Trident 643-00-36FRecalls & reported problems
Howmedica Osteonics Corp. · Elevated Rim Insert · Class 2 · Implant
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 07613327025101), received between Apr 5, 2018 and Mar 11, 2024 — classified as 8 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (4 reports).
2 other recalls cover devices cleared under the same FDA submission (K033716); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated4
- Insufficient Device Problem Information3
- Adverse Event Without Identified Device or Use Problem1
- Degraded1
- Unstable1
Patient problems
- Injury4
- Joint Dislocation4
- Unspecified Infection2
- Joint Laxity1
- Scar Tissue1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryMar 11, 2024Unstable; Unstable
TOTAL RIGHT HIP REVISION. SURGEON PERFORMED POLY LINER AND FEMORAL HEAD SWAP, TO ADDRESS STABILITY ISSUE IN HIP JOINT.
Patient problems: Joint Laxity
MAUDE report 18872817 ↗InjuryOct 20, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE CLINICAL TEAM PROVIDED AN OPERATIVE REPORT AND INTRA-OPERATIVE RECORD FOR A REVISION OF THE PATIENT'S RIGHT HIP DUE TO RECURRENT DISLOCATIONS. OP REPORT ALSO STATES THAT THE STEM HAD "MINIMAL ANTEVERSION". A STEM (IMPLANTED IN (B)(6) 2021), AN ELEVATED RIM INSERT AND +7.5 CERAMIC HEAD (IMPLANTED IN (B)(6) 2021) WERE REVISED TO A RESTORATION MODULAR STEM CONSTRUCT, +8 FEMORAL HEAD, ADM/ MDM POLY INSERT, AND MDM METAL LINER.
Patient problems: Joint Dislocation; Scar Tissue
MAUDE report 12663145 ↗InjurySep 9, 2021Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT THE PATIENT'S LEFT HIP WAS REVISED DUE TO INFECTION. A HEAD AND LINER EXCHANGE WAS PERFORMED. REP CAN PROVIDE THE PRIMARY AND REVISION USAGE SHEETS, AND CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Unspecified Infection
MAUDE report 12446176 ↗InjurySep 8, 2021Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
WHILE COVERING A REVISION OF THE PATIENT'S LEFT HIP ON (B)(6) 2021, THE REP WAS TOLD THE SURGEON PERFORMED A PRIOR I&D ON THE SAME HIP DUE TO SUSPECTED INFECTION. NO IMPLANTS WERE REVISED AND NO STRYKER REP WAS PRESENT FOR THE PROCEDURE. REP CAN PROVIDE THE PRIMARY USAGE SHEET AND CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Unspecified Infection
MAUDE report 12442347 ↗InjuryMar 30, 2020Degraded; Insufficient Device Problem Information; Degraded; Insufficient Device Problem Information
IT WAS REPORTED THAT A REVISION SURGERY IS SCHEDULED FOR THE PATIENT'S RIGHT HIP ON (B)(6) 2020 DUE TO THE HEAD AND PART OF THE STEM BREAKING OFF FROM THE REST OF THE STEM. UPDATE: "SURGEON REVISED LINER, DO TO WEAR AND TIME FRAME THE CUP HAD BEEN IN PATIENT."
Patient problems: Injury
MAUDE report 9898067 ↗InjuryJan 20, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENT'S LEFT HIP WAS REVISED DUE TO DISLOCATION OF THE 36 +0 METAL HEAD FROM THE ELEVATED RIM POLY LINER. SURGEON REPORTED THAT THERE ARE NO ALLEGATIONS AGAINST THE IMPLANTS AND REPORTED A PRE-EXISTING CONTRIBUTING FACTOR: PATIENT HAD AN AMPUTATION WITH PROSTHESIS OF THEIR RIGHT SIDE, RESULTING IN A LEG LENGTH 1" SHORTER THAN THE LEFT LEG. THE HEAD AND LINER WERE REVISED TO AN ADM/MDM LINER CONSTRUCT AND BIOLOX CERAMIC HEAD WITH SLEEVE. REP PROVIDED PRIMARY AND REVISION USAGE REPORTS, AND REPORTED THAT NO FURTHER INFORMATION WILL BE AVAILABLE.
Patient problems: Injury; Joint Dislocation
MAUDE report 9608230 ↗InjuryJan 17, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENT'S LEFT HIP WAS REVISED 2 DAYS POST IMPLANT DUE TO DISLOCATION OF THE HEAD FROM THE POLY LINER IN PHYSICAL THERAPY. PATIENT WAS REVISED FROM A 36 +5 BIOLOX HEAD AND TRIDENT X3 POLY LINER TO A 36 +7.5 BIOLOX HEAD AND SAME SIZE LINER. REP PROVIDED PRIMARY AND REVISION USAGE INFORMATION, AND REPORTED THAT NO FURTHER INFORMATION WILL BE AVAILABLE.
Patient problems: Injury; Joint Dislocation
MAUDE report 9603196 ↗InjuryApr 5, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated
PATIENT HAD A PRIMARY TOTAL HIP REPLACEMENT (B)(6) 2018. HIS POST-OP PERIOD WAS UNREMARKABLE. HE WAS DOING WELL AND EXCEEDED WHAT HE WAS SUPPOSED TO BE DOING. ON (B)(6) 2018 HE BENT OVER AT THE HIP TO PICK UP A LOG. HE DISLOCATED AND WAS BROUGHT THE EMERGENCY ROOM (E.R.) HE WAS REDUCE AND DISLOCATED AGAIN IN THE E.R. COMPONENTS WERE WELL POSITIONED. THE DOCTOR FELT THE PATIENT WOULD BENEFIT FROM A REVISION TO AN MDM. HE WAS REVISED ON (B)(6).
Patient problems: Injury; Joint Dislocation
MAUDE report 7400222 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K033716. One submission often covers many models, so check the product and lot details in each record.
- Z-2810-2015 · Jun 25, 2015 · TerminatedTrident 10 degrees X3 Insert 36 mm ID Catalog #623-10-36I Hip implant component
- Z-0909-2010 · Feb 17, 2006 · TerminatedTrident Acetabular Hip System Polyethylene Inserts, Stryker Orthopaedics, multiple sizes, Mahwah, NJ. Intended for use in cementless fixation fro use in total hip arthroplasty to
Other Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate devices
- Novation 130-32-51Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls156 reports
- Novation 130-36-53Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls79 reports
- CORAIL 3L92511Depuy (Ireland) · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate158 reports
- Acumatch A Series 138-36-29Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate3 recalls1 report
- Novation 136-32-51Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls43 reports
- Novation 130-32-53Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls33 reports
- Novation 132-32-51Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls26 reports
- Novation 136-36-53Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls22 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.