GMRS 6495-5-011Recalls & reported problems
Howmedica Osteonics Corp. · Straight Fluted Press-Fit Stem · Class 2 · Implant
All models of GMRS (32)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07613327033489), received between Mar 4, 2019 and Jun 3, 2025 — classified as 3 injuries.
The most frequently reported specific device problem is “Fracture” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K022403, K023087, K031217, K222056, K241716); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture2
- Degraded1
- Insufficient Device Problem Information1
Patient problems
- Ambulation or Postural Difficulties3
- Insufficient Information2
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJun 3, 2025Degraded; Fracture; Degraded; Fracture
THIS PI IS FOR THE REVISION SURGERY PERFORMED ON (B)(6) 2008. PATIENT UNDERWENT RESECTION SURGERY IN 1989 PROXIMAL RIGHT TIBIA AND RECONSTRUCTION WITH HMRS MODULAR PROSTHESIS STRYKER FOR OSTEOSARCOMA. ON B)(6) 2008 SURGERY FOR REVISION OF THE TIBIAL PROSTHETIC COMPONENT DUE TO STEM RUPTURE. ON THAT OCCASION, THE PATIENT REPLACED THE HMRS TIBIAL STEM WITH GMRS STRYKER TIBIAL STEM (REF 6495-5-011, LOT 7W4074A) AND FITTED NEW TIBIAL PROSTHETIC BODY BY IMPLANTING HMRS 120MM TIBIAL PROSTHETIC COROPE, GMRS TIBIAL PROSTHETIC TENSE ADAPTER, AND GMRS 50MM TIBIAL SPACER (NO REPLACEMENT OF FEMORAL COMPONENT WHICH WAS INTACT). ON (B)(6) 2016 SURGERY REPLACEMENT OF RIGHT KNEE PROSTHETIC JOINT COMPONENTS DUE TO WEAR AND TEAR. ON (B)(6) 2016 SURGICAL CLEANUP FOR INFECTION PERI-PROSTHETIC TIBIAL, RETAINING TIBIAL STEM, AND IMPLANTATION OF ANTIBIOTIC CEMENT SPACER. ON (B)(6) 2017 ALSO EXPLANTATION OF FEMORAL PROSTHETIC COMPONENT AND FURTHER SURGICAL CLEANING WITH CEMENT SPACER RENEWAL. ON 24/MAR/2017 AND 19/MAY/2017 RENEWAL CEMENT SPACER. ON (B)(6) 2017 IMPLANTATION OF MODULAR TIBIA PROSTHESIS PROXIMAL DX IN KNEE ARTHRODESIS ALSO USING COMPONENT CUSTOM GMRS EXTENSION PIECE (REF C-M106-2-000; LOT TA
Patient problems: Ambulation or Postural Difficulties; Insufficient Information
MAUDE report 22131464 ↗InjuryOct 17, 2023Fracture; Fracture
PATIENT HAD A REVISION FOR BROKEN GMRS 11 X 125 PRESS FIT FEMUR STEM ON RIGHT LEG. REVISION WAS PERFORMED DUE TO THE STEM BREAKING WHERE THE 11MM DIAMETER BEGINS TAPERING TO A WIDER DIAMETER. THE STEM, AN EXTENDER, FEMORAL COMPONENT, AND ROTATING COMPONENT WERE REVISED.
Patient problems: Ambulation or Postural Difficulties; Insufficient Information
MAUDE report 17948184 ↗InjuryMar 4, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT, FOR A PATIENT WHO WAS OPERATED SEVERAL TIMES BY THE SAME SURGEON, THE LAST SURGERY WAS THE SECOND FAILED ATTEMPT TO EXCHANGE THE GROWING PIECE OF HIS GMRS IMPLANT (HMRS/GMRS SPIGOT SEPARATOR). A LONGER GROWING PIECE IS NEEDED, GROWING PIECE THE PATIENT CURRENTLY HAS FAILED/LOST IT¿S ABILITY TO KEEP IT¿S FULL LENGTH. IT IS SUSPECTED THAT SEPARATION OF THE STEM/GROWING PIECE SPIGOT FAILED BECAUSE OF COLD FUSION. THE PATIENT ((B)(6) YEAR OLD BOY) HAS 6CM DIFFERENCE IN LEG LENGTH, THE ORIGINAL GROWING COMPONENT IS SET TO IT¿S MAXIMUM LENGTH BUT AS IT¿S COLLAPSED ONCE THE PATIENT¿S PARENTS LOST THEIR TRUST IN THE DEVICE. IT IS NOT CLEAR WHY/HOW THE DEVICE LOST IT¿S INTERNAL BLOCKAGE, BUT THE SURGEON FOUND IT COLLAPSED ( IN THE SHORTEST POSITION INSTEAD OF THE LAST SET ELONGATION ) IN THE OPERATION WHEN THEY WANTED TO DO THE LAST ADJUSTMENT ( THE LAST 5MM ) BEFORE CHANGING TO THE LONGER PIECE. AFTER THEY WENT THROUGH THE ELONGATION STEPS AGAIN - 2-3 SMALL SURGERIES - THEY REACHED FULL LENGTH AGAIN, THE DEVICE SHOWED NO SIGNS OF FAILURE/INSTABILITY.
Patient problems: Ambulation or Postural Difficulties
MAUDE report 8386838 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K022403, K023087, K031217, K222056, K241716. One submission often covers many models, so check the product and lot details in each record.
- Z-2732-2024 · Aug 5, 2024 · Open, ClassifiedCustom Made Device - GMRS EXTENSION Piece 40MM LEFT, 35 Deg EXT Rotated Tabs (Patient Specific); Catalog Number: C-MM100434-00;
- Z-2624-2010 · May 5, 2005 · TerminatedGMRS Extension Piece, 60 mm; Sterile Stryker Howmedica Osteonics; Howmedica Internation S. de. R.L. Limerick, Ireland. The Global Modular Replacement System is intended to be used
Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices
- BIOLOX DELTA 1365-32-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented551 reports
- PINNACLE ALTRX 1221-36-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented481 reports
- BIOLOX DELTA 1365-32-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death314 reports
- BIOLOX DELTA 1365-36-720Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death189 reports
- DJO SURGICAL 400-03-362Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented187 reports
- Anthology 71365705Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented176 reports
- BIOLOX DELTA 1365-40-710Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented170 reports
- PINNACLE GRIPTION 1217-32-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented169 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page