Trident 623-00-32ERecalls & reported problems
Howmedica Osteonics Corp. · 0 Degree Polyethylene Insert · Class 2 · Implant
All models of Trident (160)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 13
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 07613327039450), received between Jan 5, 2018 and Aug 26, 2024 — classified as 13 injuries, 1 malfunction.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information6
- Adverse Event Without Identified Device or Use Problem4
- Unstable2
- Device-Device Incompatibility1
- Material Discolored1
Patient problems
- Injury8
- Unspecified Infection5
- Discomfort2
- Inflammation1
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
- Pain1
- Reaction1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
InjuryAug 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED BY STRYKER SALES REP THAT PATIENT IS REQUIRED FOR REVISION DUE TO INFECTION.
Patient problems: Unspecified Infection
MAUDE report 20063109 ↗InjuryFeb 19, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT REPORTED LEFT HIP DISCOMFORT AND WANTS TO KNOW IF PARTS ARE SUBJECT TO A RECALL.
Patient problems: Discomfort
MAUDE report 18729726 ↗InjuryNov 16, 2022Insufficient Device Problem Information; Insufficient Device Problem Information
THERE WAS NO EXPLANATION PROVIDED. THE PATIENT SELECTED 'NO' FOR SATISFACTION. A FOLLOW-UP QUESTIONNAIRE WAS COMPLETED BY THE SUBJECT WHERE THEY ANSWERED "NO" FOR SATISFACTION WITH THE RESULTS OF THE STUDY HIP REPLACEMENT. THE SUBJECT REPLIED THAT THEY ARE NOT SATISFIED WITH THEIR RIGHT HIP.
Patient problems: Insufficient Information
MAUDE report 15804739 ↗InjuryJun 28, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AT FIVE YEARS POSTOPERATIVE, THE SUBJECT REPORTED VERY GOOD HEALTH AND MODERATE PAIN OR DISCOMFORT IN THE STUDY HIP. AT THIS POINT, HE ANSWERED ¿NO¿ TO BEING SATISFIED WITH HIS STUDY HIP REPLACEMENT. THE STUDY COORDINATOR BELIEVES IT IS POSSIBLE THE SUBJECT MARKED "NO" IN ERROR AS THE QUESTIONNAIRES WERE COMPLETED VIA MAIL AND THE SUBJECT NEVER EXPRESSED ANY DISSATISFACTION WITH HIS HIP REPLACEMENT. AT SIX YEARS POSTOPERATIVE, THE SUBJECT ANSWERED ¿YES¿ TO BEING SATISFIED WITH HIS STUDY HIP REPLACEMENT ON PATIENT QUESTIONNAIRES.
Patient problems: Discomfort; Pain
MAUDE report 14844872 ↗MalfunctionJan 4, 2021Device-Device Incompatibility; Device-Device Incompatibility
THE 32 SIZE E WOULD NOT SEAT IN THE 54 SIZE E TRIDENT II SOLID SHELL AFTER MULTIPLE ATTEMPTS. 36 SIZE E IMPACTED ON FIRST ATTEMPT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11113647 ↗InjuryJan 4, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PROCEDURE: PATIENT UNDERWENT INDEX THA ON (B)(6) 2015. THEN THE PATIENT HAD A PERIPROSTHETIC JOINT INFECTION, SO ALL COMPONENTS WHERE REMOVED ON (B)(6) 2019. THE PATIENT WAS PLACED ON A SPACER. THEN THE PATIENT UNDERWENT REIMPLANTATION ON (B)(6) 2019, WITH OFF-LABEL USE IMPLANTS.
Patient problems: Unspecified Infection
MAUDE report 11107761 ↗InjuryOct 21, 2019Unstable; Unstable
IT WAS REPORTED THROUGH THE FILING OF A LAWSUIT THAT ALLEGEDLY THE PATIENT WAS IMPLANTED WITH AN LFIT ANATOMIC COCR V40 FEMORAL HEAD ON HER RIGHT HIP ON OR ABOUT (B)(6) 2011 AND WAS REVISED ON (B)(6) 2018. IT IS FURTHER ALLEGED THAT SHE SUFFERED INJURIES AS A RESULT OF IMPLANTATION AND EXPLANTATION OF THE DEVICE AT ISSUE, DEVICE RECALL, AND EXCESSIVE LEVELS OF CHROMIUM AND COBALT IN HER BLOOD. UPDATE 20/SEPTEMBER/2019: INFORMATION RECEIVED FROM STUDY SITE STATES "AFTER THE HIP REPLACEMENT, PATIENT PRESENTED WITH INSTABILITY FAILURE, THEN UNDERWENT REVISION WITH REMAINING CUP, OTHER COMPONENTS EXCHANGED (B)(6) 2018."
