Secur-Fit 6052-0935SRecalls & reported problems
Howmedica Osteonics Corp. · 127 Degree Neck Angle Hip Stem · Class 2 · Implant
All models of Secur-Fit (45)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 11
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 07613327021127), received between Apr 10, 2017 and Aug 22, 2022 — classified as 11 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (5 reports).
2 other recalls cover devices cleared under the same FDA submission (K041170, K051738); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Insufficient Device Problem Information5
- Adverse Event Without Identified Device or Use Problem1
- Corroded1
- Device Damaged by Another Device1
- Fracture1
- Inadequacy of Device Shape and/or Size1
- Loss of Osseointegration1
- Migration1
Patient problems
- Injury9
- Bone Fracture(s)5
- Fall2
- Hip Fracture2
- Joint Dislocation2
- Foreign Body Reaction1
- Inadequate Osseointegration1
- Inflammation1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
InjuryAug 22, 2022Corroded; Corroded
IT WAS REPORTED THAT THE PATIENT'S LEFT HIP WAS CONVERTED FROM A HEMI TO A TOTAL HIP DUE TO THE MODULAR ENDO HEAD MIGRATING THROUGH THE ACETABULUM. INTRA-OPERATIVELY, METAL STAINING WAS NOTED AT THE JUNCTION OF THE STEM AND SLEEVE, AND SOME METAL STAINING IN THE SURROUNDING TISSUE.
Patient problems: Foreign Body Reaction
MAUDE report 15266751 ↗InjuryApr 4, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED: "LEFT HIP. REVISED STEM DUE TO PERIPROSTHETIC FRACTURE. SPOKE TO REP, WHO CONFIRMED FEMORAL FRACTURE AND THAT THERE ARE NO ALLEGATIONS AGAINST THE REVISED UNITRAX HEAD.
Patient problems: Bone Fracture(s)
MAUDE report 13993232 ↗InjuryMar 2, 2020Migration; Migration
IT WAS REPORTED THAT THE PATIENT'S RIGHT HIP WAS REVISED DUE TO SUBSIDENCE OF THE FEMORAL STEM. THE SURGEON REPORTED THE IMPLANTED STEM WAS TOO SMALL FOR THE PATIENT'S ANATOMY. THE STEM AND HEAD WERE REVISED TO A LARGER STEM AND ANOTHER FEMORAL HEAD. THERE ARE NO ALLEGATIONS AGAINST THE REVISED STEM OR HEAD. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE AVAILABLE.
Patient problems: Injury
MAUDE report 9775645 ↗InjuryFeb 12, 2020Device Damaged by Another Device; Device Damaged by Another Device
IT WAS REPORTED THAT THE PATIENT'S LEFT HIP WAS REVISED DUE TO SEVERE NOTCHING OF THE NECK OF THE STEM, LEADING TO SURGEON'S CONCERN THAT THE NECK OF THE STEM MAY BREAK OFF. SURGEON REPORTED THE REASON FOR NOTCHING WAS DUE TO IMPINGEMENT WITH THE MDM METAL LINER, NOT THE SHELL. INTRA-OPERATIVELY, A SIGNIFICANT AMOUNT OF METALLOSIS WAS NOTED IN THE PATIENT'S SOFT TISSUE. THE PATIENT'S ENTIRE HIP CONSTRUCT WAS REVISED TO A RESTORATION MODULAR STEM CONSTRUCT, TRIDENT II SHELL, POLY LINER AND FEMORAL HEAD. REP CONFIRMED THERE ARE NO ALLEGATIONS AGAINST THE REVISED ADM/ MDM POLY INSERT OR THE FEMORAL HEAD.
Patient problems: Inflammation; Injury; Reaction
MAUDE report 9702658 ↗InjuryFeb 6, 2020Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
DR. REVISED A RIGHT TOTAL HIP THAT WAS ORIGINALLY DONE ON (B)(6) 2019. PATIENT HAD DISLOCATED MULTIPLE TIMES. DR. FELT THE CUP WAS PLACED TOO VERTICALLY AND THE STEM WAS UNDERSIZED. HE REVISED THE ENTIRE HIP TO A TRIDENT II MULTI-HOLE CUP, MDM AND OMNIFIT HFX PRESS FIT STEM. UPDATE 13/JANUARY/2020: SPOKE TO REP. SURGEON ALSO FELT THAT THE IMPLANTED STEM WAS UNDERSIZED FOR THE PATIENT'S ANATOMY AND WAS REVISED.
