Skip to content

Scorpio 73-3708Recalls & reported problems

Howmedica Osteonics Corp. · Universal Dome Patella · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07613327033892), received between Nov 8, 2023 and Nov 8, 2023 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Material Discolored” (3 reports).

2 FDA recall records name this device's identifier, 2 still open.

7 other recalls cover devices cleared under the same FDA submission (K051948, K052917, K243817, K252898, K972967); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 30, 2026Open, Classified

    Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

    Z-3206-2026 · Root cause: Process change control · FDA record ↗

  2. Dec 1, 2023Open, Classified

    Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.

    Z-0896-2024 · Root cause: Nonconforming Material/Component · FDA record ↗

Most reported problems

Device problems

  • Material Discolored3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 8, 2023Material Discolored; Material Discolored

    73-3708 LOT JR5N CAME OUT OF THE BOX YELLOW

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18097331 ↗
  • MalfunctionNov 8, 2023Material Discolored; Material Discolored

    73-3708 LOT JR5N CAME OUT OF THE BOX YELLOW.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18097330 ↗
  • MalfunctionNov 8, 2023Material Discolored; Material Discolored

    OPENED STERILE PACKAGED SCORPIO UNIVERSAL PATELLA AND PRODUCT WAS SLIGHTLY DISCOLORED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18096717 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K051948, K052917, K243817, K252898, K972967. One submission often covers many models, so check the product and lot details in each record.

  • Z-3197-2026 · Jul 30, 2026 · Open, Classified
    ALL POLY CONSTRAINED INSERT 54MM 69-2854 NRG KNEE CR 82-2-0512 82-2-0518 82-2-0912 NRG KNEE P/S 82-3-0318 SCORPIO + MOBILE BEARING PCS 78-5-0910 SCORPIO C-DOME PAT
  • Z-0895-2024 · Dec 1, 2023 · Open, Classified
    SCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3508
  • Z-0894-2024 · Dec 1, 2023 · Open, Classified
    SCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3308
  • Z-0900-2024 · Dec 1, 2023 · Open, Classified
    Scorpio NRG Tibial Brg Insert Assy -Intended for knee replacement Part Number: 82-2-0910
  • Z-0897-2024 · Dec 1, 2023 · Open, Classified
    SCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3710
  • Z-0893-2024 · Dec 1, 2023 · Open, Classified
    SCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3110
  • Z-0606-2012 · Aug 17, 2011 · Terminated
    Eius Unicompartmental Knee System Stryker Howmedica Osteonics Corp. Howmedica International S. de R.L., Raheen Business Park, Limerick, Ireland. Intended for use in Unicompartment

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page