Patient problems: Inflammation; Injury; Reaction
MAUDE report 9216229 ↗InjuryMay 28, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
THIS PI IS FOR THE RIGHT HIP. IT WAS REPORTED THAT THE PATIENT HAD BILATERAL HIP REVISIONS DUE TO INFECTION (INFECTION REPORTED BY SURGEON. UNKNOWN IF CLINICALLY CONFIRMED OR IDENTIFIED). REP CAN PROVIDE PRE-REVISION X-RAYS AND REPORTED THAT NO FURTHER INFORMATION WILL BE AVAILABLE.
Patient problems: Injury; Unspecified Infection
MAUDE report 8645729 ↗InjuryDec 28, 2018Material Discolored; Material Discolored
IT WAS REPORTED THAT THA WAS DONE WITH TRIDENT HA SHELL AND X3 LINER AND STEM. THE SHELL WAS DISLOCATED IN ABOUT ONE YEAR AFTER THA. THEREFORE, ALL DEVICES WERE REMOVED AND REVISION SURGERY WAS DONE. THE REMOVED POLYETHYLENE LINER WAS YELLOWISH DISCOLORED. UPDATE DEC 17, 2018- CAUSE OF REVISION SHELL WAS DISLOCATED (MALPOSITION).
Patient problems: Injury
MAUDE report 8204034 ↗InjuryOct 16, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
PATIENT UNDERWENT PRIMARY RIGHT TOTAL HIP REPLACEMENT (B)(6) 2018. PATIENT RETURNED FOR INFECTED RIGHT TOTAL HIP REPLACEMENT REVISION. REIMPLANTATION OF IMPLANTS OCCURRED (B)(6) 2018.
Patient problems: Injury; Unspecified Infection
MAUDE report 7969682 ↗InjuryJul 10, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
PATIENT WITH PRIMARY THA ON (B)(6) 2014 BY DR. (B)(6), PATIENT DEVELOPED PJI AND REQUIRED RESECTION ARTHROPLASTY (1ST. STAGE) (ALL COMPONENTS EXCHANGED) BY DR. (B)(6) ON (B)(6) 2017. SECOND STAGE REIMPLANTATION WAS PERFORMED IN (B)(6) 2018 (OFF-LABEL) BY DR. (B)(6).
Patient problems: Injury; Unspecified Infection
MAUDE report 7674376 ↗InjuryJun 27, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
PATIENT HAD A WASH OUT OF RIGHT HIP AND THE HEAD AND LINER WERE REPLACED. REP HAS CONFIRMED THERE WAS NO INFECTION PRESENT. NO REASON FOR REVISION WAS GIVEN. HE HAS STATED NO FURTHER INFORMATION AVAILABLE.
Patient problems: Injury
MAUDE report 7641217 ↗InjuryApr 2, 2018Unstable; Unstable
REVISION DUE TO UNSTABLE LEFT HIP. IMPLANTED MDM LINER AND NEW FEMORAL HEAD".
Patient problems: Injury
MAUDE report 7390223 ↗InjuryJan 5, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT PATIENT'S LEFT HIP WAS REVISED DUE TO INSTABILITY.
Patient problems: Injury
MAUDE report 7167957 ↗
Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices
- NA 1365-51-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented753 reports
- NA 1365-50-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented412 reports
- ARTICUL/EZE 1365-21-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall280 reports
- ARTICUL/EZE 1365-11-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 death262 reports
- LFIT 6260-9-236Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, cemented256 reports
- PINNACLE GRIPTION 1217-32-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented2 deaths239 reports
- ELEOS™ 25002111EOnkos Surgical, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented3 deaths230 reports
- Novation 130-36-52Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented2 recalls124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page