Skip to content

Scorpio 72-4-7518Recalls & reported problems

Howmedica Osteonics Corp. · Total Stabilizer Tibial Insert · Class 2 · Implant

All models of Scorpio (107)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

9 other recalls cover devices cleared under the same FDA submission (K243817, K994128); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Dec 1, 2023Open, Classified

    Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.

    Z-0880-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K243817, K994128. One submission often covers many models, so check the product and lot details in each record.

  • Z-3206-2026 · Jul 30, 2026 · Open, Classified
    SCORPIO C-DOME PATELLA 73-2710 73-2910 SCORPIO U-DOME PATELLA 73-3110 73-3508 73-3708 73-3710 73-3910 Orthopaedic implant components manufactured from UHMWPE (polyet
  • Z-3197-2026 · Jul 30, 2026 · Open, Classified
    ALL POLY CONSTRAINED INSERT 54MM 69-2854 NRG KNEE CR 82-2-0512 82-2-0518 82-2-0912 NRG KNEE P/S 82-3-0318 SCORPIO + MOBILE BEARING PCS 78-5-0910 SCORPIO C-DOME PAT
  • Z-0882-2024 · Dec 1, 2023 · Open, Classified
    SCORPIO TS TIB INSERT -Intended for knee replacement Part Number: 72-4-7524
  • Z-0877-2024 · Dec 1, 2023 · Open, Classified
    SCORPIO TS TIB INSERT-Intended for knee replacement Part Number:72-4-7512
  • Z-0875-2024 · Dec 1, 2023 · Open, Classified
    SCORPIO TS TIB INSERT -Intended for knee replacement Part Number: 72-4-0321
  • Z-0881-2024 · Dec 1, 2023 · Open, Classified
    SCORPIO TS TIB INSERT -Intended for knee replacement Part Number: 72-4-7521
  • Z-0876-2024 · Dec 1, 2023 · Open, Classified
    SCORPIO TS TIB INSERT -Intended for knee replacement Part Number: 72-4-0510
  • Z-0879-2024 · Dec 1, 2023 · Open, Classified
    SCORPIO TS TIB INSERT -Intended for knee replacment Part Number: 72-4-7516
  • Z-0878-2024 · Dec 1, 2023 · Open, Classified
    SCORPIO TS TIB INSERT -Intended for knee replacement Part Number: 72-4-7514

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page