Patient problems: Injury; Joint Dislocation
MAUDE report 9678665 ↗InjuryDec 19, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT THE PATIENT'S LEFT HIP WAS REVISED DUE TO A FEMORAL PERIPROSTHETIC FRACTURE SUSTAINED IN A FALL. THE PATIENT'S HEMI HIP CONSTRUCT WAS REVISED TO A TOTAL HIP CONSTRUCT. REP CONFIRMED THERE ARE NO ALLEGATIONS AGAINST ANY OF THE REVISED IMPLANTS, PROVIDED THE PRIMARY AND REVISION USAGE SHEETS AND X-RAYS, AND CONFIRMED THAT NO FURTHER INFORMATION WILL BE AVAILABLE.
Patient problems: Bone Fracture(s); Fall; Hip Fracture; Injury
MAUDE report 9495277 ↗InjuryAug 23, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
PATIENT HAD A #9 SECURFIT MAX WITH A 48 +0 UNITRAX HEAD AND SLEEVE AND HAD A PERIPROSTHETIC FRACTURE. WE TOOK THE SECURFIT MAX STEM OUT, REDUCED THE FRACTURE, PLACED CABLES AROUND THE FEMUR, PUT IN A #8 OMNIFIT EON AND USED THE SAME HEAD AND SLEEVE THAT WE REMOVED. UPDATE: FRACTURE WAS AT PROXIMAL FEMUR. SURGEON REPORTED THE EXPLANTED SLEEVE AND HEAD HAD NO VISIBLE WEAR AND WERE RE-IMPLANTED IN THE PATIENT.
Patient problems: Bone Fracture(s); Injury
MAUDE report 7809905 ↗InjuryJun 28, 2018Loss of Osseointegration; Loss of Osseointegration
PT DISLOCATED WHILE WALKING. REVISION WAS DONE. ALL COMPONENTS REPLACED. REP. CONFIRMED DISLOCATION - HEAD DISLOCATED FROM LINER. REVISION CONSTRUCT INCLUDED AN ADM/MDM LINER CONSTRUCT. PER REP, REGARDING REVISION OF STEM AND CUP. "EVERYTHING WAS LOOSE AND NEEDED TO BE UPSIZED."
Patient problems: Inadequate Osseointegration; Injury; Joint Dislocation
MAUDE report 7645961 ↗InjuryJun 25, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT PATIENT'S LEFT HIP WAS REVISED DUE TO PAIN. REP WAS NOT MADE AWARE OF ANY LOOSE/ BROKEN DEVICES OR OTHER PATIENT FACTORS THAT COULD DIRECTLY BE ATTRIBUTED TO REPORTED PAIN, OR ITS SEVERITY. PATIENT WAS REVISED TO A RESTORATION MODULAR STEM AND AN UNKNOWN HEAD.
Patient problems: Injury; Pain
MAUDE report 7635281 ↗InjurySep 14, 2017Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT SURGEON PERFORMED A REVISION OF PATIENT'S LEFT HIP DUE TO PERIPROSTHETIC FRACTURE.
Patient problems: Bone Fracture(s); Injury
MAUDE report 6864948 ↗InjuryApr 10, 2017Fracture; Insufficient Device Problem Information; Insufficient Device Problem Information
SALES REP REPORTED A RIGHT TOTAL HIP REVISION, PATIENT FELL AND FRACTURED HIP. RESTORATION MODULAR STEM REPLACEMENT.
Patient problems: Bone Fracture(s); Fall; Hip Fracture; Injury
MAUDE report 6475681 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K041170, K051738. One submission often covers many models, so check the product and lot details in each record.
- Z-0893-2009 · Nov 13, 2008 · TerminatedSecur-Fit Max Hip Stem These devices are modular components of a total hip system. These femoral stems are intended for use with femoral heads and acetabular components in primary
- Z-0894-2009 · Nov 13, 2008 · TerminatedPrimary Secur-fit Plus Max Hip Stem These devices are modular components of a total hip system. These femoral stems are intended for use with femoral heads and acetabular componen
Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices
- BIOLOX DELTA 1365-32-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented551 reports
- PINNACLE ALTRX 1221-36-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented481 reports
- BIOLOX DELTA 1365-32-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death314 reports
- BIOLOX DELTA 1365-36-720Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death189 reports
- DJO SURGICAL 400-03-362Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented187 reports
- Anthology 71365705Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented176 reports
- BIOLOX DELTA 1365-40-710Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented170 reports
- PINNACLE GRIPTION 1217-32-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented169 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page