The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613327025293), received between Mar 27, 2019 and Feb 19, 2021 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Material Deformation” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Deformation1
- Use of Device Problem1
Patient problems
- Inadequate Osseointegration1
- Injury1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 19, 2021Material Deformation; Material Deformation
THE FOLLOWING WAS REPORTED: X3 LINER WAS IMPACTED AS USUAL. WHEN THE SURGEON CHECKED STABILITY WITH ELEVA, IT MADE A SLIGHT BENDING MOVEMENT ON ONE SIDE. SURGEON CHECKED AGAIN, BUT IT DIDN'T CHANGE. IN THE END, IT DIDN'T COME OFF, SO PROCEDURE WAS COMPLETED. WHEN SUCH AN EVENT OCCURS, WHAT COULD BE HAPPENED? ARE THERE ANY SIMILAR EVENT ?
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11347963 ↗InjuryMar 27, 2019Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT THE PATIENT'S RIGHT HIP WAS REVISED. AS REPORTED BY REP: "PREOPERATIVE DIAGNOSIS: RIGHT FAILED TOTAL HIP ARTHROPLASTY, COMPLICATION OF PROSTHETIC HIP IMPLANT¿[SURGEON] COMMENTED NO BONY INGROWTH TO CUP, ONLY FIBROUS¿" A SHELL, SCREW, INSERT, AND HEAD WERE REVISED. ON 27-FEB-2019: MEDICAL REVIEW WAS RECEIVED WHICH IDENTIFIED BONY IMPINGEMENT BETWEEN ACETABULUM AND FEMORAL STEM, ACETABULAR COMPONENT MALPOSITION AND INAPPROPRIATE SELECTION REGARDING THE TRIDENT LINER
Patient problems: Inadequate Osseointegration; Injury
MAUDE report 8457162 ↗
Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices
- BIOLOX DELTA 1365-32-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented551 reports
- PINNACLE ALTRX 1221-36-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented481 reports
- BIOLOX DELTA 1365-32-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death314 reports
- BIOLOX DELTA 1365-36-720Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death189 reports
- DJO SURGICAL 400-03-362Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented187 reports
- Anthology 71365705Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented176 reports
- BIOLOX DELTA 1365-40-710Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented170 reports
- PINNACLE GRIPTION 1217-32-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented169 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