Howmedica Osteonics Corp. · Bipolar Component · Class 2 · Implant

All models of UHR (41)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
7
Deaths reported
0
Injuries reported
4
Recalls naming it
1

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 07613327016970), received between May 29, 2019 and May 19, 2026 — classified as 4 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 30, 2026Class IIOpen, Classified

    Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

    Z-3199-2026 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Dislodged or Dislocated2
  • Adverse Event Without Identified Device or Use Problem1
  • Device Contaminated During Manufacture or Shipping1
  • Device Damaged Prior to Use1
  • Device Difficult to Setup or Prepare1
  • Fitting Problem1
  • Fracture1

Patient problems

  • No Consequences Or Impact To Patient3
  • Joint Dislocation2
  • Insufficient Information1
  • No Known Impact Or Consequence To Patient1
  • Unspecified Infection1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 19, 2026Fracture

    THE FOLLOWING INFORMATION WAS PROVIDED: REMOVAL OF BIPOLAR UHR AND INNER HEAD AND REVISION TO CEMENTED CUP/CAGE. REMOVAL OF H-GRIP AND DALL MILES CABLES , 1X CABLE BROKEN. NOTIFICATION ONLY.

    Patient problems: Insufficient Information

    MAUDE report 25214164 ↗
  • InjuryFeb 6, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THA WAS PERFORMED ABOUT 2 MONTHS AGO. TWO MONTHS AFTER SURGERY, THE PATIENT DISLOCATED HIP. THE CUP WAS DISLODGED DURING MANUAL REPOSITIONING. LATER, THE CUP WAS REPLACED. THE HEAD WAS ALSO REPLACED AND THE NECK LENGTH WAS EXTENDED.

    Patient problems: Joint Dislocation

    MAUDE report 21321724 ↗
  • InjuryJul 13, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    LEFT HIP REVISION - HEMIARTHROSCOPY TO TOTAL HIP REPLACEMENT DUE TO RECURRENT DISLOCATIONS. REMOVED BIPOLAR HEAD AND REPLACED WITH MDM + TRITANIUM HEMISPHERICAL CUP. RECURRENT DISLOCATIONS SINCE INITIAL HEMIARTHROSCOPY ON (B)(6) 2023 - REVISED TO TOTAL HIP ON (B)(6) 2023.

    Patient problems: Joint Dislocation

    MAUDE report 17322280 ↗
  • InjuryJun 12, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    LEFT HIP HEMI ARTHROPLASTY. THE REVISION TODAY WAS SECONDARY TO INFECTION. NO FURTHER INFORMATION IS AVAILABLE FROM SURGEON OR HOSPITAL.

    Patient problems: Unspecified Infection

    MAUDE report 17103044 ↗
  • MalfunctionAug 29, 2019Device Difficult to Setup or Prepare; Fitting Problem; Device Difficult to Setup or Prepare; Fitting Problem

    PRIMARY PROCEDURE, RIGHT HIP. IT WAS REPORTED THAT AFTER ASSEMBLY, THE 28 -4 BIOLOX HEAD WOULD NOT ARTICULATE FREELY WITH THE UHR BIPOLAR COMPONENT. THE DEVICES WERE WASTED AND A 28 -4 LFIT HEAD WITH A COMPETITOR BIPOLAR COMPONENT WAS USED TO COMPLETE THE SURGERY SUCCESSFULLY WITH A SURGICAL DELAY OF APPROXIMATELY 10 MINUTES. A USAGE SHEET SHOWING THE WASTED IMPLANTS WAS SUPPLIED, AND THE DEVICES ARE AVAILABLE FOR RETURN, BUT NO FURTHER INFORMATION WILL BE AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 8946999 ↗
  • MalfunctionJul 30, 2019Device Damaged Prior to Use; Device Damaged Prior to Use

    DURING THE SURGERY, THE DOCTOR NOTICED THAT TWO ITEMS WERE DEFECTIVE. ADDITIONAL INFORMATION: DURING THE SURGERY, THE OPERATOR STATED THAT THE HEAD THAT WAS SLAMMED IN THE HEAD OF UH1 WEAKLY MOVES AND WILL CAUSE DISLOCATION IN THE JOINT, HE DECIDED TO CHANGE THE HEAD AND HEAD OF UH1.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8841783 ↗
  • MalfunctionMay 29, 2019Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    UHR HEAD CONTAMINATED WITH BLUE AND BLACK STUFF. DURING SURGERY 2-3 MIN DELAY TO GET A NEW ONE FROM SHELF.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8649359 ↗

Other Prosthesis, hip, hemi-, femoral, metal devices

All Prosthesis, hip, hemi-, femoral, metal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page