Skip to content

Acetabular Insert 2099-2250Recalls & reported problems

Howmedica Osteonics Corp. · Acetabular Insert · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613327031430), received between Aug 15, 2019 and Apr 26, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 30, 2026Open, Classified

    Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

    Z-3201-2026 · Root cause: Process change control · FDA record ↗

Most reported problems

Device problems

  • Device Dislodged or Dislocated2
  • Break1

Patient problems

  • Injury1
  • Insufficient Information1
  • Joint Dislocation1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 26, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT'S RIGHT HIP WAS REVISED DUE TO DISSOCIATION OF THE INNER BIPOLAR COMPONENT FROM THE OUTER COMPONENT OF A CONSTRAINED LINER. THE LINER AND FEMORAL HEAD WERE REVISED TO ANOTHER CONSTRAINED LINER AND FEMORAL HEAD. THERE ARE NO ALLEGATIONS AGAINST THE REVISED HEAD. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Insufficient Information; Joint Dislocation

    MAUDE report 19195550 ↗
  • InjuryAug 15, 2019Break; Device Dislodged or Dislocated; Break; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT'S RIGHT HIP WAS REVISED DUE TO DISASSOCIATION OF THE CONSTRAINED LINER OUT OF THE ACETABULAR SHELL. INTRA-OPERATIVELY, THE METAL RING OF THE CONSTRAINED LINER WAS REPORTED AS BROKEN ("SNAPPED"), AND THE INNER EDGE OF THE SHELL WAS REPORTED AS BEING ASSYMETRIC DUE TO WEAR. A 54MM SERIES II ACETABULAR SHELL, 50MM HOWMEDICA OSTEONICS CONSTRAINED ACETABULAR INSERT, AND STRYKER FEMORAL HEAD WERE REVISED TO A TRIDENT II SHELL WITH 6 SCREWS, A TRIDENT CONSTRAINED LINER, AND ANOTHER STRYKER FEMORAL HEAD. REP CONFIRMED THERE WERE NO ALLEGATIONS AGAINST THE FEMORAL HEAD. THE EXPLANTS ARE AVAILABLE FOR RETURN. OTHERWISE, REP CONFIRMED THAT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Injury

    MAUDE report 8897456 ↗

Other Prosthesis, hip, constrained, cemented or uncemented, metal/polymer devices

All Prosthesis, hip, constrained, cemented or uncemented, metal/polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page