Trident 509-02-58FRecalls & reported problems
Howmedica Osteonics Corp. · Hemispherical Shell · Class 2 · Implant
All models of Trident (160)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 07613327025170), received between Mar 24, 2017 and Jun 20, 2022 — classified as 8 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated3
- Adverse Event Without Identified Device or Use Problem2
- Loss of Osseointegration2
- Insufficient Device Problem Information1
- Malposition of Device1
Patient problems
- Injury7
- Fall2
- Ambulation or Postural Difficulties1
- Bone Fracture(s)1
- Inadequate Osseointegration1
- Joint Dislocation1
- Pain1
- Unspecified Infection1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryJun 20, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
REVISED A CUP AND LINER DUE TO PAIN THE PATIENT HAD. THE ORIGINAL SURGERY WAS (B)(6) 2016 WITH THE ORIGINAL TRITANIUM CUP. IT WAS NOTED ON THE X-RAY THERE APPEARED TO BE A HALO. UPON EXTRACTION SURGEON NOTED THERE WAS MOSTLY FIBROUS INGROWTH. ORIGINAL STEM WAS A MINI STEM.
Patient problems: Pain
MAUDE report 14749420 ↗InjurySep 8, 2020Loss of Osseointegration; Loss of Osseointegration
IT WAS REPORTED THAT THE PATIENT'S RIGHT HIP WAS REVISED DUE TO LOOSENING. A SHELL, 3 SCREWS, AND A LINER WERE REVISED TO A SHELL, 6 SCREWS AND A LINER. USAGE SHEETS FOR THE CURRENT AND PREVIOUS SURGERIES WERE PROVIDED. REP CONFIRMED THERE ARE NO ALLEGATIONS AGAINST THE REVISED LINER, AND THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Injury
MAUDE report 10501231 ↗InjuryJan 28, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENT'S LEFT HIP WAS REVISED DUE TO DISLOCATION OF THE ADM/MDM POLY INSERT FROM THE MDM METAL LINER. THE SURGEON ALSO WANTED TO MEDIALIZE THE ACETABULAR SHELL. THE MDM/ ADM LINER CONSTRUCT, CERAMIC HEAD, SLEEVE (IMPLANTED (B)(6) 2019) AND ACETABULAR SHELL WITH 5 SCREWS (IMPLANTED (B)(6) 2019) WERE REVISED TO A LARGER SHELL, REIMPLANTATION OF THE 5 ORIGINAL SCREWS (REP NOTIFIED SURGEON THAT THIS WAS OFF-LABEL USE), AND FEMORAL HEAD WITH CONSTRAINED LINER. REP CONFIRMED THERE ARE NO ALLEGATIONS AGAINST THE REVISED HEAD AND SLEEVE, AND CONFIRMED THAT NO FURTHER INFORMATION WILL BE AVAILABLE.
Patient problems: Injury
MAUDE report 9639815 ↗InjuryOct 3, 2018Loss of Osseointegration; Loss of Osseointegration
IT WAS REPORTED THAT PATIENT'S RIGHT HIP WAS REVISED DUE TO SHELL LOOSENING. A 58MM HEMISPHERICAL SHELL, 4 SCREWS, AND A 40F POLY LINER WERE REVISED TO A 62MM HEMISPHERICAL SHELL WITH 5 SCREWS AND A 40G POLY LINER. THE HEAD AND STEM WERE AND REMAIN COMPETITOR DEVICES.
Patient problems: Inadequate Osseointegration; Injury
MAUDE report 7930374 ↗InjuryMay 10, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THROUGH THE SUBMISSION OF A REVISION IMPLANT SHEET THAT PATIENT'S LEFT HIP WAS REVISED. 5 RESTORATION GAP PLATE SCREWS AND A TRIDENT INSERT WERE IMPLANTED. REVISION WAS FOR PERIPROSTHETIC FRACTURE OF ACETABULUM.
Patient problems: Bone Fracture(s); Injury
MAUDE report 7502587 ↗InjuryJul 24, 2017Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem; Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED PATIENT'S LEFT HIP WAS REVISED DUE TO DISLOCATED CONSTRAINED LINER. PATIENT FELL TWO TIMES THAT CAUSED THE DISLOCATED LINER. THE SHELL AND SCREWS WERE ALL REVISED.
Patient problems: Fall; Injury
MAUDE report 6736728 ↗InjuryMay 26, 2017Device Dislodged or Dislocated; Device Dislodged or Dislocated
PATIENT UNDERWENT A REVISION OF A RIGHT COMPETITOR ACETABULAR COMPONENT ON (B)(6) 2017. PROCEDURE WAS COMPLETED SUCCESSFULLY AND THE POST OP X-RAYS LOOKED GOOD. THE PATIENT SUBSEQUENTLY FELL OUT OF BED . HE WENT TO THE EMERGENCY ROOM. X-RAYS SHOWED THAT HE WAS DISLOCATED AND THE POSITION OF HIS CUP HAD CHANGED FROM THE POST OP FILMS. HE WAS SCHEDULED FOR A REVISION ON (B)(6). THE OPERATION WAS COMPLETED SUCCESSFULLY.
Patient problems: Fall; Injury; Joint Dislocation
MAUDE report 6594790 ↗InjuryMar 24, 2017Malposition of Device; Malposition of Device
REVISION OF A TOTAL HIP DUE TO ACUTE INFECTION AND MALPOSITION OF THE ACETABULAR CUP.
Patient problems: Ambulation or Postural Difficulties; Injury; Unspecified Infection
MAUDE report 6433087 ↗
Other Prosthesis, hip, semi-constrained, uncemented, metal/polymer, porous devices
- Accolade 6021-0537Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, porous46 reports
- PROFEMUR® PHAC1252Microport Orthopedics Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Porous37 reports
- Accolade 6020-0335Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, porous35 reports
- OXINIUM 71342340Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, porous32 reports
- EXETER 0580-1-351Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, porous29 reports
- INSTRUMENT 2101-0200Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, porous27 reports
- REFLECTION 71330001Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, porous20 reports
- Accolade 6720-0330Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, porous18 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page