • InjuryAug 6, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT PATIENT'S LEFT HIP WAS REVISED AFTER MULTIPLE DISLOCATIONS. AS REPORTED BY REP: "BIPOLAR COMPONENT AND FEMORAL HEAD COMPONENT WERE RE-SIZED AND SWAPPED OUT". PATIENT REVISED FROM A 26 +0 METAL HEAD AND 44MM BIPOLAR COMPONENT TO 26 +4 METAL HEAD AND 43MM BIPOLAR COMPONENT. BIPOLAR/ HEAD CONSTRUCT DISLOCATED FROM PATIENT'S NATIVE ACETABULUM.

    Patient problems: Injury; Joint Dislocation

    MAUDE report 7754514 ↗
  • InjuryApr 19, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT PATIENT'S LEFT HIP WAS REVISED DUE TO DISLOCATION.

    Patient problems: Injury; Joint Dislocation

    MAUDE report 7443317 ↗
  • InjuryNov 9, 2017Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT PATIENT'S RIGHT HIP WAS REVISED FROM A HEMI ARTHROPLASTY TO A THR. PATIENT WAS REVISED DUE TO CHRONIC DISLOCATION OF THE HEAD FROM THE BIPOLAR COMPONENT.

    Patient problems: Injury

    MAUDE report 7014399 ↗

FDA clearance and classification

The FDA regulates product code KWL (prosthesis, hip, hemi-, femoral, metal) under 21 CFR 888.3360 as a Class II device. 21 CFR 888.3360 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the UHR UH1-44-26

Has the UHR UH1-44-26 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-3199-2026 (Class II), began on Jul 30, 2026: Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the UHR UH1-44-26?

The FDA's MAUDE database lists 7 adverse event reports for this device, received between Nov 9, 2017 and Nov 12, 2025. In those reports, the most common specific device problem is “Device Dislodged or Dislocated” (5); the most common patient problems are “Injury” (4), “Joint Dislocation” (4) and “Hematoma” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the UHR UH1-44-26?

None of the reports are classified as deaths, 7 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the UHR UH1-44-26 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the UHR UH1-44-26, and is it still sold?

The FDA's device database lists Howmedica Osteonics Corp. as the company responsible for it (brand UHR, model UH1-44-26). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Prosthesis, hip, hemi-, femoral, metal devices

All Prosthesis, hip, hemi-, femoral, metal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page