Acetabular Insert 2099-2861Recalls & reported problems
Howmedica Osteonics Corp. · Acetabular Insert · Class 2 · Implant
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613327031591), received between Nov 21, 2023 and Nov 21, 2023 — classified as 1 injury.
The most frequently reported specific device problem is “Naturally Worn” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Jul 30, 2026Open, Classified
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Z-3201-2026 · Root cause: Process change control · FDA record ↗
Most reported problems
Device problems
- Naturally Worn1
- Unstable1
Patient problems
- Ambulation or Postural Difficulties1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryNov 21, 2023Unstable; Naturally Worn; Unstable; Naturally Worn
IT WAS REPORTED THROUGH THE SUBMISSION OF A REVISION IMPLANT SHEET THAT THE PATIENT'S HIP WAS REVISED. COMPARING WITH THE USAGE SHEET FROM A PRIOR REVISION, A 28 +5 FEMORAL HEAD AND CONSTRAINED INSERT WERE REVISED TO A 28 +10 FEMORAL HEAD, CONSTRAINED INSERT, AND 2 ADDITIONAL SCREWS. ADDITIONAL INFORMATION AS PER SALES REP, "RECURRING INSTABILITY AND IMPLANT WEAR. HEAD AND LINER WAS EXCHANGED A MONTH PRIOR AND GOT INFECTED. WENT BACK IN TO REPLACE HEAD AND LINER AGAIN."
Patient problems: Ambulation or Postural Difficulties
MAUDE report 18180094 ↗
Other Prosthesis, hip, constrained, cemented or uncemented, metal/polymer devices
- Novation 134-36-45Exactech, Inc. · Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer2 recalls0 reports
- PINNACLE 1218-28-650Depuy Orthopaedics, Inc. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer89 reports
- Trident 690-00-28FHowmedica Osteonics Corp. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer1 recall37 reports
- Trident 690-00-22EHowmedica Osteonics Corp. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer1 recall36 reports
- PINNACLE 1218-32-652Depuy Orthopaedics, Inc. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer75 reports
- Trident 690-00-22DHowmedica Osteonics Corp. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer1 recall22 reports
- PINNACLE 1218-32-654Depuy Orthopaedics, Inc. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer70 reports
- Trident 69-2850Howmedica Osteonics Corp. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer1 recall11 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.