Skip to content

HMRS 6465-6-018Recalls & reported problems

Howmedica Osteonics Corp. · Bearing · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07613327048216), received between Mar 15, 2022 and Jun 13, 2022 — classified as 3 injuries.

The most frequently reported specific device problem is “Degraded” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Degraded3
  • Naturally Worn3
  • Unstable2
  • Fracture1

Patient problems

  • Joint Laxity2
  • Ambulation or Postural Difficulties1
  • Hyperextension1
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 13, 2022Degraded; Fracture; Unstable; Naturally Worn; Degraded; Fracture; Unstable; Naturally Worn

    REVISION (HMRS PARTS REPLACEMENT AFTER 12 YEARS) ACCORDING TO BFARM SPECIAL APPROVAL 92.16-5640-S-023/22. UPDATE 25MAY22: AFTER 12 YEARS , THE POLY IS WORN OUT. THERE WAS A MASSIVE INSTABILITY IN THE KNEE JOINT. ESPECIALLY THE STOPPER ON THE PROXIMAL TIBIA WAS VERY CRUSHED.

    Patient problems: Joint Laxity

    MAUDE report 14673981 ↗
  • InjuryApr 11, 2022Degraded; Unstable; Naturally Worn; Degraded; Unstable; Naturally Worn

    THE FOLLOWING WAS REPORTED: REVISION (HMRS PARTS REPLACEMENT). AFTER 23 YEARS, THE POLY IS WORN OUT. IT IS MANIFESTED BY HYPEREXTENSION OF THE KNEE JOINT AND MARKED INSTABILITY. THE DISTAL FEMUR SHOWS SIGNS OF WEAR, WHICH SHOWS UP AS METALLOSIS IN THE KNEE JOINT. "NOW AFTER THE REVISION SURGERY EVERYTHING WAS FINE AGAIN."

    Patient problems: Ambulation or Postural Difficulties; Hyperextension; Joint Laxity

    MAUDE report 14073858 ↗
  • InjuryMar 15, 2022Degraded; Naturally Worn; Degraded; Naturally Worn

    HMRS PARTS REPLACEMENT AFTER 12 YEARS. THE WEAR OF THE COMPONENTS WAS CONFIRMED DURING THE SURGERY. THE PARTS HAVE ALL BEEN REPLACED AS REQUESTED. SOME POLY PARTS WERE SEVERELY DAMAGED OR DISASSEMBLED INTO INDIVIDUAL PARTS DUE TO THE DIFFICULT DISASSEMBLY. DUE TO WEAR AND TEAR, IT WAS SUSPECTED THAT REPLACING THE COMPONENTS WOULD BE PRUDENT.

    Patient problems: Insufficient Information

    MAUDE report 13767233 ↗

Other PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page