Skip to content

Trident 690-10-22FRecalls & reported problems

Howmedica Osteonics Corp. · Constrained Acet. Liner · Class 2 · Implant

All models of Trident (160)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07613327024357), received between Jun 21, 2021 and Feb 26, 2024 — classified as 3 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 30, 2026Open, Classified

    Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

    Z-3201-2026 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Dislodged or Dislocated3
  • Detachment of Device or Device Component1

Patient problems

  • Joint Dislocation2
  • Ambulation or Postural Difficulties1
  • Pain1
  • Unspecified Infection1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 26, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT'S LEFT HIP WAS REVISED DUE TO DISLOCATION OF A COMPETITOR HEAD FROM A STRYKER CONSTRAINED INSERT. THE INSERT WAS REVISED TO AN ADM/ MDM POLY INSERT WITH MDM METAL LINER. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Joint Dislocation

    MAUDE report 18784109 ↗
  • InjuryDec 2, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THE PATIENT SUFFERED A FRACTURE IN HIS RIGHT HIP IN 2012, AND HE UNDERWENT SURGERY WITH THE PLACEMENT OF A PROSTHESIS. HOWEVER, AROUND TWO YEARS LATER, IT WAS DIAGNOSED THAT THE PROSTHESIS WAS INFECTED WITH A FISTULA. A NEW SURGICAL PROCEDURE WAS PERFORMED TO REMOVE IT OF THE PROSTHESIS, PERFORMING SURGICAL DEBRIDEMENT AND PLACING A CEMENT SPACER WITH ANTIBIOTIC. IN A SECOND MOMENT, THE PATIENT UNDERWENT A NEW INTERVENTION TO REMOVE THE SPACER AND PLACE THE NEW PROSTHESIS, THE SURGERY BEING PERFORMED ON (B)(6)2014. AFTER SURGERY, WHEN PERFORMING A REVIEW CONSULTATION WITH THE PHYSICIAN, HE FOUND THROUGH AN EXAMINATION THAT THERE HAD BEEN DISLOCATION OF ARTHROPLASTY, THAT IS, DISPLACEMENT OF THE PROSTHETIC FEMORAL HEAD FROM ITS NORMAL PLACE WITH THE ACETABULAR COMPONENT. A NEW SURGICAL PROCEDURE WAS PERFORMED ON (B)(6) 2015 TO REMOVE THE PROSTHESIS. THE PATIENT ALLEGES A DEFECT IN THE ACETABULAR PROSTHESIS, WHICH ALLOWED THE DISPLACEMENT OF THE PROSTHETIC FEMORAL HEAD FROM ITS NORMAL PLACE. ADVERSE EVENTS REPORTED BY THE PATIENT: PAIN, INFECTION, SHORTENING OF THE RIGHT LEG, MOBILITY RESTRICTED TO WHEELCHAIR.

    Patient problems: Ambulation or Postural Difficulties; Pain; Unspecified Infection

    MAUDE report 15904373 ↗
  • InjuryJun 21, 2021Detachment of Device or Device Component; Device Dislodged or Dislocated; Detachment of Device or Device Component; Device Dislodged or Dislocated

    THE PATIENT UNDERWENT SURGERY ON (B)(6) 2021 FOR THE PLACEMENT OF STRYKER PRODUCTS (TORX & TRIDENT). ON (B)(6) 2021, SHE WAS ADMITTED TO THE HOSPITAL FOR A DISLOCATION. WHEN PERFORMING X-RAY, SHE OBSERVED STRYKER PRODUCT FAILURE. THE PATIENT UNDERGOES URGENT REVISION SURGERY.

    Patient problems: Joint Dislocation

    MAUDE report 12030830 ↗

Other Prosthesis, hip, constrained, cemented or uncemented, metal/polymer devices

All Prosthesis, hip, constrained, cemented or uncemented, metal/polymer devices →

Sources and updates