Skip to content

UHR UH1-46-28Recalls & reported problems

Howmedica Osteonics Corp. · Bipolar Component · Class 2 · Implant

All models of UHR (41)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
7
Deaths reported
0
Injuries reported
5
Recalls naming it
1

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 07613327016956), received between May 8, 2018 and Dec 5, 2023 — classified as 5 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 30, 2026Class IIOpen, Classified

    Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

    Z-3199-2026 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Dislodged or Dislocated3
  • Detachment of Device or Device Component2
  • Adverse Event Without Identified Device or Use Problem1
  • Device Difficult to Setup or Prepare1
  • Fracture1
  • Mechanical Problem1

Patient problems

  • Joint Dislocation2
  • Fall1
  • Hematoma1
  • Injury1
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1
  • Unspecified Infection1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 5, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    BHA WAS PERFORMED FOR LEFT FEMORAL NECK FRACTURE ON (B)(6)2023. ON NOVEMBER 10, 2023, DISLOCATION WAS CONFIRMED. DURING THE DOCTOR ATTEMPTED TO REPOSITIONING THE DEVICE, THE INNER HEAD WAS DISLOCATED FROM THE UHR BIPOLAR CUP. THUS, THE REVISION SURGERY WAS PERFORMED ON (B)(6) 2023.

    Patient problems: Joint Dislocation

    MAUDE report 18266062 ↗
  • InjuryMay 9, 2022Detachment of Device or Device Component; Device Dislodged or Dislocated; Detachment of Device or Device Component; Device Dislodged or Dislocated

    DISLOCATION DUE TO PATIENT FALL. IT COULD NOT BE REDUCED AT ANOTHER HOSPITAL. DURING REDUCTION THE BIPOLAR CUP AND HEAD CAME OFF. THE BIPOLAR CUP AND HEAD WERE REPLACED.

    Patient problems: Fall

    MAUDE report 14335816 ↗
  • InjuryApr 8, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT'S RIGHT HIP WAS REVISED DUE TO INFECTION. AN IRRIGATION AND DEBRIDEMENT WAS PERFORMED, AND A BIPOLAR COMPONENT AND FEMORAL HEAD WERE EXCHANGED FOR ITEMS OF THE SAME CATALOG NUMBERS.

    Patient problems: Unspecified Infection

    MAUDE report 14050242 ↗
  • InjuryMar 14, 2022Detachment of Device or Device Component; Device Dislodged or Dislocated; Detachment of Device or Device Component; Device Dislodged or Dislocated

    THE 46MM BIPOLAR AND 28MM INNER HEAD DISASSOCIATED DURING REDUCTION AFTER THE PATIENT DISLOCATED POST OPERATIVELY. PATIENT WAS REVISED.

    Patient problems: Joint Dislocation

    MAUDE report 13757893 ↗
  • InjuryMay 11, 2020Fracture; Fracture

    PROCEDURE: THE PATIENT HAD A LEFT FEMORAL NECK FRACTURE AND UNDERWENT INSERTION OF ANTIBIOTICS AS A DRUG DELIVERY DEVICE FOR LEFT HIP HEMIARTHROPLASTY (B)(6) 2019. THE PATIENT HAD IRRIGATION AND DEBRIDEMENT OF LEFT HIP DEEP HEMATOMA. PATIENT UNDERWENT REVISION LEFT TOTAL HIP ARTHROPLASTY, SINGLE COMPONENT CHANGED (B)(6) 2019.

    Patient problems: Hematoma; Injury

    MAUDE report 10043600 ↗
  • MalfunctionDec 28, 2018Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare

    WHEN THE METAL HEAD 6001-2810 IS PRESSED INTO DUAL MOBILE UH1-46-28 INTROPERATIVELY,HEAD CAN'T MOVE FREELY WITHOUT PUSHING WITH HAND HARD. CLUNK OCCURS WHEN MOVING. MAYBE THESE TWO MODELS CAN'T BE USED AND NEEDED CHANGING. EVENT UPDATE FROM (B)(6) 2018: SURGICAL DELAY IS NEARLY 1 HOUR. THE SURGEON SPENT NEARLY 20 MINUTES TO FIND ROOT CAUSE. AND THEN, THE SURGEON USED NEARLY 40 MINUTES TO CONFIRM THE FINAL REPLACEMENTS SIZE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8202357 ↗
  • MalfunctionMay 8, 2018Mechanical Problem; Mechanical Problem

    IT IS REPORTED THAT A STRYKER HEAD WAS NOT SLIDING SMOOTHLY IN THE UHR BIPOLAR CUP.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7493722 ↗