Skip to content

Trident 690-10-28KRecalls & reported problems

Howmedica Osteonics Corp. · Constrained Acet. Liner · Class 2 · Implant

All models of Trident (160)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jul 30, 2026Class IIOpen, Classified

    Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

    Z-3201-2026 · Root cause: Process change control · Recall details · FDA record ↗

Other Prosthesis, hip, constrained, cemented or uncemented, metal/polymer devices

All Prosthesis, hip, constrained, cemented or uncemented, metal/polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page