Skip to content

Patella 3044-0028Recalls & reported problems

Howmedica Osteonics Corp. · Circular - All Polyethylene · Class 2 · Implant

All models of Patella (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07613327033793), received between Jun 11, 2019 and Oct 13, 2023 — classified as 3 injuries.

The most frequently reported specific device problem is “Malposition of Device” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (K863668); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 1, 2023Open, Classified

    Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.

    Z-0870-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Malposition of Device2
  • Unstable1

Patient problems

  • Pain2
  • Synovitis2
  • Ambulation or Postural Difficulties1
  • Injury1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 13, 2023Malposition of Device; Malposition of Device

    THE PATIENT UNDERWENT ARTHROSCOPIC EXAM, DEBRIDEMENT OF THE SYNOVIUM, AND LATERAL RELEASE TO ADDRESS PATELLAR IMPINGEMENT. NO DEVICES WERE REVISED.

    Patient problems: Synovitis

    MAUDE report 17928591 ↗
  • InjuryMay 28, 2021Malposition of Device; Malposition of Device

    IN PERFORMING FOLLOW-UP FOR A PI, REP PROVIDED A REVISION OPERATIVE REPORT FOR (B)(6) 2013 FOR THE PATIENT'S RIGHT KNEE. PREOPERATIVE DIAGNOSIS: LOOSE RIGHT REVISION KNEE ARTHROPLASTY WITH MALTRACKING OF PATELLA. OP REPORT NOTES: "IN LESS THAN 1 YEAR SHE DEVELOPED INCREASING PAIN AND DISCOMFORT. STUDIES ARE SUSPICIOUS FOR POSSIBLE LOOSENING OF FEMORAL COMPONENT. THERE WAS ONLY SOME DISTAL CEMENT AROUND THE FEMORAL COMPONENT. THE STEM APPEARED TO BE REACTIVE LINE AROUND IT SUGGESTING LOOSENING OF FEMORAL COMPONENT...THE PATELLA WAS ALSO TILTING AND MALTRACKING¿FEMORAL COMPONENT CAME OUT VERY EASILY WITH NO FIXATION AT ALL. THE CEMENT WAS BONDED TO THE IMPLANT WITH AUGMENTS. THE STEM CAME OUT AND THERE WAS A FIBROUS MEMBRANE AROUND THE STEM COMPLETELY. THERE WAS NO CEMENT ON THE STEM AT ALL CONSISTENT WITH FEMORAL COMPONENT LOOSENING...WE HAD TO REMOVE THE PATELLAR COMPONENT TO POSITION IT MORE CENTRALLY..."

    Patient problems: Ambulation or Postural Difficulties; Pain

    MAUDE report 11900264 ↗
  • InjuryJun 11, 2019Unstable; Unstable

    IT WAS REPORTED THROUGH THE COMMUNICATION OF AN ATTORNEY THAT ALLEGEDLY THE PATIENT WAS REVISED DUE TO "PAINFUL RIGHT TOTAL KNEE ARTHROPLASTY 2008 SECONDARY TO FLEXION INSTABILITY AND LOOSENING OF THE TIBIAL COMPONENT. SEVERE POLYETHYLENE WEAR LATERAL TIBIAL PLATEAU OF THE ARTICULAR COMPONENT WITH ASSOCIATED SYNOVITIS".

    Patient problems: Injury; Pain; Synovitis

    MAUDE report 8686457 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K863668. One submission often covers many models, so check the product and lot details in each record.

  • Z-0869-2024 · Dec 1, 2023 · Open, Classified
    SCORPIO RECESSED PATELLA- Intended for Knee Replacement Part Number: 3044-0026
  • Z-0871-2024 · Dec 1, 2023 · Open, Classified
    SCORPIO RECESSED PATELLA- Intended for knee replacement Part Number: 3044-0030
  • Z-0872-2024 · Dec 1, 2023 · Open, Classified
    SCORPIO RECESSED PATELLA-Intended for knee repalcment Part Number: 3044-0032

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page