Trident 690-10-22FRecalls & reported problems
Howmedica Osteonics Corp. · Constrained Acet. Liner · Class 2 · Implant
All models of Trident (160)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07613327024357), received between Jun 21, 2021 and Feb 26, 2024 — classified as 3 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Jul 30, 2026Open, Classified
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Z-3201-2026 · Root cause: Process change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Dislodged or Dislocated3
- Detachment of Device or Device Component1
Patient problems
- Joint Dislocation2
- Ambulation or Postural Difficulties1
- Pain1
- Unspecified Infection1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryFeb 26, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENT'S LEFT HIP WAS REVISED DUE TO DISLOCATION OF A COMPETITOR HEAD FROM A STRYKER CONSTRAINED INSERT. THE INSERT WAS REVISED TO AN ADM/ MDM POLY INSERT WITH MDM METAL LINER. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Joint Dislocation
MAUDE report 18784109 ↗InjuryDec 2, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE PATIENT SUFFERED A FRACTURE IN HIS RIGHT HIP IN 2012, AND HE UNDERWENT SURGERY WITH THE PLACEMENT OF A PROSTHESIS. HOWEVER, AROUND TWO YEARS LATER, IT WAS DIAGNOSED THAT THE PROSTHESIS WAS INFECTED WITH A FISTULA. A NEW SURGICAL PROCEDURE WAS PERFORMED TO REMOVE IT OF THE PROSTHESIS, PERFORMING SURGICAL DEBRIDEMENT AND PLACING A CEMENT SPACER WITH ANTIBIOTIC. IN A SECOND MOMENT, THE PATIENT UNDERWENT A NEW INTERVENTION TO REMOVE THE SPACER AND PLACE THE NEW PROSTHESIS, THE SURGERY BEING PERFORMED ON (B)(6)2014. AFTER SURGERY, WHEN PERFORMING A REVIEW CONSULTATION WITH THE PHYSICIAN, HE FOUND THROUGH AN EXAMINATION THAT THERE HAD BEEN DISLOCATION OF ARTHROPLASTY, THAT IS, DISPLACEMENT OF THE PROSTHETIC FEMORAL HEAD FROM ITS NORMAL PLACE WITH THE ACETABULAR COMPONENT. A NEW SURGICAL PROCEDURE WAS PERFORMED ON (B)(6) 2015 TO REMOVE THE PROSTHESIS. THE PATIENT ALLEGES A DEFECT IN THE ACETABULAR PROSTHESIS, WHICH ALLOWED THE DISPLACEMENT OF THE PROSTHETIC FEMORAL HEAD FROM ITS NORMAL PLACE. ADVERSE EVENTS REPORTED BY THE PATIENT: PAIN, INFECTION, SHORTENING OF THE RIGHT LEG, MOBILITY RESTRICTED TO WHEELCHAIR.
Patient problems: Ambulation or Postural Difficulties; Pain; Unspecified Infection
MAUDE report 15904373 ↗InjuryJun 21, 2021Detachment of Device or Device Component; Device Dislodged or Dislocated; Detachment of Device or Device Component; Device Dislodged or Dislocated
THE PATIENT UNDERWENT SURGERY ON (B)(6) 2021 FOR THE PLACEMENT OF STRYKER PRODUCTS (TORX & TRIDENT). ON (B)(6) 2021, SHE WAS ADMITTED TO THE HOSPITAL FOR A DISLOCATION. WHEN PERFORMING X-RAY, SHE OBSERVED STRYKER PRODUCT FAILURE. THE PATIENT UNDERGOES URGENT REVISION SURGERY.
Patient problems: Joint Dislocation
MAUDE report 12030830 ↗
Other Prosthesis, hip, constrained, cemented or uncemented, metal/polymer devices
- Novation 134-36-45Exactech, Inc. · Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer2 recalls0 reports
- PINNACLE 1218-28-650Depuy Orthopaedics, Inc. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer89 reports
- Trident 690-00-28FHowmedica Osteonics Corp. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer1 recall37 reports
- Trident 690-00-22EHowmedica Osteonics Corp. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer1 recall36 reports
- PINNACLE 1218-32-652Depuy Orthopaedics, Inc. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer75 reports
- Trident 690-00-22DHowmedica Osteonics Corp. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer1 recall22 reports
- PINNACLE 1218-32-654Depuy Orthopaedics, Inc. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer70 reports
- Trident 69-2850Howmedica Osteonics Corp. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer1 recall11 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page