MRS 6485-3-000Recalls & reported problems
Howmedica Osteonics Corp. · Straight Cemented Stem · Class 2 · Implant
All models of MRS (25)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07613327048278), received between Jun 6, 2019 and Mar 22, 2022 — classified as 4 injuries.
The most frequently reported specific device problem is “Fracture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fracture3
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Failure to Osseointegrate1
Patient problems
- Ambulation or Postural Difficulties2
- Injury2
- Inadequate Osseointegration1
- Pain1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryMar 22, 2022Fracture; Adverse Event Without Identified Device or Use Problem; Fracture; Adverse Event Without Identified Device or Use Problem
PATIENT COMPLAINED OF LEG PAIN. XRAY SHOWED BENT OR BROKEN FEMUR STEM. REVISION SURGERY PERFORMED WITH NEW STEM.
Patient problems: Ambulation or Postural Difficulties; Pain
MAUDE report 13844968 ↗InjuryNov 17, 2020Fracture; Failure to Osseointegrate; Fracture; Failure to Osseointegrate
IT WAS REPORTED BY THE SALES REP THAT THE SURGEON INFORMED OF AN ISSUE OF A BROKEN GMRS STEM ON (B)(6) 2020 (CURRENT STEM DIAMETER 10MM REF 6485-3-000, LN 187138A) AND THE NEED TO REVISE TO A THICKER STEM (11 OR 13MM). CURRENT STEM WAS IMPLANTED ON (B)(6) 2019 (PRIMARY SURGERY). REVISION SURGERY IS PLANNED ON (B)(6) 2020. UPDATE: "SURGEON MENTIONED IMPLANT LIKELY HAVE LOOSENED AND BROKEN DUE TO PATIENT RIDING MOTORCYCLE AND USING HER LEGS TO STOP THE BIKE (EXCESSIVE STRESS TO IMPLANT) PATIENT HAD BROKEN IMPLANT IN (B)(6) 2020 BUT WAS UNABLE TO COME IN DUE TO COVID, HENCE THE DELAY TO (B)(6) SURGERY."
Patient problems: Ambulation or Postural Difficulties; Inadequate Osseointegration
MAUDE report 10854626 ↗InjuryDec 30, 2019Fracture; Fracture
InjuryJun 6, 2019Break; Break
IT WAS REPORTED THAT THE PATIENT'S LEFT KNEE WAS REVISED DUE TO A BROKEN MRS FEMORAL STEM. THE SURGEON REPORTED THAT THE MOST LIKELY CAUSE OR CONTRIBUTOR WAS THAT THE PATIENT WAS MOVING FURNITURE AT THE TIME. REP PROVIDED PRE- AND POST-REVISION X-RAYS AND EXPLANT PICTURES, AND REPORTED THAT NO FURTHER INFORMATION WILL BE AVAILABLE.
Patient problems: Injury
MAUDE report 8674329 ↗
Other PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER devices
- ELEOS™ 25001208EOnkos Surgical, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER3 deaths193 reports
- ELEOS™ THSMWRS01MOnkos Surgical, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER190 reports
- "UNITED" USTAR II Hip System 1115-1413United Orthopedic Corp. · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer3 recalls3 reports
- "UNITED" USTAR II Hip System 1115-1213United Orthopedic Corp. · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer3 recalls0 reports
- "UNITED" USTAR II Hip System 1115-1417United Orthopedic Corp. · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer3 recalls0 reports
- "UNITED" USTAR II Hip System 1115-1217United Orthopedic Corp. · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer3 recalls0 reports
- Endo-Model Modular Knee Prosthesis System 15-3816/12WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer2 recalls6 reports
- Endo-Model Modular Knee Prosthesis System 15-3816/11WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer2 recalls6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page