Trident 69-2852Recalls & reported problems
Howmedica Osteonics Corp. · Constrained Acetabular Insert · Class 2 · Implant
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613327025484), received between Oct 28, 2024 and Jun 10, 2026 — classified as 2 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Jul 30, 2026Open, Classified
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Z-3201-2026 · Root cause: Process change control · FDA record ↗
Most reported problems
Device problems
- Device Dislodged or Dislocated1
- Fracture1
Patient problems
- Ambulation or Postural Difficulties1
- Joint Dislocation1
- Pain1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJun 10, 2026Device Dislodged or Dislocated
THE FOLLOWING WAS REPORTED: THE FIRST SURGERY WAS DONE ON (B)(6) 2026. PATIENT HAS COME BACK WITH DISLOCATION OF HIP AND REVISION WAS DONE. LEFT SIDE.
Patient problems: Joint Dislocation
MAUDE report 25455749 ↗InjuryOct 28, 2024Fracture; Fracture
TRIDENT ALL POLY CONSTRAINED ACETABULAR INSERT HAS BEEN BROKEN WHILE IMPLANTED IN PATIENTPATIENT CAME TO THE SURGEON AFTER 2 MONTHS OF THE THR REVISED SURGERY WITH COMPLAINT OF THE HIP PAIN. SURGEON GOT HIS HIP XRAY DONE AND FOUND THAT TRIDENT CONSTRAINED ACETABULAR INSERT WAS BROKEN.
Patient problems: Ambulation or Postural Difficulties; Pain
MAUDE report 20553251 ↗
Other Prosthesis, hip, constrained, cemented or uncemented, metal/polymer devices
- Novation 134-36-45Exactech, Inc. · Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer2 recalls0 reports
- PINNACLE 1218-28-650Depuy Orthopaedics, Inc. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer89 reports
- Trident 690-00-28FHowmedica Osteonics Corp. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer1 recall37 reports
- Trident 690-00-22EHowmedica Osteonics Corp. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer1 recall36 reports
- PINNACLE 1218-32-652Depuy Orthopaedics, Inc. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer75 reports
- Trident 690-00-22DHowmedica Osteonics Corp. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer1 recall22 reports
- PINNACLE 1218-32-654Depuy Orthopaedics, Inc. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer70 reports
- Trident 69-2850Howmedica Osteonics Corp. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer1 recall11 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.